Page 3
User Manual INSTRUCTIONS FOR USE Version 1.1 July 2015 Version 1.1 July 2015 Pg. 3...
Page 4
User Manual Implemented Revision Approved Approval Version Reason Date Date Ben Smith G. Ball June ‘15 Original copy Corrective actions by Ben Smith July ‘15 Cranage EMC Testing Limited. (IEC 60601-1) Version 1.1 July 2015 Pg. 4...
User Manual The NoDec AIRSTREAM 2 replaces the mattress on any hospital, ® style bed frame. This mattress replacement system is quiet, comfortable and simple enough for a single caregiver installation. The frequent relief of contact (interface)pressure has been demonstrated to reduce the risk of pressure injury which is caused by reduced capillary blood flow in the tissue surrounding the bony prominences.
User Manual Contraindications As with all alternating mattress systems, the NoDec AIRSTREAM ® is contraindicated for use where patients have unstable fractures or spinal injuries. Descriptions of Symbols Throughout this user manual and on the pump unit itself, the NoDec AIRSTREAM 2 has numerous symbols: ®...
User Manual Unpack your NoDec AIRSTREAM 2 mattress system and check ® that you have the following: NoDec AIRSTREAM 2 Pump Unit: ® Please note the Name, Model and the Serial Number of your pump in case you need to reference these when contacting Rober or your local distributor.
User Manual Before setting up the mattress, it is important to read all of the safety instructions below. Any actions or operations performed outside the guidance of this manual are carried out at the user’s risk. Rober and/or the supplier are not liable for any such actions. ...
Page 11
User Manual Do not place any components of the mattress system in either heat or steam autoclaves. Risk of Suffocation: The cover is not air permeable. Ensure the patient can use the product safely. Do not drop the pump unit and do not drop objects onto it. ...
User Manual Please follow these instructions to install your NoDec AIRSTREAM 2 mattress replacement safely and ® appropriately. Place the mattress onto the bed frame, ensuring the light grey top cover is facing upwards and the air supply hose (umbilical) is positioned at the foot end of the bed.
Page 13
After installation carefully test the functions of the bed to ensure the mattress moves freely with the bed platorm. The hooks on the back of the NoDec AIRSTREAM 2 have been ® designed to accommodate a variety of bed frames and can be adjusted easily by pushing down on the hooks and rotating.
Page 14
User Manual Hang the unit onto the bed frame at the foot end of the bed with the control panel facing outwards. Check that the CPR clip is in the closed position and plug the umbilical hose into the dock. Version 1.1 July 2015 Pg.
Page 15
(waterfall). This will help reduce any hazards in relation to the power cable. Wait for approximately 20-30 minutes and your NoDec ® AIRSTREAM 2 Mattress Replacement System is ready for use. Version 1.1 July 2015 Pg. 15...
Loading There will be 4 cells shown slowly inflating simultaneously on the screen whilst the NoDec unit pumps up the mattress. Unlocking 60 seconds after either full inflation or mode change, the system will automatically lock out the display panel.
User Manual Different Modes / Displays The NoDec AIRSTREAM 2 mattress is configured into two decks ® of air cells, one deck (Integrated Cells) with lateral cells sitting on top of another with longitudinal cells. While the bottom deck remains inflated, including two side formers, the 18 air cells of the top deck are pneumatically connected into two groups.
Page 18
User Manual Alternating Mode The air cells inflate and deflate in a 1 in 2 pattern with a cycle time of 10 minutes. When the "A" cells deflate, pressure is reduced in that portion of the body in contact with those cells. After 5 minutes, the "A"...
User Manual The NoDec AIRSTREAM 2 has four different modes: ® Firm Alternating Firm Static Soft Alternating Soft Static When pressing the mode button multiple times the screen will cycle through these different modes until the desired mode is found.
The mattress remains at the lower range of pressure settings and alternation ceases. The ‘SOFT’ setting of the NoDec AIRSTREAM 2 is suitable ® for frail underweight patients and children. N.B. Both Static Modes revert to Alternating Mode after a preset period of 30 or 120 minutes, depending on your local distributor.
User Manual Each of the following alarm displays will be accompanied by a fixed tone audible alarm. The audible alarm can be cancelled by pressing the “Mode” Button, however if the cause of the alarm is not resolved after 15 minutes the audible alarm will reactivate. To turn off the alarm permanently, or reset the pump after fixing the cause of the alarm, simply switch the pump unit off and then on again.
User Manual Low Pressure Alarm This condition could be caused by releasing the CPR valve or a leak in the mattress. This alarm is shown by FOUR flat deflated cells FLASHING RED. Electrical Failure If the 3-way valves within the pump should fail then a FLASHING RED lighting bolt will appear across a...
User Manual Power Failure Alarm If there is a failure of the mains power supply or the unit is accidently unplugged form the mains then a plug symbol with a RED FLASHING cross will show. N.B. The Power Failure Alarm is powered through a small battery. The sole function of this battery is to operate the alarm in case of a mains power failure.
Page 24
User Manual In the case of a patient emergency where CPR may be necessary, the NoDec AIRSTREAM 2 has been equipped with a CPR release ® which rapidly deflates the mattress so that cardiac resuscitation can be performed on a firm surface.
User Manual In this configuration the mattress can be transported easily around the hospital. Preparation When separating the pump unit from the mattress the unit should first be SET TO STATIC MODE. When all cells are inflated the pump should be SWITCHED OFF and the umbilical hose seperated.
User Manual Cleaning Procedures Cleaning the mattress system every week is recommended, or sooner if heavily soiled. Please make sure that hands are cleaned and appropriate protective clothing is worn before commencing cleaning process. Cleaning the mattress Leave the cover on the mattress and disconnect the mattress from the unit.
User Manual Wipe the umbilical hose with hot water and detergent. Dry with a paper towel and repeat wiping down using hypochlorite or equivalent solution. If required, the cover can be removed and machine-washed at a temperature of 80°C for a minimum of 10 minutes. The cover may also be sterilised using ETO or Draeger processes.
Page 28
User Manual Power Unit Model NoDec AIRSTREAM 2 ® Dimensions (L x W x H) 400 x 160 x 250 mm 2.8 Kg Weight 230V~ ±10%, 50 / 60 Hz Power Supply Power Consumption 20w max IP42 Ingress Protection Rating...
Page 30
User Manual Version 1.1 July 2015 Pg. 30...
Page 31
User Manual Version 1.1 July 2015 Pg. 31...
Page 32
User Manual Rober Limited Rober (India) Private Limited Subsidiary of Rober Limited UK Wheatbridge Road Chesterfield No 1, Ground Floor, RNG Derbyshire Palazzo S40 2AB K.P.E. Extension Bangalore 560001 Karnataka India t +44 (0)1246 245404 t +91 (0) 8065555683 f +44 (0)1246 245400 e info@roberlimited.in e info@roberlimited.com w www.roberlimited.in...