parkell INTEGRA D680 Instructions For Use Manual

Ultrasonic scaler

Advertisement

Quick Links

This manual, in whole or in part, should not be considered a substitute for formal training in Ultrasonic Scaling.
Appropriate professional education is STRONGLY RECOMMENDED prior to using this device clinically.
U LT R A S O N I C S C A L E R
DEVICE DESCRIPTION
The Parkell Integra™ is a magnetostrictive, auto-
tuning, ultrasonic tooth scaler that operates at either
25 KHz or 30 KHz. It has an integrated, two-bottle,
pump-driven fluid supply, and is also capable of
obtaining fluid from an external water supply.
The Integra automatically detects whether the
handpiece contains a 25 KHz or 30 KHz insert, and
switches to the correct operating frequency. The
Integra has an expanded low-power range to improve
patient comfort, and a power-boosting, user-controlled
TURBO feature to increase scaling power when needed
for heavy deposits.
INTENDED USE/INDICATIONS
For removal of calculus, plaque and oral debris from
teeth during dental prophylaxis/scaling.
CONTRAINDICATIONS TO USE
Because of the potential for electromagnetic
interference, this device should not be used on
patients or by clinicians with cardiac pacemakers,
internal defibrillators, intracorporeal fluid pumps or
any other implantable electronic devices, or in close
proximity to sensitive patient monitoring devices such
as pulse oximeters.
IMPORTANT INFORMATION REGARD-
ING USE OF NON-PARKELL INSERTS:
Parkell ultrasonic scalers are compatible with most
major brands of ultrasonic inserts. For optimal
performance of your scaler and insert, we recommend
using Parkell ultrasonic inserts. Parkell cannot
guarantee that all non-Parkell inserts available on the
market today or in the future will work properly with
our scalers. For more information, contact Parkell.
Shown with Parkell's Burnett
Power-Tip Insert (not included)
CAUTION: RX ONLY
WARNINGS
• The water supply to the scaler should always be
turned off whenever the device is being connected,
disconnected, or when not in use.
• Do not run this device dry. Water must always flow
through tip during use to cool handpiece and insert.
• The power supply line cord functions as the AC
power disconnect device for the scaler.
• The transformer should be located above and away
from any sources of water. As with all electrical
devices, the unit should not be immersed in water
or other liquids. Do not reach for the device if it has
fallen into liquid until power is disconnected, and
do not use the device after it has fallen into liquid.
Return the device to Parkell for servicing.
• Modification of this device will void the warranty, and
may violate safety codes, endangering the patient
and/or the clinician.
• This equipment is not suitable for use in the
presence of a flammable anesthetic gas mixture
(when used along with air or oxygen).
• This equipment produces electromagnetic energy
and may cause interference with other nearby
electronic devices. Should this occur, changing the
position or location of the device
may be necessary.
INSTRUCTIONS FOR USE
STOCK NO. D680
24-HOUR SAFETY HOTLINE
1-800-535-5053
A00545revC0317

Advertisement

Table of Contents
loading

Summary of Contents for parkell INTEGRA D680

  • Page 1 For optimal performance of your scaler and insert, we recommend using Parkell ultrasonic inserts. Parkell cannot guarantee that all non-Parkell inserts available on the market today or in the future will work properly with our scalers. For more information, contact Parkell.
  • Page 2: Clinical Precautions

    However, they may be included in a scaler package that you have chosen to purchase. Parkell • Protect the patient’s clothing from water damage scalers are compatible with most major brands of 25 when using this device.
  • Page 3 USING THE INTERNAL FLUID SUPPLY BOTTLES The Parkell Integra is equipped with two internal bottles which can supply the fluid that flows through the handpiece for cooling and lavage during use. The bottles may be filled with drinking-quality water, or diluted with anti-microbials at a ratio of 4 parts water to 1 part chlorhexidine or mouthwash.
  • Page 4: Electrical Connection

    The scaler must only be powered via the Transformer “External”, from the rear connector to the city water, Assembly that is supplied with the device by Parkell. The or “Internal”, from the bottles inside the Integra. scaler and its separate transformer generate a minimal •...
  • Page 5 Press the foot-pedal to activate the scaling function. Note: There is a small switch on the rear of the foot Release the foot-pedal to deactivate. Whenever you pedal to allow you to enable or disable the POWER release the pedal, the ACTIVE light on the front panel button.
  • Page 6 • Insufficient water flow. to wake it up. • Use of a heated water supply causing excessive water • Foot pedal or cable is damaged. Contact Parkell. temperature at the insert tip. • Unit fuse has failed. Contact Parkell. • Scaling with the point of the insert instead of the side.
  • Page 7: Warranty

    “Submit” button. Store the printout in a safe place. Within the US, all repairs must be made by Parkell. We maintain a complete service and parts facility in our For full Warranty and Terms of Use information, factory at 300 Executive Drive, Edgewood, NY 11717.
  • Page 8 IS SERVICED IN THE UNITED STATES OF AMERICA BY: 300 Executive Drive, Edgewood, NY 11717 Toll-Free: 1-800-243-7446 • Phone: 631-249-1134 Fax: 631-249-1242 • E-mail: info@parkell.com www.parkell.com European Authorized Representative (Not a dealer/distributor): Directa AB - P.O. Box 723, Finvids väg 8 SE-194 27 Upplands Väsby - Sweden...