JB02 Junction box user manual
Content
-
Specifications
Installation guidelines
-
Dimensions
-
-
Mounting
Electrical connections
-
Calibration
-
-
I Corner adjustment
II Span adjustment
-
III Trouble-shooting
-
Specifications
Die-cast aluminium housing
Plastic cable glands
Available for 2, 3 or 4 load cells connection
Excitation trim (28 turns, 5M Ohm)
4 wire load cell connection
4 wire output connection
Environmental protection IP65
Screw-clamp terminals and soldering connections
Installation guidelines
The junction box model JB02 provides a low cost answer for a variety of weighing applications. Excitation
trim makes it fast and easy to adjust the load cell outputs.
Model JB02-2 offers connection and individual adjustment for up to 2 load cells. Models JB02-3 and JB02-
4 offer connection and sectional adjustment for 3 and 4 load cells respectively and offer the user the
possibility to connect further junction boxes where required. Cable glands that are not used need to be
closed with suitable plugs.
Model JB02 has a 5-way terminal strip for each load cell, enabling connection of 4 and 6-wire load cells,
with screen. Connection to the instrument is achieved via a 5-way terminal strip and is also possible by
soldering points. Terminal strips are screw-operated cage clamp design.
To overcome disturbance, make sure not to create ground-loops whilst connecting the cable screens.
Confirm ( via the load cell data sheet) whether or not the load cell has its screen connected to its body.
Trimming of signal output is accomplished using 28-turn trim potentiometers. The operating temperature
range of the Junction Box is -40°C to 85°C.
Dimensions
Width 65mm
Height 38mm
Length 135mm
(dimensions without cable glands)\
Zemic Europe B.V.
Leerlooierstraat 8
4871 EN Etten-Leur
The Netherlands
Nr. 2015.10 JB02 Junction box user manual Rev0
1/3
T: +31 76 50 39480
F: +31 76 50 39481
info@zemic.nl
www.zemic.nl

Advertisement

loading

Summary of Contents for Zemic JB02

  • Page 1 Model JB02 has a 5-way terminal strip for each load cell, enabling connection of 4 and 6-wire load cells, with screen. Connection to the instrument is achieved via a 5-way terminal strip and is also possible by soldering points.
  • Page 2 Fig 1 (load cell and indicator connection) It is possible for the all JB02 to be used with fewer load cells than for which they are designed. Calibration After all wiring is completed and the scale instrument is powered up, turn each of the potentiometers fully anti-clockwise (maximum 28 turns, click will be heard and/or felt), to obtain the highest possible output from each load cell.
  • Page 3 - EMC directive 2004/108/EC in relation to the harmonised standards EN61000-6-1:2001 and EN61000-6-3:2001. - RoHS directive 2002/95/EC. Zemic Europe B.V. T: +31 76 50 39480 Leerlooierstraat 8 Nr. 2015.10 JB02 Junction box user manual Rev0 F: +31 76 50 39481 4871 EN Etten-Leur info@zemic.nl The Netherlands...