CERTIFIED 052-0059-8 Instruction Manual

CERTIFIED 052-0059-8 Instruction Manual

Pocket analog multimeter

Advertisement

model no. 052-0059-8 | contact us 1-800-689-9928
POCKET ANALOG MULTIMETER
IMPORTANT:
INSTRUCTION
Before using your analog multimeter, read
MANUAL
and understand this owner's guide.

Advertisement

Table of Contents
loading

Summary of Contents for CERTIFIED 052-0059-8

  • Page 1 052-0059-8 | contact us 1-800-689-9928 POCKET ANALOG MULTIMETER IMPORTANT: INSTRUCTION Before using your analog multimeter, read MANUAL and understand this owner’s guide.
  • Page 3: General Description

    GENERAL DESCRIPTION This meter is a rectifier type, permanent-magnet moving coil instrument for measuring DC voltage, AC voltage, DC current , resistance, decibels and for testing battery strength. INTRODUCTION 1. Pointer Figure 1 2. Zero set control 3. 0Ÿ adjuster knob 4.
  • Page 4: Safety Information

    052-0059-8 | contact us 1-800-689-9928 SAFETY INFORMATION The meter has been designed according to IEC-61010 concerning electronic measuring instruments with a measurement category III ( CAT III 300V ) and pollution degree 2. WARNING To avoid possible electric shock or personal injury, follow these guidelines: Do not use the meter if it is damaged.
  • Page 5 When using the probes, keep your fingers behind the finger guards on the probes. Connect the black test lead before you connect the red test lead. When you disconnect test leads, disconnect the red test lead first. Remove the test leads from the meter before you open the battery cover or the case.
  • Page 6: Electrical Symbols

    052-0059-8 | contact us 1-800-689-9928 CAUTION To avoid possible damage to the meter or to the equipment under test, follow these guidelines: Disconnect circuit power and discharge all capacitors before testing resistance and continuity. Use the proper function and range for your measurements.
  • Page 7: Technical Specification

    TECHNICAL SPECIFICATION Measuring Range DC Current: 0.6/12/600 mA Accuracy at FSD: 5% DC Voltage: 3/12/60/120/300 V Accuracy at FSD: 5% AC Voltage: 12/60/120/300 V Accuracy at FSD: 5% Resistance: Ÿ × 10 ( scale centre at 50 Ÿ ) Ÿ × 1k ( scale centre at 5 kŸ ) Accuracy: 5% of arc Battery Test: 1.5 V Decibel: -20 to 50 dB...
  • Page 8 052-0059-8 | contact us 1-800-689-9928 SCALE INSTRUCTION Figure 2 Common Markings for Voltage and Current Scales There are five scales: The upper scale is for resistance measurements. The second scale " " is for the measurements of DC voltage, AC voltage ( except AC 12 V range ), and DC current.
  • Page 9: Operating Instruction

    OPERATING INSTRUCTION Points to be Observed before Operation Before measuring, check the pointer to see whether it exactly rests at the "0" position, and if not, adjust the pointer with the zero set control ( Figure 3 ). Insert the red test lead to the "+" jack and black test lead to the "−"...
  • Page 10 052-0059-8 | contact us 1-800-689-9928 DC Voltage Measurement Set the selector switch to the desired " V " range. Connect the test leads across the circuit to be measured. Note: The black test lead must be connected to the negative potential of the circuit;...
  • Page 11 AC V Read Range Following Multiply Setting Scale Reading By: 0 - 12 0 - 60 0 - 12 0 - 300 DC Current Measurement Set the selector switch to the desired " mA " range. Discharge all capacitors. Break the circuit path to be measured, connect the red test lead to the more positive side of the break and the black test lead to the more negative side of the break.
  • Page 12 052-0059-8 | contact us 1-800-689-9928 Resistance Measurement Set the selector switch to the desired " Ÿ " range. Short the red and black test leads and turn the 0 Ÿ adjuster knob until the pointer rests in "0" position of the upper scale.
  • Page 13 G. Battery Measurement Set the range switch to the desired BATT. range ( 9 V or 1.5 V ). Connect the red test lead to the positive terminal of the battery to be tested and the black test lead to the negative terminal of the battery.
  • Page 14 052-0059-8 | contact us 1-800-689-9928 BATTERY AND FUSE REPLACEMENTS If the pointer can not be set to zero on the top scale by the 0 Ÿ adjuster knob while the test leads are being shorted, you should replace the battery.
  • Page 15 This Certified product carries a one (1) year warranty against defects in workmanship and materials. Should this Certified product become defective within the stated warranty period, return it to the store with proof of purchase, and it will be replaced or repaired free of charge.