Advertisement

E
volution
TM
O X Y G E N C O N S E R V E R
PRODUCT INFORMATION
AND INSTRUCTIONS
OM-900

Advertisement

Table of Contents
loading

Summary of Contents for CHAD Evolution OM-900

  • Page 1 volution O X Y G E N C O N S E R V E R PRODUCT INFORMATION AND INSTRUCTIONS OM-900...
  • Page 2: Table Of Contents

    TABLE OF CONTENTS General Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 Important Safety Rules and Precautions Operating the Unit .
  • Page 3 TABLE OF CONTENTS Care and Maintenance Hygienic Preparation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16 Intervals .
  • Page 4: General Information

    GENERAL INFORMATION This manual provides the information necessary to operate the EVOLUTION electronic oxygen conserver with built-in regulator in accordance with a physician’s prescription. The EVOLUTION conserver can be used with any Cylinder Valve CGA 870 post-valve cylinder (see Fig . A) at home Toggle or away from home to provide your specific oxygen requirements .
  • Page 5: Important Safety Rules And Precautions

    IMPORTANT SAFETY RULES & PRECAUTIONS Read this instruction manual carefully to ensure your complete understanding before operating your EVOLUTION electronic conserver . This manual is part of the unit and must be available at all times . Use the unit for the designated purpose only (see “Purpose” on page 7) . For your own safety and that of your patients, please observe the following points: OPERATING THE UNIT WARNING!
  • Page 6: Operating The Unit

    IMPORTANT SAFETY RULES & PRECAUTIONS WARNING! (Cont.) Please remember that this is for your own safety! • Keep all parts free of oil and grease. Hydrocarbon compounds such as oil, grease, petroleum- based products, cleaning agents containing alcohol, hand cream or adhesive bandages can cause explosive reactions if they come into contact with highly compressed oxygen .
  • Page 7: Accessories/Repairs

    . • Servicing and repair work must only be carried out by the manufacturer (CHAD Therapeutics) or by trained personnel . Refer repairs to authorized personnel .
  • Page 8: Introduction

    INTRODUCTION PURPOSE The EVOLUTION electronic conserver is designed for use as part of a portable ambulatory oxygen system for the purpose of facilitating long-term oxygen therapy (LTOT) . It provides mobile LTOT patients with an extended operating time of their mobile system . The EVOLUTION conserver includes a combination of a low-pressure regulator and an oxygen conserver and is capable of delivering a precise amount of supplemental oxygen at the optimal point in the breathing cycle .
  • Page 9: Function

    INTRODUCTION FUNCTION When we breathe, approximately one-third of the time is spent inhaling, and two-thirds exhaling . As a result, oxygen delivered by continuous flow is wasted during exhalation . By eliminating oxygen flow during exhalation, a two-thirds savings is possible . Additionally, the application of the EVOLUTION is based on the fact that only oxygen inhaled at the beginning of the breath actually reaches the alveoli and is absorbed by the body .
  • Page 10: Description Of Parts And Controls

    DESCRIPTION OF PARTS & CONTROLS FIGURE B Low Battery Selector EVOLUTION Cylinder Conserver Indicator Light Knob Adjustment Top View Handle Selector Button OXYGEN INLET CONNECTION Cylinder Oxygen Alignment Pressure Seal Washer Pins Gauge FIGURE C EVOLUTION Conserver Side View Battery FIGURE D Oxygen Supply Holder...
  • Page 11: Description Of Parts And Controls

    DESCRIPTION OF PARTS & CONTROLS • Low Battery Indicator Light: The EVOLUTION conserver incorporates a visual indicator light that alerts you when a battery change is needed . The following signal will be displayed as your batteries become depleted: Blinking red light - Low battery energy level . Replace batteries . •...
  • Page 12: Assembly And Use

    ASSEMBLY AND USE • Make certain that your hands are free of oil, grease, and other contaminants . • Inspect the unit to insure that it has a manufacturer-specified seal washer in good working condition attached to the inlet nozzle . •...
  • Page 13: Installing Alkaline Batteries

    ASSEMBLY AND USE INSTALLING ALKALINE BATTERIES The EVOLUTION conserver uses advanced technology that prolongs battery life through efficient power use . With normal use of four (4) hours per day at any setting, your batteries should last a minimum of 2 years . Two (2) AA alkaline batteries are packaged in the box with your EVOLUTION conserver .
  • Page 14: Monitoring Battery Energy Level

    ASSEMBLY AND USE MONITORING BATTERY ENERGY LEVEL: The EVOLUTION conserver is equipped with a low battery indicator light that alerts you when a battery change is needed . The following signal will be displayed as your batteries become depleted: Blinking red light – Low battery energy level . Replace batteries . Disposal of Batteries: Do not dispose of used batteries in the household waste .
  • Page 15: Operating Instructions

    ASSEMBLY AND USE OPERATING INSTRUCTIONS: STEP 1: Make sure the EVOLUTION conserver is set to the “PULSE” position . Battery saving technology has eliminated the need for an “OFF” position . The unit will go to “Sleep Mode” after three minutes of non-use . The unit will turn on again when an inspiration is detected or the Selector Button is pressed .
  • Page 16: Oxygen Cylinder Duration

    OXYGEN CYLINDER DURATION Because the total delivery of oxygen via the EVOLUTION conserver is related to breathing rates, it is user adaptive in that the total oxygen delivered per minute will automatically adjust with user need, as expressed by increased or decreased breathing rates . For example, at all settings, twice as much oxygen per minute will be delivered if one breathes twenty (20) times per minute as compared with ten (10) times per minute .
  • Page 17: Care And Maintenance

    CARE AND MAINTENANCE The EVOLUTION conserver is designed for a long and accurate life; however, as with any electronic device, prudent care is required . The unit should be kept clean and free from moisture and dust, as well as extreme temperature . Do not expose the unit to water, such as when bathing or swimming . It is advisable to keep the device in a carrying bag to afford a degree of protection .
  • Page 18: Procedure

    CARE AND MAINTENANCE PROCEDURE Carry out hygienic preparation of the unit and accessories as described on page 16 . We recommend using a 0 .5% TERRALIN solution for wipe disinfection . Follow the instructions enclosed with the disinfectant . WARNING! •...
  • Page 19: Functional Check

    CARE AND MAINTENANCE FUNCTIONAL CHECK CHECKING FOR LEAKS 1 . Close the valve on the oxygen cylinder . 2 . Depressurize the EVOLUTION by inhaling several times using the nasal cannula . The gauge indicator should drop to zero . 3 .
  • Page 20: Products, Spare Parts, Accessories

    PRODUCTS, SPARE PARTS, ACCESSORIES STANDARD PRODUCT ORDER NUMBER DESCRIPTION OM-900 Standard EVOLUTION with CGA 870 Connection SPARE PARTS ORDER NUMBER DESCRIPTION RP-3040 Black “S” Cylinder Adjustment Handle FR-870G Seal Washers (U .S . Only), bag of 10 ACCESSORIES ORDER NUMBER DESCRIPTION OP-150-800 3-in-1 carry bag, fits M4, M6, M7, and M9 cylinders...
  • Page 21: Troubleshooting Guide

    TROUBLESHOOTING GUIDE PROBLEM PROBABLE CAUSE SOLUTION Open the battery door and remove the Software needs to be reset batteries . Wait 10 seconds . Replace the batteries and close the battery door . Replace or recharge batteries. Dead batteries. Make sure battery polarity is correct. Unit does not Batteries installed pulse.
  • Page 22: Classifications And Specifications

    CLASSIFICATIONS AND SPECIFICATIONS TECHNICAL DATA SPECIFICATIONS EVOLUTION Product class according to 93/42/EEC IIb Dimensions (L x H x W) 6 .1” L (15 .5 cm) x 2 .5” H (6 .4 cm) x 3 .1” W (7 .9 cm) Weight Approximately 14 .9 ounces (422 grams) with batteries Input pressure 200 to 3000 PSI (13 .8 bar to 206 .8 bar)
  • Page 23 CLASSIFICATIONS AND SPECIFICATIONS SPECIFICATIONS (Cont.) Oxygen delivery at level: Liter Flow Equivalency Continuous flow 2 LPM ±0 .5 LPM Maximum breathing rate 40 breaths per minute Power supply • Batteries (2) x 1 .5 V alkaline AA or LR6-type rechargeable batteries Low battery indicator light: •...
  • Page 24 CLASSIFICATIONS AND SPECIFICATIONS SPECIFICATIONS (Cont.) Classification according to EN 60601-1 • Degree of protection against electric shock Type BF • Degree of protection against harmful ingress of water IPX1 as per IEC 60529 Electromagnetic compatibility (EMC) according to EN 60601-1-2 •...
  • Page 25: Separation Distances

    CLASSIFICATIONS AND SPECIFICATIONS SEPARATION DISTANCES Recommended separation distances between portable and mobile RF Communications equipment and the EVOLUTION The EVOLUTION is intended for use in the electromagnetic environment in which radiated RF disturbances are controlled . The customer or user of the EVOLUTION can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications (transmitters) and the EVOLUTION...
  • Page 26: Symbols Key

    CLASSIFICATIONS AND SPECIFICATIONS SYMBOLS KEY SYMBOL MEANING Warning, consult accompanying documents Refer to the instruction manual TYPE PLATE Year manufactured Degree of protection against electric shock: type BF unit Type of protection against electric shock: protection class II unit Do not dispose of the unit in the household waste Serial number of the unit IPX1 The EVOLUTION...
  • Page 27: Limited Warranty

    Therapeutics’ obligation shall be limited to the replacement or repair of any such units or parts that prove, by CHAD’s inspection, to be defective within two years from the date of purchase. Any abuse, operation other than the intended use of the product as outlined in this manual, negligence, accident or repair by other than authorized service professionals shall immediately void this warranty .
  • Page 28: Important Information To Record

    IMPORTANT INFORMATION TO RECORD Your Name: __________________________________________________________________ Date You Received Your Unit: ____________________________________________________ Prescribed Oxygen Flow Setting: • At Rest: ____________________________ • During Exercise: _____________________ Home Care Provider’s Name: _____________________________________________________ Home Care Provider’s Phone Number: (_______) _____________________________________ Physician’s Name: ______________________________________________________________ Physician’s Phone Number: (_______) ______________________________________________ Notes: __________________________________________________________________...
  • Page 29: Notes

    NOTES...
  • Page 30 NOTES...
  • Page 31 NOTES...
  • Page 32 Drive Medical Design & Manufacturing 99 Seaview Boulevard PM-00040/13/F Port Washington, NY 11050 Printed in the U .S .A . Toll-free: 877-224-0946 CHAD is a registered trademark of Inovo, Inc . www.drivemedical.com EVOLUTION is a trademark of Inovo, Inc .

Table of Contents