Anton Sprint Pro 2 Quick Start Manuals

Advertisement

Quick Links

Summary of screen icons
Select, pick or action
Up item or entry
Down item or entry
Left field
Right field
Zero pressure
Start timed test
Stop timed test
Restart timed test
Yes, pass or done
No or fail or cancel
Purge or pump running icon
Select temp 1 (flow)
(Differential Temperature Test)
Select temp 2 (return)
(Differential Temperature Test)
Save log
Export report/log to IR Printer
Export report/log to Mobile
App
Delete
Edit value
Increase value
Decrease value
Descend through character
list
(string edit)
Ascend through character list
(string edit)
CO Alarm
Select Net/Gross/NetHE
Efficiency
IR Printer Alignment
Unit 6 Greenhill House, 26 Greenhill Crescent, Watford Business Park, Watford WD18 8JA
Tel. No: +44 (0)870 428 0075
Quick Start Guide - Sprint Pro 2 - 6
IMPORTANT NOTES:
Please read the user manual for full instructions on the use of Sprint Pro this guide should be used for a
quick reference only (full user manual is available at www.anton-group.com).
Observe all warnings and instructions detailed in the user manual or as prompted by the analyser.
Before use ensure Sprint Pro is in good repair and do not use if damaged or if calibration has expired.
Sprint Pro must only be used with the Sprint Pro Flue Probe Assembly and always with the water trap
and filters in place.
The water trap must be dried after use and before Sprint Pro Flue Probe Assembly is returned to carry
case.
Connect the Flue Probe Assembly to Sprint Pro before switching on and do not insert probe into flue
until after the zeroing process is complete.
Always start Sprint Pro in clean air.
Sprint Pro will lock out its Test Menu once calibration has expired.
SWITCH-ON
Press and hold the ON/OFF/ENTER key for two seconds until you hear two beeps. Ensure the gas exhaust
outlet on the side of the FGA is not blocked. After warm-up is complete, Sprint Pro will display the Main Menu
screen. You can now navigate the menu system using the
analyser.
SWITCH-OFF
Press and hold the ON/OFF/ENTER button for approximately two seconds until you hear a rising beep. Press the
ESC key to abort switch off sequence and return to Test Menu. Note: Unit will purge and shut down once
purge complete.
Charging battery
PERFORMING TESTS
To enter any test screen, select the required test from the menu and use the soft key marked
To end any test, press the ESC key.
FLUE GAS ANALYSIS
Select Flue gas analysis from the Test Menu. The first time this option is entered after each switch on the Sprint
Pro will perform a 'zero'. Following a successful 'zero', Sprint Pro will begin taking measurements. Use the soft
keys to log or print the results.
Press the ON/OFF/ENTER button to display the various screens available:
Screen 1: O
, CO, CO
, CO/CO
2
2
Screen 2: O
, excess (XS) air, Temperature Flue and Efficiency
2
Screen 3: Combines flue elements from previous screens.
Screen 4: (if NO sensor not fitted): Temperature Flue, Temperature Inlet, Net Temperature
Screen 4: (if NO sensor fitted): NO, NOx, CO, CO
Screen 5: (if NO sensor fitted): Temperature Flue, Temperature Inlet, Net Temperature
To perform a manual pressure 'zero', use the soft key marked
To change the efficiency readings
gas screen 2.
Note: The pump will remain on and continue purging for some time after leaving this option to clear residual gas
PRESSURE AND DIFFERENTIAL PRESSURE TESTS
Before carrying out the test, the pressure must be zeroed with the tubes connected to the instrument but not to
the pressure source. A thermocouple may be connected to provide temperature readings in these tests.
1. Select Pressure Menu from the Test Menu. Then select Pressure or Diff Pressure from the menu.
2. Connect tubes to pressure inlets on instrument.
3. Zero pressure using the soft key marked
4. Connect tube(s) to pressure source(s) to begin differential pressure tests. Sprint Pro will give a warning if the
pressure points are the wrong way round.
5. There is an option to time this test: Sprint Pro will display the duration time on the screen.
Press the soft key
to start optional timer.
To stop timer press the soft key
To reset the timer press the soft key
6. To re-zero the pressure at any time press
LET-BY, STABILISATION AND TIGHTNESS TEST
1.
Select Pressure Menu from the Test Menu. From the Pressure Menu select Let-by / Tightness.
2.
Connect tube to positive pressure inlet on the instrument.
3.
Zero pressure, using the soft key marked
Anton Industrial Services Limited
Website: www.anton-group.com
Part number: INS29504 Issue 1 September 2018
and
ratio and pressure
2
, CO/CO
Ratio, O
2
2
in flue gas screen 1 or 3.
(Net, Gross – for NetHE see user manual),
.
then .
Email
Page 1 of 2
soft keys and the
and ESC keys on the
or
.
2
use the soft key marked η in flue
:
sales@anton-group.com

Advertisement

loading

Summary of Contents for Anton Sprint Pro 2

  • Page 1 Quick Start Guide - Sprint Pro 2 - 6 IMPORTANT NOTES: Please read the user manual for full instructions on the use of Sprint Pro this guide should be used for a quick reference only (full user manual is available at www.anton-group.com).
  • Page 2 6. Sprint Pro will emit continuous clicks like a Geiger counter. If higher gas levels are detected the bar graph will increase in readings and the sounder will increase in pitch. Anton Industrial Services Limited Unit 6 Greenhill House, 26 Greenhill Crescent, Watford Business Park, Watford WD18 8JA Tel.

This manual is also suitable for:

Sprint pro 3Sprint pro 4Sprint pro 6Sprint pro 5