Summary of Contents for RKI Instruments 61-1000RK
Page 1
61-1000RK / 61-0140RK Combustible Gas Detector Operator’s Manual Part Number: 71-0067RK Revision: E Released: 10/26/18 www.rkiinstruments.com...
Page 2
Typical calibration frequencies for most applications are between 3 and 6 months, but can be required more often or less often based on your usage. 61-1000RK/61-0140RK Combustible Gas Detector...
Page 3
Product Warranty RKI Instruments, Inc. warranties gas alarm equipment sold by us to be free from defects in materials, workmanship, and performance for a period of one year from date of shipment from RKI Instruments, Inc. Any parts found defective within that period will be repaired or replaced, at our option, free of charge.
± 5% of reading or ± 2% LEL (whichever is greater) WARNING: When using the 61-1000RK, you must follow the instructions and warnings in this manual to assure proper and safe operation of the 61-1000RK and to minimize the risk of personal injury. Be sure to maintain and periodically calibrate the 61-1000RK as described in this manual.
Description This section describes the components of the 61-0140RK and 61-1000RK detectors. The 61-1000RK includes the 61-0140RK combustible gas detector and a junction box. A four point terminal strip is provided inside the junction box for detector connections. The 61-0140RK does not include a junction box.
61-1000RK detector. 10. If using shielded cable, connect the cable’s drain wire to an available chassis ground at the controller. RKI controllers typically have a ground stud that is a convenient grounding location. 8 • 61-1000RK/61-0140RK Combustible Gas Detector...
Page 9
Cable Shield Controller White Detector Green Terminals Black J-Box Terminal Strip Black Green Detector W ires White Combustible Detector Figure 3: Wiring the Combustible Gas Detector to a Controller 61-1000RK/61-0140RK Combustible Gas Detector • 9...
If the display reading is 0% LEL, start up is complete. The combustible detector is in normal operation. If the display reading is not 0% LEL, continue with step 3. Perform a zeroing operation at the controller. See the controller operator’s manual for directions. 10 • 61-1000RK/61-0140RK Combustible Gas Detector...
4. If the calibration/response difficulties detector requires continue, contact RKI for further calibration once every instruction. three months. Some applications may require a more frequent calibration schedule. 61-1000RK/61-0140RK Combustible Gas Detector • 11...
(see Figure 3). See the controller operator’s manual and the controller’s detector head specification sheet for the 61-1000RK detector to verify the connections to the controller are correct.
Place the controller into its calibration program or disable external alarms. NOTE: Calibrating the combustible detector may cause alarms. Be sure to put the controller into its calibration program or disable external alarms before continuing. 61-1000RK/61-0140RK Combustible Gas Detector • 13...
NOTE: If you do not allow the gas reading to decrease below the alarm points, then unwanted alarms may occur. Verify that the controller display reading decreases and stabilizes at 0 %LEL. Store the components of the calibration kit in a safe and convenient place. 14 • 61-1000RK/61-0140RK Combustible Gas Detector...
Parts List Table 3 lists replacement parts and accessories for the 61-1000RK combustible gas detector. Table 3: Parts List Part Number Description 18-0400RK-01 Junction box with spacers 61-0140RK Replacement combustible gas detector 71-0067RK 61-1000RK Combustible Gas Detector Operator’s Manual (this document)