INDEX 02 PRESENTATION OF MANUAL 04 IDENTIFICATION OF EQUIPMENT 04 Indication of Equipment 04 Purpose of the equipment 05 Principles and fundamentals applied to the product functioning 05 Description of Equipment 06 MODULES, ACCESSORIES, OPTIONS AND MATERIALS OF CONSUMPTION 09 TECHNICAL SPECIFICATIONS 09 General features 11 Electromagnetic emissions 15 Dimensions...
IDENTIFICATION OF EQUIPMENT Dear Customer Congratulations. You have made a good choice when you decided to buy a GNATUS QUALITY product comparable to the best products available in the World. This manual is a general presentation of your product and it will give you important details to help you to solve possible problems.
To guarantee the safe functioning of your equipment, use only the assemble configurations (Dental Chair, Dental and Water Units and Dental Light) supplied by Gnatus authorized Dealers / Technical Assistance.
MODULES, ACCESSORIES, OPTIONS AND MATERIALS OF CON- SUMPTION The contents of this page are of an informative nature, the equipment being able to differ from that illustrated. So, upon acquiring the product check the technical compatibilty between equipment, coupling and accessories. WARNING For the chair with integrated connection box is recommended to fix the base of the chair to the floor, as oriented under the item "Installation"...
Page 7
MODULES, ACCESSORIES, OPTIONS AND MATERIALS OF CON- SUMPTION...
Page 8
MODULES, ACCESSORIES, OPTIONS AND MATERIALS OF CON- SUMPTION 01 - Bi articulated headrest ("click" system). 02 - Bi articulated headrest (driven by knob). 03 - Bi articulated headrest (driven lever). 04 - "Wireless system" pedal 05 - Spare bounding box kit. 06 - Anti-smashing device.
TECHNICAL SPECIFICATIONS General features Models Syncrus G8 Supply voltage 127/220 V~ (selectable internally) Frequency 50/60 Hz Electric shock protection type Class one equipment - Applied Part Type B Mode of Operation Continuous with intermittent load: T-on 1min - T-off 4min...
Page 10
List of pieces and circuit scheme Gnatus Company declares that the supply of the circuit scheme, list of pieces or any other information that propitiate technical attendance for the user, can be request if there is an agreement between the user and Gnatus Company.
TECHNICAL SPECIFICATIONS Electromagnetic emissions Electromagnetic Emissions Syncrus G8 Dental Chair is made to be used in the electromagnetic environments specified below. The client or the user of the Syncrus G8 Dental Chair must be sure that it is used in such environment.
Page 12
TECHNICAL SPECIFICATIONS Guidelines and manufacturer's declaration - electromagnetic immunity Syncrus G8 Dental Chair is made to be used in the electromagnetic environments specified below. The client or the user of the Syncrus G8 Dental Chair must be sure that it is used in such environment.
Page 13
TECHNICAL SPECIFICATIONS Guidelines and manufacturer's declaration - electromagnetic immunity Syncrus G8 Dental Chair is made to be used in the electromagnetic environments specified below. The client or the user of the Syncrus G8 Dental Chair must be sure that it is used in such environment.
Page 14
TECHNICAL SPECIFICATIONS Recommended distances between portable and mobile RF communication equipments and the Syncrus G8 Dental Chair Syncrus G8 Dental Chair is made to be used in an electromagnetic environment in which Syncrus G8 Dental Chair RF disturbances are controlled. The client or the user of the...
TECHNICAL SPECIFICATIONS Packing symbols It determines the maximum Packing to be transported and / quantity of boxes which can be or stored avoiding humidity, rains stacked during transportation and and wet floor. storage “as per packaging”. The packing must be stored and Packing to be transported and / transported away from direct sun or stored with the harrows up.
Page 17
TECHNICAL SPECIFICATIONS Product symbols General warning: If the instructions are not properly General sign for mandatory action. followed, the use can lead to damage to the product, user or patient. Turned on position. Turned off position. Lift seat. Lower seat. Lift backrest.
Intensity - Massager. INFRA-STRUCTURE SPECIFICATION The perfect functioning and the durability of the Gnatus dental set are directly linked to the preinstallation, which must be carried out by duly trained professionals according to the instructions contained in the "Pre installation plan" which is found with the Gnatus authorized technical representative.
Page 19
INSTALACIÓN DEL EQUIPAMIENTO Attention The Gnatus dental set is dimensioned so that the stability of the set composed of a Dental Chair, Delivery Unit, Dental Light and Water Unit remains stable in normal conditions of use. If, by option of the customer, it is necessary to install other accessories in the dental set.
OPERATION OF EQUIPMENT Initiating After positioning the chair in the chosen place: - Turn on the general key - on/off switch located on the engine cover trim side (a). - Thru the command pedal, adjust the chair to the desired position. - Adjust the headrest position.
Page 21
To remove it, just pull. To place it again, find the correct position and push, fixing the upholstery to the backrest. Bump preventing device The Syncrus G8 chair features a leg bump preventing device (e). When the chair was descending, any obstruction below the chair would trigger this function, interrupting all movements and preventing further complications.
Page 22
OPERATION OF EQUIPMENT Operation of Integrated Foot control “Chip-Blower” 01 - Operating Position 1/2 02 - Operating Position 3/4 03 - Return to the last position / Spitting Position 04 - Activation of the dental light / inverting the rotation of the electric micromotor 05 - Return to Zero 06 - Lever for electro-pneumatic operation 07 - Chip-Blower System (double function operation key - “Spray”...
Page 23
Working Positions: - The Syncrus G8 Dental Chair has four working positions. In order to program them, just place the chair in the position and the dental light at the desired intensity and press and hold for 3 seconds the “working position 1 and 2”...
Page 24
OPERATION OF EQUIPMENT “Go back to the last position” and “Spitting Position”: When pressing the button “Go back to the last position” “Spitting Position” (03), the light source will turn off (if it was on), water will then flow in the basin (for the duration of the programmed time or if this has not been programmed, for 4 minutes), the chair back will raise completely to the Spitting Position, when pressing it again, the back support will return to the previous position and the light source will turn on (if it was on).
Page 25
OPERATION OF EQUIPMENT Backrest / seat simultaneous movement - Trendelemburg position When lowering of rising backrest, the seat automatically moves, improving the patient’s comfort. At the backrest lowest position the pa- tient can lie down in the Trendelemburg position - legs and pelvis (01) are loca- ted in a higher position than thorax and head (02).
Page 26
OPERATION OF EQUIPMENT Massager control To activate the massager function, press in the equipment control panel (On/Off) massager (B). After selecting the function, choose the operation option: 01 - Numeric display – speed and intensity percentage (0 to 99%). Control panel of the 02 - Speed or intensity control.
Page 27
OPERATION OF EQUIPMENT Massager Control Functions To turn the massager on, press the key (06) and to turn it off, keep it pressed for one second. With the massager already on, press the key (06) to select how much time it will remain on, 10 or 20 minutes, after that time the massager will automatically turn off.
PRECAUTIONS, RESTRICTIONS AND WARNINGS Recommendations for the dental equipment maintenance Your Gnatus equipment has been designed and developed according to the standards of modern techology. Similarly to other kinds of equipment, it requires special care, which is many times neglected due to several reasons and circunstances.
- It is not convenient that the Syncrus G8 Dental Chair be used near to or Piled over other equipment, in case of the use near to or piled over is necessary, it is convenient that the Syncrus G8 Dental Chair be observed to check if it is working properly in the configuration to be used.
- The equipment should only be operated by duly enabled and trained technicians (Dental Surgeons, Capacitated Professionals). - If any maintenance should be required, only use services of the Gnatus Authorized Technical Assistance. - The equipment has been manufactured to handle both continuous and intermittent operation; so follow the cycles described in these Instructions for Use.
(failures, causes and solutions). If it is not possible to resolve the problem, turn off the equipment, remove the power supply cable from the socket and contact your representative (Gnatus). Precautions to be adopted against foreseeable or uncommon risks, related to...
For protecting your equipment, look for a Gnatus’ technical assistance for periodic reviews as preventive maintenances. Corrective Maintenance Gnatus states that the supplying of the circuits’ diagram, Part lists or any other information that permits the technical assistance by the user, can be requested, since previously agreed between the buyer and Gnatus.
Among the care you have to take with your equipment, the most important is regarding of the spare parts replacement. To ensure the lifetime of your device, only replace original spare parts from Gnatus. They have the assurance of the standards and technical specifications required by the Gnatus representative.