Advertisement

Quick Links

INSTRUCTION MANUAL – V1.1x B
INTRODUCTION
myPCLab is a very compact Data Acquisition tool that connects to a PC via a USB port and monitors two universal input
analog variables along with one digital input. From hobbyists to scientists, from simple technical tasks to complex
engineering activities, myPCLab can be an invaluable tool for on-line monitoring and data logging in school, laboratory
research, machine data recording and industrial understanding.
It comes with intuitive and easy-to-use Windows software which plots and records data, shows gauges, bargraphs and
digital readouts. The USB port is accessed as a virtual serial port, which makes myPCLab compatible with SCADA
software with Modbus RTU driver or with any program with ASCII serial communication.
Multiple units of myPCLab can be connected to the same computer, easily expanding the number of inputs. The
myPCLab software can simultaneously communicate with multiple modules.
This manual describes software installation and connection of input signals to a myPCLab module. Instructions on using
the software are available on its help system.
READ THIS RECOMMENDATION TO AVOID DAMAGE TO YOUR COMPUTER.
myPCLab inputs are not electrically isolated from the USB port. Before connecting signals from sensors or
transducers, make sure that they are not in contact with any source of electrical power (wires or conductive
surfaces which may get energized). Input channels are not isolated from each other.

SPECIFICATIONS

Analog Input Signals: Configurable for channels 1 and 2. The input signals and measuring ranges are listed in the
following table.
Thermocouples: J, K, T, E, N, R, S and B types.
Pt100:
Linear Signals:
J Thermocouple
K Thermocouple
T Thermocouple
E Thermocouple
N Thermocouple
R Thermocouple
S Thermocouple
B Thermocouple
Pt100 RTD
0 to 50 mVdc
0 to 10 Vdc (Channel 1)
4 to 20 mAdc
NOVUS AUTOMATION
my
PCLab
Not compatible with Windows® 10.
Input Impedance >> 1 MΩ
Three wires connection, 625 µA bias current, α = 0.00385.
0 to 50 mVdc: Input Impedance >> 1 MΩ
0 to 10 Vdc: Input Impedance = 1 MΩ. Available only in channel 1.
4 to 20 mAdc: Input Impedance: 22 Ω (+ 2,0 Vdc)
Input Type
-130,0 to 940,0 °C (-202,0 to 1724,0 ºF)
-200,0 to 1370,0 °C (-328,0 to 2498,0 ºF)
-200,0 to 400,0 °C (-328,0 to 752,0 ºF)
-100,0 to 720,0 °C (-148,0 to 1328,0 ºF)
-200,0 to 1300,0 °C (-328,0 to 2372,0 ºF)
0,0 to 1760,0 °C (-32,0 to 3200,0 ºF)
0,0 to 1760,0 °C (-32,0 to 3200,0 ºF)
500,0 to 1800,0 °C (932,0 to 3272,0 ºF)
-200,0 to 650,0 °C (-328,0 to 1202,0 ºF)
Adjustable from -32767 to +32767
Adjustable from -32767 to +32767
Adjustable from -32767 to +32767
myPCLab module sensors and signals
Maximum Range
1/9

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the myPCLab and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for Novus myPCLab

  • Page 1: Specifications

    INTRODUCTION myPCLab is a very compact Data Acquisition tool that connects to a PC via a USB port and monitors two universal input analog variables along with one digital input. From hobbyists to scientists, from simple technical tasks to complex engineering activities, myPCLab can be an invaluable tool for on-line monitoring and data logging in school, laboratory research, machine data recording and industrial understanding.
  • Page 2 Operating environment: 0 to 50°C, 10 to 90% relative humidity, non-condensing. Electromagnetic Compatibility: EN 50081-2, EN 50082-2. Built-in cold junction compensation for thermocouples and wire resistance for Pt100. ABS case, dimensions: 70 x 60 x 18 mm. See figure. NOVUS AUTOMATION...
  • Page 3: Thermocouple Wiring

    If compensation cable is not used or is not connected with the correct polarity, large measurement errors will arise. 4-20 mA WIRING Connections of channel 1 are on terminals 6 and 7. For channel 2 on terminals 2 and 3. Observe indicated polarity. NOVUS AUTOMATION...
  • Page 4: Dry Contact Digital Input Wiring

    The contact connected to digital input must not have any electrical voltage. VOLTAGE DIGITAL INPUT WIRING Digital signals are connected to channel 3, terminals 1 and 2. Observe indicated polarity. Voltage levels applied to this channel must be within the specified limits. See “Specifications”. NOVUS AUTOMATION...
  • Page 5: Usb Driver Installation

    6. If a warning message regarding Windows® XP compatibility appears, select “Continue Anyway”. 7. The myPCLab driver files will be copied to your computer and, when concluded, a window will show up informing that the wizard has finished installing the software. Select “Finish”.
  • Page 6: Software Installation

    SERIAL PORT (COM PORT) ASSIGNMENT A few seconds after connection of a myPCLab, Windows operating system assigns a COM port number for communication. The assigned COM port number will not change in future connections to the same USB port. Users can easily identify and modify the assigned COM port in: Control Panel / System / Hardware / Device Manager / Ports (COM &...
  • Page 7: Warranty

    PCLab WARRANTY Warranty conditions are available on our web site www.novusautomation.com/warranty. NOVUS AUTOMATION...
  • Page 8: Appendix 1 - Serial Communication Protocols

    APPENDIX 1 – SERIAL COMMUNICATION PROTOCOLS Each myPCLab module installed in a PC is accessible as a distinct serial COM port. A module can communicate in one of two protocols: ASCII (delimited text) or Modbus RTU. Any program with serial communication capability using one of these protocols may be used to gather measurement data from the modules.
  • Page 9 ASCII – DELIMITED TEXT (AUT0-SEND MODE ENABLED) When configured to the auto-send mode, the myPCLab module transmits to its USB port a line of text containing the current value of all measurement channels. A new line of text is transmitted after each measurement interval, according to the module configuration.

Table of Contents