INSTRUCTIONS FOR USE
Slit Lamp
BQ 900
®
18. Edition / 2016 – 02
1
© HAAG-STREIT AG, 3098 Koeniz, Switzerland - HS-Doc. no. 1500.7220589-04170 – 2016 – 02

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the Slit lamp BQ 900 and is the answer not in the manual?

Questions and answers

Summary of Contents for Haag-Streit Slit lamp BQ 900

  • Page 1 SVENSKA NEDERLANDS PORTUGUÊS ESPAÑOL ITALIANO FRANÇAIS DEUTSCH ENGLISH INSTRUCTIONS FOR USE Slit Lamp BQ 900 ® 18. Edition / 2016 – 02 © HAAG-STREIT AG, 3098 Koeniz, Switzerland - HS-Doc. no. 1500.7220589-04170 – 2016 – 02...
  • Page 2 This device is equipped with high intensity light emitting diodes. Exces- sive exposure of patients in treatment with certain medication may lead to phototoxic adverse reactions, due to higher photosensitivity. © HAAG-STREIT AG, 3098 Koeniz, Switzerland - HS-Doc. no. 1500.7220589-04170 – 2016 – 02...
  • Page 3: Table Of Contents

    Fixation star ......................9 Immunity for non-life support devices (standard table 4) ........15 Microscope and eyepiece ..................9 Safe distances for non-life support devices (standard table 6) ......16 © HAAG-STREIT AG, 3098 Koeniz, Switzerland - HS-Doc. no. 1500.7220589-04170 – 2016 – 02...
  • Page 4: Safety

    "Purpose of use". • Check the appliance for damage after it is unpacked. Return defective appliances in the appropriate packaging. © HAAG-STREIT AG, 3098 Koeniz, Switzerland - HS-Doc. no. 1500.7220589-04170 – 2016 – 02...
  • Page 5: Light Toxicity

    Slit lamp illumination 1.8 Warranty and product liability power supply • Haag-Streit products must be used only for the purposes and in the manner des- Test symbol of CSA with ap- Background illumination cribed in the documents distributed with the product.
  • Page 6: Overview

    28. Mounting screw for breath shield 3.1 Microscope and illumination • The slit lamp is packaged and delivered fully assembled. The transport lock must be removed prior to commissioning. © HAAG-STREIT AG, 3098 Koeniz, Switzerland - HS-Doc. no. 1500.7220589-04170 – 2016 – 02...
  • Page 7: Power Supply

    (out) makes it easier. It should at least be set so hard that the slit cannot close on its own. A © HAAG-STREIT AG, 3098 Koeniz, Switzerland - HS-Doc. no. 1500.7220589-04170 – 2016 – 02...
  • Page 8: Preparing The Patient

    90° using the locking screw (23) and tilted in 5° steps using the latch (31), and the illumination and microscope turned to the central position (0°). • Front-lens glasses and contact glasses are used to examine the ocular fundus. © HAAG-STREIT AG, 3098 Koeniz, Switzerland - HS-Doc. no. 1500.7220589-04170 – 2016 – 02...
  • Page 9: Setting The Filters & Diaphragms

    Slit image width 0 – 14 mm continuous Slit image length 1 – 14 mm continuous Illumination field circle ø 8 / 5 / 3 / 2 / 1 / 0.2 mm Test mark With fixation star Slit image rotatability ± 90° © HAAG-STREIT AG, 3098 Koeniz, Switzerland - HS-Doc. no. 1500.7220589-04170 – 2016 – 02...
  • Page 10: Stereo Microscope

    Only original HS replacement parts may be used. Dust cover, large (for several instruments) HS no. 1001434 The LED illumination can be operated maintenance-free for its entire service life. © HAAG-STREIT AG, 3098 Koeniz, Switzerland - HS-Doc. no. 1500.7220589-04170 – 2016 – 02...
  • Page 11: Appendix

    Instrument table HSM 901 Imaging • The EN 60601-1 standard must be observed when using different medical and/or Instrument table HSM 901 Workstation non-medical electrical appliances in combination. © HAAG-STREIT AG, 3098 Koeniz, Switzerland - HS-Doc. no. 1500.7220589-04170 – 2016 – 02...
  • Page 12: Classification

    E. Standards EN 60601-1 EN 60601-1-2 EN ISO 15004-1 EN ISO 15004-2 EN ISO 10939 © HAAG-STREIT AG, 3098 Koeniz, Switzerland - HS-Doc. no. 1500.7220589-04170 – 2016 – 02...
  • Page 13: Information And Manufacturer's Declaration Concerning Electromagnetic Compatibility (Emc)

    Class A Emission of harmonics according to domestic purposes. EN 61000-3-2 Voltage fluctuations / flicker emissions Fulfilled according to EN 61000-3-3 © HAAG-STREIT AG, 3098 Koeniz, Switzerland - HS-Doc. no. 1500.7220589-04170 – 2016 – 02...
  • Page 14: Immunity (Standard Table 2)

    NOTE: U = the AC mains voltage prior to application of the test level. © HAAG-STREIT AG, 3098 Koeniz, Switzerland - HS-Doc. no. 1500.7220589-04170 – 2016 – 02...
  • Page 15: Immunity For Non-Life Support Devices (Standard Table 4)

    Over the frequency range 150 kHz to 80 MHz, fi eld strengths should be less than 3 V/m. Possible shorter distances outside the ISM bands do not contribute to improved application in this table. © HAAG-STREIT AG, 3098 Koeniz, Switzerland - HS-Doc. no. 1500.7220589-04170 – 2016 – 02...
  • Page 16: Safe Distances For Non-Life Support Devices (Standard Table 6)

    NOTE 3: These guidelines may not apply in all situations. Electromagnetic wave propagation is influenced by absorption and reflection of buildings, objects and people. © HAAG-STREIT AG, 3098 Koeniz, Switzerland - HS-Doc. no. 1500.7220589-04170 – 2016 – 02...
  • Page 17 SVENSKA NEDERLANDS PORTUGUÊS ESPAÑOL ITALIANO FRANÇAIS DEUTSCH ENGLISH © HAAG-STREIT AG, 3098 Koeniz, Switzerland - HS-Doc. no. 1500.7220589-04170 – 2016 – 02...
  • Page 18 Should you have any further questions, please contact your HAAG-STREIT dealer at: http://www.haag-streit.com/contact/contact-your-distributor.html HAAG-STREIT AG Gartenstadtstrasse 10 3098 Koeniz, Switzerland Phone +41 31 978 01 11 +41 31 978 02 82 eMail info@haag-streit.com Internet www.haag-streit.com © HAAG-STREIT AG, 3098 Koeniz, Switzerland - HS-Doc. no. 1500.7220589-04170 – 2016 – 02...