Gilson HM-378 Operating Manual page 6

Binder ignition system
Table of Contents

Advertisement

Gilson Company, Inc.
Binder Ignition Programming Sequence:
Furnace must be plugged in and the toggle switch
must be in the "ON" position.
All programming starts from the IdLE display. Press
<ANY> key except <STOP> to obtain the IdLE display.
Insure the °C indicator in the display is lit. If not, press
<ENTER>, then <°F/°C>. "CHG" will display. Press
<ENTER> again and °C will be lit.
Displays noted with "/" alternate between the two
displays shown.
Program #1 may be confirmed, reprogrammed, or
altered by the following steps:
Action
Starting display.
Press <PROGRAM>
Press <ENTER>. Press <ENTER>
again to accept displayed value, or
input new main chamber preheat tem-
perature, then press <ENTER>.
Press <ENTER>. Press <ENTER>
again to accept displayed value, or
input new main chamber burnout tem-
perature, then press <ENTER>.
Press <ENTER>. Press <ENTER>
again to accept displayed value,
or input new hold time, then press
<ENTER>.
Press <ENTER>. Press <ENTER>
again to accept displayed value, or in-
put new Afterburner setpoint tempera-
ture, then press <ENTER>.
Press <ENTER>. Press <ENTER>
again to accept displayed value, or
input new exhaust fan activation tem-
perature, then press <ENTER>.
Ending display.
After completing the Programming sequence as
shown, the furnace is ready to run using Program
#1.
The fan-on temperature (CFN) is set at or below the
Afterburner temperature setting (CAb) to allow faster
preheating. DO NOT attempt to test samples if the
fan is not on.
To review the program last used or just programmed,
start from IdLE, press <ENTER>, then press <PRO-
GRAM REVIEW>. Programmed values will display
in sequence.
Display
IdLE/XXX
-
Current
main
chamber temperature.
Pro/1 (for Program 1)
CPR/500 - Indicates 500°C is
set as main chamber preheat
temperature.
Cbo/538 - Indicates 538°C is
set as main chamber burnout
temperature.
HOLd/00.15 - Indicates 15
minutes hold time for decok-
ing.
CAb/850 - Indicates 850°C
is setpoint for Afterburner
chamber.
CFN/850 - Indicates exhaust
fan will activate at 850°C After-
burner temperature.
IdLE - Programming sequence
is complete, ready for contin-
ued operation.
Page 6
Gilson Binder Ignition System: HM-378
To confirm the active program during firing, press
<PROGRAM REVIEW> at any time after -ON- has
been displayed (as described below).
To recall a program from memory or review, proceed
from IdLE as in the above sequence until the desired
program number is displayed (PRO/1, etc.). Press
<ENTER>, then press <STOP>. The desired program
is then ready for review as described above.
Delayed Firing:
In cases where the furnace must be ready for testing at
the beginning of a shift, it can be programmed to begin
the preheat cycle at a delayed time. Follow this procedure
to initiate a Delayed Firing:
Action
Starting display.
Press <DELAY>.
Input desired delay time, for example:
12 hrs, 30 min.
Press <ENTER>.
Press
<START>.
Initiates
Countdown.
Press <STOP> at any time to cancel the delayed
firing sequence or if a programming error is made.
The display will return to IdLE.
Delayed Firing programming must be repeated for
each use. Program is not retained in memory.
Preheating the Furnace:
Action
Starting display.
Press <ENTER>.
No action required.
No action required.
Red lights on the front of the furnace will be lit when
main chamber and after-burner heating elements are
active.
Display
IdLE
DELA/XX.XX
-
Indicates
hh.mm until furnace starts
pre-heating. Maximum time is
99hr, 99min.
DELA/12.30
IdLE - Programming is com-
plete.
DELA/XX.XX - Counts down
Delay
time remaining. Furnace will
begin preheating in 12 hrs,
30 min.
Display
IdLE/XXXX - Main chamber
temperature is indicated.
-ON- - Controller runs a self-
diagnostic routine for 5—10
seconds.
PRHT/XXXX - Main chamber
temperature.
dELA/XX.XX - Time remaining
if delayed firing is active.

Hide quick links: