Schele Electronic GA-S1 User Manual

Portable analyzer for o2 and co2

Advertisement

Quick Links

User manual
Copyright © 2015 Schele Electronic UG & Co KG
Portable analyzer GA-S1
for O
and CO
2
2
Version 2015-5

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the GA-S1 and is the answer not in the manual?

Questions and answers

Summary of Contents for Schele Electronic GA-S1

  • Page 1 User manual Copyright © 2015 Schele Electronic UG & Co KG Portable analyzer GA-S1 for O and CO Version 2015-5...
  • Page 2: Table Of Contents

    Image 1: Calibrating gas connection with the included hose ..............7 Image 2: Connection of calibrating gas with a testing bottle ..............8 OVERVIEW TABLES Table 1: Automatic charger ........................18 Table 2: Spare parts ..........................18 Portable analyzer 2 of 18 GA-S1 Version 2015...
  • Page 3: Introduction

    Schele Electronic UG & Co KG 1.1 About this manual This manual contains instructions about how to work with the portable analyzer of Schele Electronic in order to lead to safe and efficient operations. Please read the entire manual carefully before starting up the portable analyzer.
  • Page 4: Contact

    1.4 Contact For questions and remarks you can contact: Schele Electronic UG & Co KG Antoine Schele Zum Brachsen 3/2 88662 Überlingen Germany Tel: +49 (0)7551 7211 Fax: +49 (0)7551 9499994 E-Mail: info@schele-electronic.de 2. Guarantee When using the portable analyzer appropriately as described in this user manual the guarantee period is 12 months.
  • Page 5: Analyzer Start Up

     = trend upwards  = trend downwards READY = ready for use Battery full indication Battery low indication 0: 4: 23 = Operating time [h:mm:ss] is 0 hours 4 minutes 23 seconds Portable analyzer 5 of 18 GA-S1 Version 2015...
  • Page 6: Checking The Analyzer

    The user is entirely responsible if the user carries out calibration himself. In this case, Schele Electronic cannot be held to account for any damages. In the event of no sample gas suction, the suction hoses and the connected filter have to be checked.
  • Page 7: Control Procedure

    The quantity of gas in the bottle has to be a bit larger than the quantity drawn by the analyzer to avoid air from the outside from entering, see Image 2: Connection of calibrating gas with a testing bottle. This can be identified by gas escaping the testing bottle. Portable analyzer 7 of 18 GA-S1 Version 2015...
  • Page 8: Image 2: Connection Of Calibrating Gas With A Testing Bottle

    Connec tion of portable analyser Ansc hlu am tragbare Me ger te ß ß ä Raccordement S url’analyseur portable (5%) escaping calibrating gas (1%) testing bottle Image 2: Connection of calibrating gas with a testing bottle Portable analyzer 8 of 18 GA-S1 Version 2015...
  • Page 9: Calibration

    Calibration can be interrupted by pressing function key F1 (EXIT)  If the measured value is stable, save the value with function key F4 (SAVE)  To stop the calibration press function key F1 (EXIT) Portable analyzer 9 of 18 GA-S1 Version 2015...
  • Page 10: Calibrating The Co Sensor With Reference Gas

     If you use the testing bottle: Take out the hose connected to the analyzer of the bottle  Press function key F4 (NEXT) Calibration can be interrupted by pressing function key F1 (EXIT) Portable analyzer 10 of 18 GA-S1 Version 2015...
  • Page 11  If the measured value is stable, save the value with function key F4 (SAVE)  To stop the calibration press function key F1 (EXIT) Portable analyzer 11 of 18 GA-S1 Version 2015...
  • Page 12: Save And Reload Default Settings

    Press the right arrow key until the following screen is displayed  Press function key F3 (DEFAULT)  Confirm default read with function key F4 (YES) or cancel with function key F1 (NO) Portable analyzer 12 of 18 GA-S1 Version 2015...
  • Page 13: Reset To Factory Settings

    Press the right arrow key until the following screen is displayed:  Press function key F2 (RESET), see above  Confirm reset read with function key F4 (YES) or cancel with function key F1 (NO) Portable analyzer 13 of 18 GA-S1 Version 2015...
  • Page 14: Battery

    Mistreatment may lead to premature ageing and defects, and even to fire and explosion in an extreme case.  Keep the battery out of the reach of children. Portable analyzer 14 of 18 GA-S1 Version 2015...
  • Page 15 Before a protracted period of storage batteries should be charged up to the storage voltage of around 3.8 ... 3.9 Volts / cell. After about 3 to 5 months the packs should be recharged again. Do not keep or Portable analyzer 15 of 18 GA-S1 Version 2015...
  • Page 16 Exclusion of liability Since Schele Electronic is unable to monitor the way you handle these batteries, we expressly deny any liability and any claim under guarantee if the battery is charged, discharged or handled incorrectly.
  • Page 17: Display Operating Time

    Press function key F4 (SCREEN) The operating time is displayed in [h:mm:ss], the number of days and total number of hours.  Return to current measurements screen by pressing function key F1 (EXIT) or F4 (NEXT) Portable analyzer 17 of 18 GA-S1 Version 2015...
  • Page 18: Technical Specifications

    Automatic charger GS40A24-P1J Input voltage 230V AC 50Hz Output voltage 24VDC 1,67A 40W Table 1: Automatic charger Description Type Article code Hydrophobic 30mm PTFE 0,2 μm Filter 0ANA100023 Table 2: Spare parts Portable analyzer 18 of 18 GA-S1 Version 2015...