Use this guide to help change the battery for a DataTrace MPIII data logger. This quick
start guide is intended to use the tools provided in the MPIII Maintenance Kit (part num-
ber 124808-500). Replacement batteries part number: 124804-500 (pkg of 20)
1
Clean and dry the data logger, making sure
that the surfaces are clear of contaminants or
debris. Then, loosen the battery cap with the
logger gripper provided by holding the logger
in one hand with the probe facing toward the
ground, and twisting the battery cap counter-
clockwise with the other hand using the log-
ger cap wrench provided. (PN DT170002)
4
Examine the battery compartment and
threads, making sure that the wave springs
located in the cap and inside the battery
compartment are in place. If one of the
springs has fallen out, reinsert it.
Mesa Laboratories, Inc.
12100 W. 6th Avenue
Lakewood, CO 80228 USA
MPIII Battery Change Quick Start Guide
PN DT170002
2
Lift off the battery cap and set it to the
side, taking care not to misplace the springs
inside the logger or cap.
3
Remove the old battery by tipping the
logger to its side until the battery falls out.
Dispose of the battery appropriately per local
regulations.
5
Remove and discard the O-ring to be
replaced using the pick tool provided. Clean
the O-ring groove with a clean, lint-free cloth.
Check for any scratches or nicks in the metal
before reassembly. If damage is detected or
contamination is noted, return the logger to
an authorized servicer for repair.
datatracesales@mesalabs.com
Tel: +1-303-987-8000
datatrace.mesalabs.com
Rev. 05OCT2016

Advertisement

loading

Summary of Contents for DataTrace MPIII

  • Page 1 MPIII Battery Change Quick Start Guide Use this guide to help change the battery for a DataTrace MPIII data logger. This quick start guide is intended to use the tools provided in the MPIII Maintenance Kit (part num- ber 124808-500). Replacement batteries part number: 124804-500 (pkg of 20)
  • Page 2 Apply silicone grease to a new O-ring Holding the logger in the same posi- by rubbing a small amount on your index tion as step 1, place a new battery in the finger and thumb and slowly pulling the battery compartment of the logger, taking O-ring through the grease on your fingers care to put the “+”...