Hach SensION156 Quick Reference Instructions
Hach SensION156 Quick Reference Instructions

Hach SensION156 Quick Reference Instructions

Multiparameter meters

Advertisement

54650-23
and
Multiparameter Meters

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the SensION156 and is the answer not in the manual?

Questions and answers

Summary of Contents for Hach SensION156

  • Page 1 54650-23 Multiparameter Meters...
  • Page 2 pH Menu Options Table 1 How to Change pH Menu Options Setup How to Get There Setup 1 From any reading mode, press SETUP Time Press ENTER Use the number keys to change the time. Press to accept the time. ENTER Setup 2 From any reading mode, press...
  • Page 3 pH Calibration Using Automatically Recognized Buffers 1. Prepare two or three pH buffers according to the electrode Note: Use a 6.86 or 7.0 pH buffer for the instruction manual. Choose from 1.68, 4.01, 7.00 (or 6.86), 10.01, mid-range buffer. and 12.45 pH buffers. Choose the pH setup 2.
  • Page 4 Conductivity Menu Options Table 2 How to Change Conductivity Menu Options Setup How to Get There Setup 5 From conductivity reading mode, press SETUP Display Lock Press the until the Setup number is 5. UP ARROW Press to toggle the display lock off and on. ENTER When the desired option is selected, press EXIT to return to the reading mode.
  • Page 5 Conductivity Calibration with a Known Standard 1. Make sure the meter is in Conductivity reading mode. 2. Make sure that the reference temperature in Conductivity Setup 9 matches the reference temperature of the sample. 3. Place the probe in a conductivity standard that is in the expected range of the samples.
  • Page 6 Dissolved Oxygen Menu Options Table 3 How to Change Dissolved Oxygen Menu Options Setup How to Get There Setup 5 From dissolved oxygen reading mode, press SETUP Display Lock Press the until the Setup number is 5. UP ARROW Press to toggle display lock off and on.
  • Page 7 Dissolved Oxygen Calibration in Water Saturated Air 1. Plug in the DO probe at least 30 minutes before calibrating to polarize the probe. 2. Secure the calibration and storage chamber to the probe cable. 3. Prepare the calibration and storage chamber by holding it under water and squeezing it a couple of times to pull a small amount of water into the lower chamber through the inlet.
  • Page 8 In the U.S.A. - Call toll-free 800-227-4224 Outside the U.S.A. - Contact the HACH office or distributor serving you. On the Worldwide Web - www.hach.com ; E-mail - techhelp@hach.com © Hach Company, 2000. All rights reserved. Printed in the U.S.A. ap/dk 1/00 1ed...