E Instruments AQ Quick Reference Operating Instructions

Incuba air quality monitor

Advertisement

Quick Links

1. Push the On/Off power button to activate the monitor & sampling pump. Before doing any monitoring allow the unit to run for
approximately 40-60 seconds for the CO
2. If needed, connect the external black Temperature and Humidity module (Part# TRH Mod) as shown in Figure 1 above.
3. Select a Monitoring Position:
a. If the incubator has a glass/see-thru door and the AQ Incuba can fit safely inside the incubator without
damaging the instrument (ex: extremely high humidity), place the AQ Incuba with the TRH Module connected
inside the incubator. If this is not feasible or the instrument needs to remain outside the incubator, please see
part b.
b. If the instrument needs to remain outside the incubator, connect the Filter/hose assembly (AQ650078) shown
in Figure 2 to the Center "Sample-In" / "A" inlet at the bottom of the AQ Incuba. If temperature and humidity
need to be measured connect the optional TRH-EXT probe shown in Figure 3 to the bottom of the AQ Incuba.
The proper connections are shown in Figure 4 both. Figure 5 shows proper usage of the TRH-EXT and
AQ650078 together.
Figure 2
AQ650078 (Included)
E Instruments
AQ Incuba Air Quality Monitor
Quick Reference Operating Instructions
NOTE: FIND FULL DETAILED MANUAL ON INCLUDED THUMBDRIVE
NOTE: DO NOT LIFT OR HANDLE THE INSTRUMENT BY THE BLACK
EXTERNAL TEMP, %RH, PRESSURE SENSOR HOUSING!
sensor to stabilize & calibrate.
2
Figure 4
Figure 1
Figure 3
TRH-EXT (Optional)
Figure 5

Advertisement

loading

Summary of Contents for E Instruments AQ

  • Page 1 If the instrument needs to remain outside the incubator, connect the Filter/hose assembly (AQ650078) shown in Figure 2 to the Center “Sample-In” / “A” inlet at the bottom of the AQ Incuba. If temperature and humidity need to be measured connect the optional TRH-EXT probe shown in Figure 3 to the bottom of the AQ Incuba.