Technical Specification - S.P.L. HF-400 User Manual

Electrosurgical analyzer for function tests of hf surgical equipment
Table of Contents

Advertisement

S.P.L.

2 Technical specification

Line voltage:
Power consumption:
Safety class:
ambient
temperature:
Storage temperature:
Measuring ranges
HF current :
HF-Peak Current (V2):
Discrimination:
or HF output power:
Crest Faktor (V2)
HF leakage current:
Discrimination:
Neutral electrode test:
Range:
Measuring principle:
Load resistors:
Swing in time:
Rated voltage:
Output power:
83 V – 264 V ac, 50 / 60 Hz
50 VA
1
+ 5 - +40 °C
-10 - +50 °C
0 - 5000 mA
0 - 5000 mA
0,1 mA
0 - 500 Watt
1 - 10 (at > 1000 mA)
0 - 250 mA
0,1 mA
0 - 1000 Ohm
0,3 - 10 MHz
Thermal electric converter
10 Ohm,
25 Ohm - 6375 Ohm
in steps of 25 Ohm
< 3 sec.
max. 1000 V
500 watt 1 min on, 5 min off
Permanent max. 200 watt at 25°C
environmental temperature (50 – 800 Ohm)
HF-400 User manual
5