MODEL 2911 HIGH FREQUENCY SHAKER ASSEMBLY INSTRUCTION MANUAL San Juan Capistrano, California, U.S.A. IM2911 Revision A 28 September 2007 DOCUMENT SUBJECT TO U.S. EXPORT CONTROLS. COMPLIANCE APPROVAL REQUIRED PIOR TO DISTRIBUTION.
Page 2
IM2911 Revision A 28 September 2007 DOCUMENT SUBJECT TO U.S. EXPORT CONTROLS. COMPLIANCE APPROVAL REQUIRED PIOR TO DISTRIBUTION.
ENDEVCO 2911 INSTRUCTION MANUAL TABLE OF CONTENTS SECTION 1: DESCRIPTION PAGE EQUIPMENT DESCRIPTION EQUIPMENT USE SECTION 2: CONTROLS AND CONNECTORS INTRODUCTION SECTION 3: OPERATING INSTRUCTIONS UNPACKING AND INSPECTION INSTALLATION CHECKOUT OPERATION SECTION 4: THEORY OF OPERATION THEORY OF OPERATION SECTION 5: MAINTENANCE AND CALIBRATION...
The armature is made of Beryllium, which is quite stiff and light. This combination results in a resonant frequency around 40 kHz. This helps give the 2911 its wide operating frequency range in a small, compact and affordable package.
INTRODUCTION The Model 2911 has an input connector for the drive signal. See S ection 7 for more detailed shaker connection information. The ¼ inch plumbing connector is for clean dry air free of oil, water or particles bigger than 10 microns.
(Consult IM68357 for specific instructions.) Then selecting and starting a sweep that specifies the 2911 shaker to be used will activate the shaker. After the audible “click” of the air solenoid, push down on the armature slightly. If it moves freely this means that the system applied the air and the system configuration and the connections are correct.
ENDEVCO 2911 INSTRUCTION MANUAL SECTION 4 THEORY OF OPERATION THEORY OF OPERATION The shaker armature has a coil around it inside of a permanent magnetic field. When current is passed through the coil an opposite or attractive magnetic force field is generated depending on the phase of the drive signal. The interaction of the magnetic fields is what causes the armature to move up and down.
IM68357 document supplied with the system. Other means of calibration may not be traceable to the National Institute of Standards & Technology (NIST) and are not recommended. If you lack the equipment and/or the expertise to calibrate this instrument, return it to ENDEVCO for calibration. Calibration is recommended every 6 to 12 months.
ENDEVCO warrants each new instrument to be free from defects in material and workmanship for one year from the date of sale to the original purchaser. This warranty does not extend to units that have been mistreated, used in violation of Endevco recommendations, or to units that have been altered or repaired outside of the ENDEVCO factory.
Adapter Bushing, 1/4-28 UNF to 2-56 UNC EHM374 Hex Wrench, 1/8 in. EHM1020 Hex Wrench, 1/4 in. ACCESSORIES OPTIONAL The following accessories for use with the Model 2911 are available from ENDEVCO: QTY. PART NUMBER DESCRIPTION 34699 Isolation Mounting Kit *...
ENDEVCO 2911 INSTRUCTION MANUAL SECTION 7 DRAWINGS AND DIAGRAMS GENERAL CONNECTOR PINOUT TABLE 1 Shaker Control, DB25 Plug Description (NU = not used) Description (NU = not used) Armature ground Armature ground PA Out+ PA Out+ PA Out+ PA Out+...