HOLDPEAK 700
Instruction Manual
Warnings!
It is important that users of this tester
read,understand,and follow all
warnings,cautions,safety information,and
instructions in this manual before operating or
servicing this tester.Failure to follow instructions
could result in death or serious injury.
Risk of electric shock and burn.Contact with live
circuits could result in death or serious injury.
Use caution with voltages above 30 V AC as a
shock hazard may exist.
A steady red glow and continuous beep
indicates voltage present.If no indication,voltage
could still be present.
In
bright
light
indicators will be less visible.
Do not use if green LED is not illuminated.
Do not use if tester appears damaged or if the
tester is not operating properly.If in doubt,replace the
tester.
Do not apply more than the rated voltage as
marked on the tester(1000 volts AC).
Always wear approved eye protection.
Before each use,verify operation by testing a
known working circuit that is within the rating of this
tester.
Do not rely on this tester with shielded
wire.Operation may be affected by differences in
socket design and insulation thickness and type.
Comply
with
requirements.
If this tester is used in a manner not specified by
the manufacturer,protection provided by the tester
conditions,the
LED
local
and
national
may be affected.
CAUTION
Do not attempt to repair this tester.It contains no
serviceable parts.
Do not expose the product to extremes in
temperature or high humidity.
Technical Data
Voltage range: 50...1000V AC
Frequency range: 50Hz...500Hz
Display: Clear, omnidirectional flashing beacon
IP-Protection: IP40
Overvoltage categorty : CAT III,600V
Pollution degree: 2
Altitude: up to 2000m
Temperature range: 0℃ up to 40℃
Humidity: <80%
Power supply: 2x1,5V batt. IEC Lr03
Dimensions: 145x25x21 mm
Weight: Approx.42g
Display and Control Elements:
1) Test probe
2) LED for voltage indication
3) Battery case cover and Clip
4) Switch
visual
SYMBOLS ON TESTER
Warning.Risk of electric shock.
Rish
important
read,understand,and
information,and instruction in this manual before
operating or servicing this tester.Failure to follow
instructions could result in death or serious injury.
Double insulated
OPERATING INSTRUCTIONS:
safety
Turn unit on:
Press the power button on the end-cap for 1
second.Listen for a single-beep sound and watch for
a steady green LED to illuminate in the tip of the
②
①
of
danger.Important
that
users
of
follow
all
④
③
information.It
is
this
tester
warning,safety