AquaLab 4TE DUO Quick Start
1. Getting Started—Chapter 4
• Place AquaLab on a level surface.
• Use AquaLab where environmental temperature remains fairly stable.
• Plug in AquaLab and turn on power (switch in back).
• For best results, let AquaLab warm-up for 15 min.
2. Sample Preparation—Chapter 8
• Fill sample cup no more than half full (-7ml). Do Not Overfill!
• Minimum sample amount should cover bottom of sample cup
(small gaps OK).
• Ensure rim and outside of sample cup are clean.
• Make sure sample temperature is not more than 4°C above chamber
temperature.
3. Taking a Reading—Chapter 9
• Place sample cup in sample chamber. Close lid carefully to avoid spillage.
• Use the up and down arrows to select the desired moisture content model.
WARNING: Selecting the incorrect model will cause you to get an
inaccurate moisture content reading.
NOTE: Contact AquaLab if you don't have a model for a product.
• Move latch left to start measurement.
• AquaLab beeps once when starting
a
• First
value will display after one to two minutes, but reading is not
w
complete until save icon appears.
• Sample's final
4. AquaLab Series 4TE DUO Menu—Chapter 5
Measurement

Calibration Menu

• Press Menu Icon on main screen.
• Press Enter twice to start calibration.
a
reading, temperature and moisture content will display.
w
Configuration
Enter Icon Menu Icon
a
measurements.
w
Data

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the 4TE DUO and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for AquaLab 4TE DUO

  • Page 1: Calibration Menu

    WARNING: Selecting the incorrect model will cause you to get an inaccurate moisture content reading. NOTE: Contact AquaLab if you don’t have a model for a product. • Move latch left to start measurement. • AquaLab beeps once when starting measurements.
  • Page 2 • For cleaning instructions, see Chapter 10. Cautions • Do not overfill sample cups. • Never tip or move AquaLab with sample loaded in sample chamber. • Sample chamber contamination can damage AquaLab or cause inaccurate results. Call for AquaLab technical support •...

Table of Contents