Contents Before First Use ..............................3 Positioning and safely using your MATOS Aria Medical refrigerator ..............3 Transporting and moving your MATOS Aria Medical refrigerator ................. 3 Energy saving tips ..............................4 Product Information ............................ 5 Important Information ............................5 Product Specifications ............................
40°. In the event that the cabinet is tilted more than 40°, the power supply must not be connected until the appliance has stood upright for at least 24 hours. Rollex Medical – MATOS ARIA Eco 150L User Manual v.02 24/04/2018...
Ensure that products being stored in the appliance are below room temperature upon entry. • Make sure the door is opened as little as possible whilst in use and closed as quickly as possible to prevent unnecessary temperature fluctuations. Rollex Medical – MATOS ARIA Eco 150L User Manual v.02 24/04/2018...
This refrigerator has been designed and built with the intention of the product being used within a Pharmacy and is designed to store contents between +2°C to +8°C. Please see below for specific product information: Rollex Medical – MATOS ARIA Eco 150L User Manual v.02 24/04/2018...
There are also 2 roller castors at the rear of the appliance to ease product movement. Additional or replacement parts are available from our spare parts department. See back of booklet for contact details. Rollex Medical – MATOS ARIA Eco 150L User Manual v.02 24/04/2018...
Controller overview and operation Menu Lock keypad Light (not available on all models) Maximum temperature / Up Power Mute Minimum temperature / Down Set / Enter Rollex Medical – MATOS ARIA Eco 150L User Manual v.02 24/04/2018...
The set point is the point at which the refrigerator normally holds the temperature at; in some circumstances such as high/low room temperatures it may be necessary to adjust this slightly. Changing the temperature scale between Celsius (°C) and Fahrenheit (°F) Rollex Medical – MATOS ARIA Eco 150L User Manual v.02 24/04/2018...
(e.g. a 5ml vaccine). The temperature of a refrigerator’s contents will change more slowly (up or down) than the air temperature within that refrigerator should there be a fault or the door is left open. Rollex Medical – MATOS ARIA Eco 150L User Manual v.02 24/04/2018...
Remote alarm contacts All MATOS Aria Medical Pharmacy Refrigerators include remote alarm contacts and these terminal contacts enable the user to connect to a remote control station; these contacts can be found on the rear of the cabinet. WARNING: These contacts must only be accessed by a competent electrician.
Remove shelves/basket and wipe with a soft cloth. DO NOT put in a dishwasher. • Detergents containing abrasives or acids are not suitable for the cleaning. • When all cleaning has been completed and accessories replaced, reconnect the plug with dry hands. Rollex Medical – MATOS ARIA Eco 150L User Manual v.02 24/04/2018...
Install the hinge on the top, left hand side (figure 5). Figure 4 Figure 5 Remove the bottom, left hand feet and base plate (figures 6 and 7). Figure 6 Figure 7 Rollex Medical – MATOS ARIA Eco 150L User Manual v.02 24/04/2018...
Page 14
Remove the caps from the screws holding the handle in position. Remove the handle screws. Remove the caps on the opposite side of the door. Screw the handle into position and add the caps to cover the screws. Rollex Medical – MATOS ARIA Eco 150L User Manual v.02 24/04/2018...
3. Empty, defrost and clean the refrigerator. 4. Make sure the door is left slightly ajar to prevent any bad odours forming within the cabinet. Rollex Medical – MATOS ARIA Eco 150L User Manual v.02 24/04/2018...
Until the moment of cooling, the alarm is repeated every 15 minutes. Such an operating mode is normal until the required temperature settings are reached. Green & Yellow 13 Amp Fuse Brown Blue Cord clamp Rollex Medical – MATOS ARIA Eco 150L User Manual v.02 24/04/2018...
A relevant label has been placed on the refrigerators packaging (see below). The product has been manufactured of recyclable materials. Ask your local environmental care authority for information concerning location of such facilities. Troubleshooting Guide Rollex Medical – MATOS ARIA Eco 150L User Manual v.02 24/04/2018...
Alternative Manual Location A digital copy of this manual will be available for free download in the following locations: www.rollexmedical.co.nz/support/ www.rollexmedical.com.au/support/ Rollex Medical – MATOS ARIA Eco 150L User Manual v.02 24/04/2018...
Australia: Sales@rollexmedical.com.au 1300 880 441 Spares and After Sales Service: New Zealand: service@rollexmedical.co.nz 09 274 0707 Australia: service@rollexmedical.com.au 1300 880 441 For any other information please visit: www.rollexmedical.co.nz/support/ www.rollexmedical.com.au/support/ Rollex Medical – MATOS ARIA Eco 150L User Manual v.02 24/04/2018...