ParTech 7200 Instruction Manual

Field mount
Table of Contents

Advertisement

INSTRUCTION MANUAL
7200 Monitor
Field Mount

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the 7200 and is the answer not in the manual?

Questions and answers

Harry Drewett
September 2, 2025

What is the password to the patch 7200 monitor

Summary of Contents for ParTech 7200

  • Page 1 INSTRUCTION MANUAL 7200 Monitor Field Mount...
  • Page 2 ** This page is intentionally left blank **...
  • Page 3: Table Of Contents

    7200 Monitor Instruction Manual Table of Contents 1 Health & Safety Information............................6 2 Introduction...................................7 2.1 General................................7 2.2 Display................................7 2.3 Key Pad................................7 2.4 Alarms................................7 2.5 Analogue Output..............................7 2.6 Sensors................................7 3 Manual Conventions..............................8 4 MCERTS Product Certification............................9 5 Installation...................................10 5.1 Location................................10 5.2 Mounting................................10 5.3 Electrical Connections.............................11...
  • Page 4 7200 Monitor Instruction Manual 10.4 Fault Alarm Setup............................23 10.5 Delay to Set..............................23 10.6 Delay to Reset...............................23 10.7 Test Alarm Output............................23 10.8 Alarm Test Cycle............................23 11 Analogue Output...............................24 11.1 Set Output Low..............................24 11.2 Set Output High.............................24 11.3 Select Output Type............................24 11.4 Fault Output Mode............................25 11.5 Fault Output Setup............................25...
  • Page 5 7200 Monitor Instruction Manual 17 Diagnostics................................36 17.1 Sensor Input..............................36 17.2 Sensor %...............................36 17.3 Calibration Values............................36 17.4 Factory Values...............................36 17.5 Diagnostic Values............................36 17.6 Diagnostic Download.............................36 17.7 Opto Inputs..............................36 17.8 ADC Inputs..............................36 18 Information................................37 18.1 Statistics.................................37 18.2 Software Version............................37 18.3 Contact Information............................37 19 Instrument Security..............................38...
  • Page 6: Health & Safety Information

    7200 Monitor Instruction Manual 1 Health & Safety Information The following information is a guide and does not in anyway replace legal requirements for the handling of the equipment and chemicals or any safe working practise defined outside of this manual.
  • Page 7: Introduction

    The measurement title, units and display range can all be user set to suit the sensor being used and the parameter being monitored. Key Pad A six key tactile membrane keypad is used for all configuration and operation of the 7200 Monitor, each key has a very positive operation enabling use even whilst wearing gloves. Alarms Two user configurable alarms are provided within the 7200, which can be used to signal high, low or fault alarm conditions.
  • Page 8: Manual Conventions

    The manual has been written on the basis that the user has a basic knowledge of instrumentation and an understanding of the type of measurement being made. Training in the use of the 7200 Monitor can be provided by Partech, please contact sales for further information.
  • Page 9: Mcerts Product Certification

    The scheme provides a framework within which instruments and measurements can be made in accordance with the Agency's quality requirements. Partech Instrument's Certified Turbidity System consists of the 7200 Monitor and Turbi-Tech 2000LS Sensor. The system underwent laboratory testing in accordance with the MCERTS performance standard and can only be described as an MCERTS certified system when installed and operated as a system in accordance with recommended procedures.
  • Page 10: Installation

    5.1 Location The field mounting version of the 7200 Monitor is housed in an IP65 enclosure and therefore is suitable for mounting outdoors, preferably located close to the sensor. It is advisable to mount the monitor in a location where it has some protection from the elements, this could be inside a GRP or metal cabinet, in a control room, or with a protective shield.
  • Page 11: Electrical Connections

    If the cable being used is too large for the PG11 glands, the cable must be terminated externally to the unit and a suitable cable used for connection to the 7200. If any fitted cable glands are not used, these must be sealed to prevent the ingress of moisture into the unit.
  • Page 12 Note 1: The normal condition for the alarm outputs is when the instrument is powered and the alarm is off. Note 2: The RS232 connections can be used to collect data from the 7200 and using a suitable serial cable for connection to a PC.
  • Page 13: Basic Operational

    Measure Mode The normal operating mode for the 7200 Monitor is its measure mode and this is active when the measured value is displayed in large numbers. Whilst in measure mode the display value is continually updated based upon the current sensor input signal.
  • Page 14: Configuration

    7200 Monitor Instruction Manual 7 Configuration 7.1 Menu Mode To access the main menu, press the MENU key when the instrument is in measure mode. Pressing the MENU key when the main menu is displayed will return the instrument to the measure mode. When the main menu is displayed, keys can be used to move the selection bar.
  • Page 15: Data Entry

    Data Entry To provide the user with a simply interface, the 7200 uses three methods of data entry for the setting of all the user configurable parameters. The various methods of user data entry used on the 7200 are described below: 7.2.1...
  • Page 16: User Text Entry

    7200 Monitor Instruction Manual 7.2.3 User Text Entry Using this method of data entry, the user is able to enter or edit text and is used for the user defined measurement title and units. The text string is displayed with a flashing cursor indicating the selected character. The keys are used ...
  • Page 17: Calibration

    8 Calibration Calibration General Calibration of the 7200 Monitor is very important, errors in the calibration process will affect the measurement performance of the instrument. The following section describes the various aspects of calibrating the 7200 Monitor and has been written with the understanding that the user has knowledge of the measurement being made.
  • Page 18: Calibration Wizard

    Calibration menu, select Set Zero and press the OK key. Immerse the sensor into the zero solution and press the OK key. The 7200 Monitor collects data for the zeroing operation for approximately 5 seconds to ensure that the zero is set accurately.
  • Page 19: Set Span

    Step 2. Calibrating As indicated by the on display instructions, immerse the sensor into the calibration solution and press the OK key. The 7200 Monitor collects data for the spanning operation for approximately 5 seconds to ensure that the span is set accurately.
  • Page 20: Take Sample

    (The display value must be stable – this can be checked when in measure mode) press the OK key. The 7200 Monitor collects data for the take sample operation for approximately 5 seconds; during this period do not move the sensor.
  • Page 21: Setup Wizard

    9 Setup Wizard Setup Wizard The 7200 setup wizard simplifies the process of setting up the instrument when first installed by guiding the user through the basic (but important!) configuration steps. Prior to using the setup wizard, the following information should...
  • Page 22: Alarm Outputs

    10 Alarm outputs The 7200 Monitor has two alarm outputs designated Alarm 1 and Alarm 2. Each alarm has a single pole changeover output that signals the alarm condition. During normal operation when the alarm is not active the alarm output will be in its NORMAL condition, the N/O (normally open) contact will be open and the N/C (normally closed) contact will be closed.
  • Page 23: Fault Alarm Setup

    7200 Monitor Instruction Manual The alarm hysteresis value can be adjusted between 0 – 10% of the selected display range using the keys.  Pressing OK key will enter the new value, pressing the MENU key cancels any changes made and exits from the hysteresis set up.
  • Page 24: Analogue Output

    7200 Monitor Instruction Manual 11 Analogue Output The following section describes the setting up of the analogue output. The analogue output has three modes of operation as shown in the table below: Output Type High 4 – 20 mA 4 mA 20 mA 0 –...
  • Page 25: Fault Output Mode

    7200 Monitor Instruction Manual 11.4 Fault Output Mode When any of the fault conditions occur, the output can be configured to go to one of three values, two of which can be used to remotely detect the instrument fault. Last Valid Measurement: The output will be held at the last value that was measured before the fault occurred and will remain held until the fault is cleared.
  • Page 26: Sensor Configuration

    7200 Monitor Instruction Manual 12 Sensor Configuration Page 26 of 50 167300IM-Issue 12 Issue Date 14/10/2014...
  • Page 27: Select Sensor Type

    12.1 Select Sensor Type The sensors that can be configured for use with the 7200 Monitor are displayed as a list, a small tick indicates the currently selected sensor type. To select a different sensor type, use the keys to highlight the required sensor and ...
  • Page 28: Set Damping Rate

    12.4 Air Clean Option If the 7200 Monitor is being used with a Partech Air Clean System, this menu is used to enable the air clean function. Selecting yes will enable air clean operation and the air clean setup menu will be displayed on the Instrument Config menu.
  • Page 29: Advanced Sensor Setup

    These values along with the Y values are used to perform linearisation on certain sensor types by means of a lookup table. The X values represent the input values to the table and the Y values are used to determine the linear output values. For further information on the use of lookup table linearisation please contact Partech. Adjust Y Values See Adjust X values above.
  • Page 30: Instrument Setup

    7200 Monitor Instruction Manual 14 Instrument Setup 14.1 Measurement Title This menu is used to select the measurement title displayed in measure mode. The user can select from a list of predefined titles or enter a custom title. To select a measurement title, use the keys to highlight the required title ...
  • Page 31: Bleeper

    7200 Monitor Instruction Manual 14.4 Bleeper This option allows the bleeper to be turned on or off, a small tick indicates the current setting. To change the bleeper setting, use the keys to select the required setting and press OK. Press MENU to exit the bleeper menu.
  • Page 32: Clean Setup

    The Partech range of self cleaning sensors are designed such that the 7200 Monitor determines when a clean cycle is performed. During a clean cycle, the 7200 Monitor display value, alarm outputs and analogue output are all frozen, and are only unfrozen once the clean cycle is complete.
  • Page 33: Clean Information

    7200 Monitor Instruction Manual 15.3 Clean Information This display provides general information regarding the sensor cleaning process. Time until clean This indicates the time until the next clean cycle in hours, minutes and seconds. Clean Count The clean count indicates the number of cleans that the sensor has performed since it was installed or last serviced, and the set service life that has been set.
  • Page 34: Air Clean Setup

    The air clean setup and manual air clean menu’s are only display if the air clean option has been enabled on the sensor setup menu. The air clean option should only be enabled when the 7200 is being used with a Partech air clean system.
  • Page 35: Clean Information

    7200 Monitor Instruction Manual 16.4 Clean Information This display provides general information regarding the sensor cleaning process. Time until clean This indicates the time until the next clean cycle in hours, minutes and seconds. Clean Count The clean count indicates the number of air cleans that have been completed since the air pump was installed or last serviced.
  • Page 36: Diagnostics

    17 Diagnostics The diagnostics menu provides information regarding the internal setup and operation of the instruments and is a very useful facility in assisting problem diagnosis. If a problem has occurred, contact Partech technical support for assistance. 17.1 Sensor Input...
  • Page 37: Information

    This display indicates the software version and serial number of the instrument and is useful for recording and problem diagnosis. 18.3 Contact Information This display provides contact information for Partech (Electronics) Ltd. 167300IM-Issue 12 Issue Date 14/10/2014 Page 37 of 50...
  • Page 38: Instrument Security

    19 Instrument Security The 7200 Monitor can be locked to prevent unauthorised access to the configuration menu’s by using a password. The user password is a 4 digit number that can be set to any value the user chooses. When supplied new, the user password is set to 1000, we recommend that this is changed when the instrument is installed.
  • Page 39: Error Messages

    If nothing is found remove and reapply power to the monitor and use manual clean to activate a clean cycle. If the fault reoccurs, contact Partech technical support for assistance. 167300IM-Issue 12 Issue Date 14/10/2014...
  • Page 40: Calibration

    7200 Monitor Instruction Manual 20.4 Calibration Not Available - Sensor Fault If one of the above sensor faults is present, it is not possible to perform a calibration. Clear the sensor fault and try again. Sensor Input Too High (Set Zero) The data collected for the zero operation was above the allowed maximum for a zero operation.
  • Page 41: Menu Structure

    7200 Monitor Instruction Manual 21 Menu Structure Manual Clean* Sensor Defaults Clean Setup* Calibration Clean Interval Calibration Wizard Clean Mode Set Zero Clean Information Set Span Service Life Take Sample Reset Clean Count Enter Sample Result Air Clean Setup** Instrument Configuration...
  • Page 42: Sensor Defaults

    7200 Monitor Instruction Manual 22 Sensor Defaults The sensor default values are shown in the table below: Sensor Type Calibration Value Display Range Units Equation 0 – 100.0 mg/l IR15 10000 0 – 10000 mg/l IR40 1500 0 – 10000...
  • Page 43: Preparation Of Fullers Earth Solution

    7200 Monitor Instruction Manual 23 Preparation of Fullers Earth Solution The Fuller's Earth is supplied in a sealed 20g package, the package contains sufficient to make a 20,000 mg/l (20 g/l) solution when combined with 1 litre of water. Alternate solutions can be obtained by increasing the amount of water used as shown in the table below.
  • Page 44: Preparation Of Formazin Turbidity Standardise

    7200 Monitor Instruction Manual 24 Preparation of Formazin Turbidity Standardise Health & Safety Precautions The chemicals used when following this procedure are harmful, therefore the correct safety precautions must be carried out. During handling, avoid inhalation and contact with the eyes or skin. Wash hands thoroughly after use.
  • Page 45: Technical Support

    Application details • Description of fault • 25.1 Returning Equipment for Repair If equipment needs to be returned to Partech for repair or service the following address should be used: SERVICE DEPARTMENT PARTECH INSTRUMENTS ROCKHILL BUSINESS PARK HIGHER BUGLE ST AUSTELL...
  • Page 46: Technical Specification

    7200 Monitor Instruction Manual 26 Technical Specification 26.1 General Supply Voltage ....... 85-264VAC, 50/60 Hz, Power Consumption......15 VA Supply Fuse........1.6A Anti surge 20 x 5mm (Fitted Internally) Operating Temperature....0 to 50°C Storage Temperature......-20 to 70°C Display Resolution......0.1% Response Time.......0.2 to 120 Seconds (User selectable) Linearity...........(Refer to sensor specification)
  • Page 47: Declaration Of Conformity

    7200 Monitor Instruction Manual 27 Declaration of Conformity DECLARATION OF CONFORMITY According to EN 45014 We, Partech Instruments, Rockhill Business Park, Higher Bugle, St Austell, Cornwall, PL25 3NN, UK declare under our sole responsibility that the product: Product Name: 7200 Monitor...
  • Page 48 7200 Monitor Instruction Manual 167300IM-Issue 12 Issue Date 14/10/2014 Page 48 of 50...
  • Page 49 7200 Monitor Instruction Manual 167300IM-Issue 12 Issue Date 14/10/2014 Page 49 of 50...
  • Page 50 Partech Instruments Rockhill Business Park, Higher Bugle, St Austell, Cornwall, PL26 8RA, UK Tel: +44(0)1726 879800 Email: info@partech.co.uk Web: www.partech.co.uk...

Table of Contents