Page 5
User manual TUMBLER 1000+ Specification plate Main body Switch box Stand Air vent grid Drive vent grid Power plug incl. fuse Plate for electronic sensor (inside switch box)
User manual TUMBLER 1000+ 2 Warnings Please carefully read these warnings before operating this apparatus! Technical staff which is not familiar with this user manual is not allowed to operate the apparatus! Ensure that this user manual is always available at the operation location! Children and people below the age of 16 are not allowed to operate the apparatus and should be kept away from the connected apparatus! The working area has to be secured in an appropriate way and must meet the...
User manual TUMBLER 1000+ 3 Unpacking / Installation / First operation 3.1 Unpacking Before using the apparatus for the first time check the delivery for damages and completeness on the following components: Main body Tumbler-Box Dust bin ...
User manual TUMBLER 1000+ 4 Operating instructions Connect the power plug to a suitable electricity supply and switch on the main switch („On“). The tumbler starts an adjusting sequence (display shows “Adjusting”) and stops automatically afterwards. The indicator light will be on and the display will show the amount of turns and time remaining before starting the test.
Page 9
User manual TUMBLER 1000+ The display will show the remaining turns and time. The tumbler will stop automatically after 500 rotations and 10 minutes. Then the display shows “done”. Do not open the apparatus door before the tumbler fully stopped. Unlock the locking latch of the tumblerbox and remove it to the front.
User manual TUMBLER 1000+ 4.1 Guidelines for cleaning Before any cleaning or servicing disconnect the power plug! The dust bin should be discharged periodically according to the frequency of usage. Therefore you should take out the tumblerbox first for easier handling. On the back side of the apparatus is a small blue lever on the air vent grid (see page 4 pos.
User manual TUMBLER 1000+ 5 Guidelines for testing The determination of the durability of pellets is accurately described in the standard ISO 17831-1. If you conduct a formal test according to the standard, follow all instructions precisely. 6 Principle The instrument is exactly designed according the requirements of the ISO 17831-1 chapter 5.1 and the controller (function principle) ensures the requirements of the ISO 17831-1 chapter 7.1.
User manual TUMBLER 1000+ 8 Fault finding Problem Possible cause Corrective action Tumbler will not function Power plug not connected Connect apparatus Main switch in “Off” position Switch main switch to “On” Apparatus door open Close door Emergency stop switch pushed Pull the emergency stop switch Fuse is damaged Change fuse...