Do you have a question about the 8000D MIXER/MILL and is the answer not in the manual?
Questions and answers
JOSHI PRADNYESH VINAYAK
April 2, 2025
safety switch to be attended
1 comments:
Mr. Anderson
April 2, 2025
For the SPEX CertiPrep 8000D MIXER/MILL, the following safety switch issues should be addressed:
1. Lid Open: The unit will not run if the lid is open. Ensure the lid is closed and latched properly.
2. Safety Interlock Switch: If the unit does not run even though the lid is closed and the timer counts down, the safety interlock switch may need adjustment.
3. Time-Delay Latch: This latch prevents the lid from opening during and immediately after operation. Wait for the audible click before opening the lid. Never force the lid open before the Time-Delay Latch unlocks.
This answer is automatically generated
Subscribe to Our Youtube Channel
Summary of Contents for SPEX CertiPrep 8000D MIXER/MILL
G-forces, enough to pulverize the toughest rocks, slag and ceramics. The dual clamps of the 8000D Mixer/Mill not only allow twice as many samples to be ground in a given time, but they also move in balance, reducing vibration and extending the life of the mill’s components.
8000D Mixer/Mill 2. SPECIFICATIONS Type of mill: High-Energy shaker mill Grinding container: Canister with balls Weight (empty, without vial): 68 lbs. (31 Kg) Dimensions: 22 in (56 cm) x 15 1/4 in (39 cm) x 11 in (28 cm) Grinding Vials Available (see pages 16-19): 8001 Hardened Steel Vial Set.
8000D Mixer/Mill 3. UNPACKING Any visible damage to the shipping container should immediately be reported to the carrier. If there is no visible damage, remove all packing documents from the exterior of the box, and after completion of your inspection, file with your records.
8000D Mixer/Mill 4. SETTING UP Remove the plastic bag containing the instructions and spare fuses. Then loosen the locking tab on the clamps, and unscrew the clamps a few turns. Cut and remove the tie holding the clamps together. Tilt the cabinet about 45 degrees, and note the four bolts on the underside of the cabinet. They hold the floating base plate to the cabinet for shipping.
Page 7
8000D Mixer/Mill 4.3 CE Safety Features CE directives require that electricity to the motor and timer be interrupted whenever the lid is open, and that the clamp arm must have stopped completely before the lid can open. Safety Interlock - The Safety Interlock cuts off power when the lid is opened; the switch is under the lid on the upper right-hand side.
8000D Mixer/Mill 5. OPERATION Plug the power cord into the mill’s electrical inlet, and then into an approved grounded outlet. Ensure that the On/Off switch (next to the power inlet) is in the off position, before plugging the unit in. After the unit is plugged in, switch the power on by means of the rocker switch.
Page 9
8000D Mixer/Mill 5. OPERATION (Continued) A typical mixing load is up to 25 ml, or about 40% of the vial’s volume. If a sample is not being ground fine enough, one can decrease the amount of sample, increase grinding time, use a denser grinding medium, and/or add a grinding aid or a liquid. Wet grinding...
This should minimize the buildup of any powders, mold/mildew, or unsightly gunk. To maintain the exterior of the unit, first disconnect the 8000D Mixer/Mill, then spray it with a mild window cleaner or similar product, and wipe it down. REMINDER: unplug the Mixer/Mill before any cleanup or maintenance work.
8000D Mixer/Mill 7.0 Accessories: Grinding/Mixing Vials (If not included in your order, they are available for purchase at additional cost) 8001 Hardened Steel Vial Set Vial size 2 ¼ n. dia. x 3 in. Vial body and cap liner of hardened tool steel.
Page 12
8000D Mixer/Mill 7.0 Accessories: Grinding/Mixing Vials (If not included in your Order, they are Available for Purchase at Additional Cost) 8005 Zirconia Ceramic Vial Set Vial size 2 in. diameter x 2-11/16 in. Solid zirconia ceramic vial, cap, and two 1/2 in. balls; seven corprene gaskets. Grinding load approximately 2-8ml;...
Page 13
Grinding load 0.2- 0.6 ml, mixing load 1ml. 8010 Multiple Sample Adapter Can be used in 8000D Mixer/Mill with 3116 vials, or with 3111, 3114, and 3117 vials if 8012 adapter is also used. Page 13 of 21 SPEX CertiPrep 8.21.03...
Page 14
(If not included in your Order, they are Available for Purchase at Additional Cost) 8011 Multiple Sample Adapter Can be used in 8000D Mixer/Mill with 3118, 5004, and 6133 vials. To use 3127 vials in 8000D, 8012 adapter is also necessary.
8000D Mixer/Mill 8.0 PARTS LIST & ITEM LOCATOR Location Item Part No. Description (Diagram, Page #) A, 17 37356 Single Clamp Assembly A, 17 50013 Bearing for Clamp (2 required per clamp) A, 17 33297 Clamp Retaining Spring Assembly A, 17...
8000D Mixer/Mill 10.0 TROUBLESHOOTING GUIDE PROBLEM CAUSE SOLUTION The unit does not turn on No AC power Make sure power cord is plugged into outlet and check outlet for power and correct as required The unit does not turn on...
Repairs, replacements, or parts are guaranteed for 30 days or for the remaining original warranty period (whichever is greater) for the item that was repaired or replaced. Items not produced by SPEX CertiPrep carry the manufacturer’s warranty only.
Need help?
Do you have a question about the 8000D MIXER/MILL and is the answer not in the manual?
Questions and answers
safety switch to be attended
For the SPEX CertiPrep 8000D MIXER/MILL, the following safety switch issues should be addressed:
1. Lid Open: The unit will not run if the lid is open. Ensure the lid is closed and latched properly.
2. Safety Interlock Switch: If the unit does not run even though the lid is closed and the timer counts down, the safety interlock switch may need adjustment.
3. Time-Delay Latch: This latch prevents the lid from opening during and immediately after operation. Wait for the audible click before opening the lid. Never force the lid open before the Time-Delay Latch unlocks.
This answer is automatically generated