• Snap the PMA
2000 Adapter gently into place on the side of either
®
the PMV
2000 (clear) or PMV
®
Tracheostomy & Ventilator Swallowing and Speaking Valve. The
curvatures of the two parts should fit tightly to one another, with no
gaps or interference from the small Secure-It
the side of the Passy-Muir
(see Fig.1B) of the PMA 2000 Adapter should overhang the front edge
of the Passy-Muir Valve in order to deliver oxygen directly in front of the
diaphragm of the valve.
•
After the PMA 2000 Adapter (with oxygen tubing attached) has been
snapped onto the Passy-Muir Valve, the PMV can then be placed on
the hub of the tracheostomy tube in the conventional manner (see
Passy-Muir Valve Instruction Booklet).
!
WARNING:
AS COMPARED TO DELIVERY OF OXYGEN VIA A TRACH
COLLAR, THE PMA 2000 ADAPTER IS AN OPEN OXYGEN DELIVERY SYSTEM.
THEREFORE, DELIVERY OF F
DIFFERENT THAN WITH THE USE OF A TRACH COLLAR. THE CLINICIAN
SHOULD USE METHODS SUCH AS PULSE OXIMETRY TO ENSURE PROPER
TITRATION OF O
WHILE USING THE PMA 2000 ADAPTER.
2
REMOVAL:
• Remove the PMV from the hub of the tracheostomy tube (as per Passy-
Muir Valve Instruction Booklet).
• Remove the PMA 2000 Adapter from the Passy-Muir Valve with a gentle
pulling motion.
• Remove the oxygen tubing from the PMA 2000 Adapter.
CARE AND CLEANING:
• The PMA 2000 Adapter should be cleaned daily to prevent debris from
accumulating within the oxygen port and chamber. Before cleaning, remove
the adapter from the Passy-Muir Valve and remove the oxygen tubing from
the PMA 2000 Adapter. Wash the PMA 2000 Adapter and Passy-Muir Valve
as separate units (see separate cleaning instructions for the Passy-Muir
Valve, provided in the Passy-Muir Valve Instruction Booklet).
• Swish the PMA 2000 Adapter in soapy, warm water.
• Rinse the PMA 2000 Adapter thoroughly with warm water.
• Allow the PMA 2000 Adapter to air dry thoroughly before placing it in
the storage container that is packaged with the Passy-Muir Valve.
• Do not apply heat to dry PMA 2000 Adapter. Do not use hot water,
peroxide, bleach, vinegar, or alcohol to clean PMA 2000 Adapter. Do
not autoclave.
© Passy-Muir, Inc. All rights reserved.
2001 (Purple Color
) Low Profile
®
™
®
ring that is molded into
®
Valve. When properly attached, the cowling
O
VIA THE PMA 2000 ADAPTER MAY BE
I
2
3
Passy-Muir
Passy-Muir
for use with PMV
For technical questions regarding utilization of the PMA 2000 Adapter
or to receive our free research literature packet, "Physiological
Benefits" CD-ROM and clinical inservice DVD, please
contact our respiratory and speech clinical specialists.
Touching Lives and Advancing Patient Care Through Education
Passy-Muir, Inc
Irvine, CA USA 92614-6813
949.833.8255 • 800.634.5397 • Fax: 949.833.8299
website: www.passy-muir.com
email: info@passy-muir.com
IBO
/R0317
2
Valve Oxygen Adapter
®
(PMA
2000)
®
Instructions For Use
Valve Oxygen Adapter (PMA
®
2000 (clear) and PMV
®
Low Profile Tracheostomy & Ventilator
Swallowing and Speaking Valves
2000)
®
2001 (Purple Color
)
®
™
0050