• Snap the PMA
2000 Adapter gently into place on the side of either
®
the PMV
2000 (clear) or PMV
®
Tracheostomy & Ventilator Swallowing and Speaking Valve. The
curvatures of the two parts should fit tightly to one another, with no
gaps or interference from the small Secure-It
the side of the Passy-Muir
(see Fig.1B) of the PMA 2000 Adapter should overhang the front edge
of the Passy-Muir Valve in order to deliver oxygen directly in front of the
diaphragm of the valve.
After the PMA 2000 Adapter (with oxygen tubing attached) has been
snapped onto the Passy-Muir Valve, the PMV can then be placed on
the hub of the tracheostomy tube in the conventional manner (see
Passy-Muir Valve Instruction Booklet).
!
WARNING:
AS COMPARED TO DELIVERY OF OXYGEN VIA A TRACH
COLLAR, THE PMA 2000 ADAPTER IS AN OPEN OXYGEN DELIVERY SYSTEM.
THEREFORE, DELIVERY OF F
DIFFERENT THAN WITH THE USE OF A TRACH COLLAR. THE CLINICIAN
SHOULD USE METHODS SUCH AS PULSE OXIMETRY TO ENSURE PROPER
TITRATION OF O
WHILE USING THE PMA 2000 ADAPTER.
2
REMOVAL:
• Remove the PMV from the hub of the tracheostomy tube (as per Passy-
Muir Valve Instruction Booklet).
• Remove the PMA 2000 Adapter from the Passy-Muir Valve with a gentle
pulling motion.
• Remove the oxygen tubing from the PMA 2000 Adapter.
CARE AND CLEANING:
• The PMA 2000 Adapter should be cleaned daily to prevent debris from
accumulating within the oxygen port and chamber. Before cleaning, remove
the adapter from the Passy-Muir Valve and remove the oxygen tubing from
the PMA 2000 Adapter. Wash the PMA 2000 Adapter and Passy-Muir Valve
as separate units (see separate cleaning instructions for the Passy-Muir
Valve, provided in the Passy-Muir Valve Instruction Booklet).
• Swish the PMA 2000 Adapter in soapy, warm water.
• Rinse the PMA 2000 Adapter thoroughly with warm water.
• Allow the PMA 2000 Adapter to air dry thoroughly before placing it in
the storage container that is packaged with the Passy-Muir Valve.
• Do not apply heat to dry PMA 2000 Adapter. Do not use hot water,
peroxide, bleach, vinegar, or alcohol to clean PMA 2000 Adapter. Do
not autoclave.
© Passy-Muir, Inc. All rights reserved.
2001 (Purple Color
) Low Profile
®
®
ring that is molded into
®
Valve. When properly attached, the cowling
O
VIA THE PMA 2000 ADAPTER MAY BE
I
2
3
Passy-Muir
Passy-Muir
for use with PMV
For technical questions regarding utilization of the PMA 2000 Adapter
or to receive our free research literature packet, "Physiological
Benefits" CD-ROM and clinical inservice DVD, please
contact our respiratory and speech clinical specialists.
Touching Lives and Advancing Patient Care Through Education
Passy-Muir, Inc
Irvine, CA USA 92614-6813
949.833.8255 • 800.634.5397 • Fax: 949.833.8299
website: www.passy-muir.com
email: info@passy-muir.com
IBO
/R0317
2
Valve Oxygen Adapter
®
(PMA
2000)
®
Instructions For Use
Valve Oxygen Adapter (PMA
®
2000 (clear) and PMV
®
Low Profile Tracheostomy & Ventilator
Swallowing and Speaking Valves
2000)
®
2001 (Purple Color
)
®
0050

Advertisement

loading

Summary of Contents for Passy-Muir PMA 2000

  • Page 1 Passy-Muir Valve. Passy-Muir, Inc • Do not apply heat to dry PMA 2000 Adapter. Do not use hot water, Irvine, CA USA 92614-6813 peroxide, bleach, vinegar, or alcohol to clean PMA 2000 Adapter. Do 949.833.8255 •...
  • Page 2 • Oxygen conservation devices that require flow triggering cannot be used WARNING: NOT INTENDED FOR USE IN-LINE DURING MECHANICAL with the PMA 2000 Adapter, as it may not allow the patient to properly VENTILATION. trigger the gas source, thereby decreasing the amount of oxygen DESCRIPTION: delivered to the patient.