Harvard Apparatus Pump 11 Elite OEM User Manual

Pump 11 elite oem module series

Advertisement

Quick Links

Pump 11 Elite OEM
Module Series
U SER'S MANUAL
Harvard Apparatus Pump 11 Elite OEM Module Series
Publication 5420-007 Rev B
800.272.2775
phone
bioscience@harvardapparatus.com
email

Advertisement

Table of Contents
loading

Summary of Contents for Harvard Apparatus Pump 11 Elite OEM

  • Page 1 Pump 11 Elite OEM Module Series U SER’S MANUAL Harvard Apparatus Pump 11 Elite OEM Module Series Publication 5420-007 Rev B 800.272.2775 phone bioscience@harvardapparatus.com email...
  • Page 2: Table Of Contents

    Run Commands ..................15 Rate Commands ..................15 Volume Commands ..................16 Time Commands ..................17 Digital I/O Commands ................18 Internal Commands ..................18 Serial Command Notes ................18 508.893.8999 508.429.5732 bioscience@harvardapparatus.com phone e-mail www.harvardapparatus.com • Pump 11 Elite OEM Module Series User’s Manual Publication 5420-007 Rev B...
  • Page 3: Table Of Contents

    Legacy RJ-11 Option ................26 Appendix D: Maintenance ...................27 Maintenance .....................27 Upgrading Pump 11 Elite Series Software ..........28 Troubleshooting ..................29 Appendix E: Ordering Information ...............30 508.893.8999 508.429.5732 bioscience@harvardapparatus.com phone e-mail www.harvardapparatus.com • Pump 11 Elite OEM Module Series User’s Manual Publication 5420-007 Rev B...
  • Page 4: General Information

    OBSERVE ALL WARNING LABELS ON PRODUCT Read all labels on product to ensure proper usage. CAUTION Protective Refer to Manual Ground Terminal 508.893.8999 508.429.5732 bioscience@harvardapparatus.com phone e-mail www.harvardapparatus.com • Pump 11 Elite OEM Module Series User’s Manual Publication 5420-007 Rev B...
  • Page 5: Warranty And Repair Information

    Harvard Apparatus warranties this instrument for a period of two years from date of purchase. At its option, Harvard Apparatus will repair or replace the unit if it is found to be defective as to workmanship or materials. This warranty does not extend to damage resulting from misuse, neglect or abuse, normal wear and tear, or accident.
  • Page 6: Specifications

    24.13 x 17.15 x 10.67 cm (9.5 x 6.75 x 4.2 in) Weight 2.3 kg (5.08 lbs) 2.3 kg (5.08 lbs) 508.893.8999 508.429.5732 bioscience@harvardapparatus.com phone e-mail • Pump 11 Elite OEM Module Series User’s Manual Publication 5420-007 Rev B www.harvardapparatus.com...
  • Page 7 61010-1 ; CB Scheme 61010-1 ; CB Scheme EMC Declaration FCC 47CFR 15B Class A; EN61326-1 FCC 47CFR 15B Class A; EN61326-1 508.893.8999 508.429.5732 techsupport@harvardapparatus.com phone e-mail www.harvardapparatus.com • Pump 11 Elite OEM Module Series User’s Manual Publication 5420-007 Rev B...
  • Page 8: Product Overview

    Up to 16 kg (35 lbs) of linear force is available. This force Programmable is user adjustable to ensure the right force is applied for the various syringe sizes. 508.893.8999 508.429.5732 bioscience@harvardapparatus.com phone e-mail www.harvardapparatus.com • Pump 11 Elite OEM Module Series User’s Manual Publication 5420-007 Rev B...
  • Page 9: Operating Instructions

    FIREWIRE PORTS ON A COMPUTER. DAMAGE MAY OCCUR TO THE Best performance guaranteed when using 24V power supply. PUMP OR COMPUTER. 508.893.8999 508.429.5732 bioscience@harvardapparatus.com phone e-mail www.harvardapparatus.com • Pump 11 Elite OEM Module Series User’s Manual Publication 5420-007 Rev B...
  • Page 10: Syringe Loading

    The Pump 11 Elite OEM Module can accommodate 6. Secure the syringe flanges by pushing the from 0.5 µl to 60 ml single syringe and 0.5 µl to 10 ml retaining brackets (4,5) tightly against the dual syringe.
  • Page 11: External Pump Control

    Hyperterminal, turn off the pump, unplug next to this in parenthesis. then re-plug the USB cable, turn on the pump and re-open Hyperterminal. 508.893.8999 508.429.5732 bioscience@harvardapparatus.com phone e-mail www.harvardapparatus.com • Pump 11 Elite OEM Module Series User’s Manual Publication 5420-007 Rev B...
  • Page 12: Pump Chain Commands

    A space must • <lf>[##]<prompt> is the response unless follow the command if arguments are included. otherwise noted. 508.893.8999 508.429.5732 bioscience@harvardapparatus.com phone e-mail www.harvardapparatus.com • Pump 11 Elite OEM Module Series User’s Manual Publication 5420-007 Rev B...
  • Page 13: Error Messages

    [9600 | 19200 | 38400 | 57600 | 115200 | 128000 | 230400 | 256000 | 460800 | 921600] <lf>[##]<prompt> Query response: <lf>[##:]# baud<cr> <lf>[##]<prompt> 508.893.8999 508.429.5732 bioscience@harvardapparatus.com phone e-mail www.harvardapparatus.com • Pump 11 Elite OEM Module Series User’s Manual Publication 5420-007 Rev B...
  • Page 14 METRICS Displays the pump metrics. The operation of this command is covered in the 11 Elite Configuration document. Command format: metrics 508.893.8999 508.429.5732 bioscience@harvardapparatus.com phone e-mail www.harvardapparatus.com • Pump 11 Elite OEM Module Series User’s Manual Publication 5420-007 Rev B...
  • Page 15: Run Commands

    STOP / STP <lf>[##]<prompt> Stops the pump. Command format: <lf>[##:] {volume}, {unit}<cr> (this line is repeated stop for each syringe size) <lf>[##]<prompt> 508.893.8999 508.429.5732 bioscience@harvardapparatus.com phone e-mail www.harvardapparatus.com • Pump 11 Elite OEM Module Series User’s Manual Publication 5420-007 Rev B...
  • Page 16: Rate Commands

    For commands requiring rate units, use the following format: m, u, n, p/h, m, s. ex: m/m = milliliter/minutes ex: n/s = nanoliter/second 508.893.8999 508.429.5732 bioscience@harvardapparatus.com phone e-mail www.harvardapparatus.com • Pump 11 Elite OEM Module Series User’s Manual Publication 5420-007 Rev B...
  • Page 17: Volume Commands

    Sets or displays the target volume. Quick Start mode Query response: only. Command format: <lf>[##:]# seconds<cr> <lf>[##]<prompt> tvolume [{target volume} {volume units}] <lf>[##:]##:##:##<cr> <lf>[##]<prompt> 508.893.8999 508.429.5732 bioscience@harvardapparatus.com phone e-mail www.harvardapparatus.com • Pump 11 Elite OEM Module Series User’s Manual Publication 5420-007 Rev B...
  • Page 18: Digital I/O Commands

    “w or W” is displayed. If no limit switch was hit or the pump does not have limit switches, “.” is displayed. <lf>[##:]##:##:##<cr> <lf>[##]<prompt> The Pump 11 Elite OEM Module does not contain limit switches, so “.” will be displayed. Digital I/O commands INPUT Flag three is the stall status and will be “S”...
  • Page 19: Appendices

    0.5 µl to 60 ml single syringe and 0.5 µl to 10 ml dual 10 ml. syringe. 508.893.8999 508.429.5732 bioscience@harvardapparatus.com phone e-mail www.harvardapparatus.com • Pump 11 Elite OEM Module Series User’s Manual Publication 5420-007 Rev B...
  • Page 20: Appendix B: Min/Max Flow Rates

    Appendices APPENDIX B: PUMP 11 ELITE OEM MODULE (SINGLE & DUAL) MIN/MAX FLOW RATES Nominal Minimum/Maximum Flow Rates for Various Syringes. (Actual Limits will vary depending on syringe manufacturer) Minimum Maximum Syringe Size Syringe ID Units Units Rate Rate 0.5 µl 0.103 mm...
  • Page 21 8.585 mm 3.990 nl/min 4.144 ml/min 5 ml 11.989 mm 7.782 nl/min 8.082 ml/min 10 ml 14.430 mm 11.270 nl/min 11.700 ml/min 508.893.8999 508.429.5732 bioscience@harvardapparatus.com phone e-mail www.harvardapparatus.com • Pump 11 Elite OEM Module Series User’s Manual Publication 5420-007 Rev B...
  • Page 22: Appendix C: External Connections

    Electrical Specifications Inputs: ≥ 2V ≤ 20µA ≤ 0.4V ≤ 0.5mA Outputs: V ≥ 3.8V ≤ 6mA ≤ 0.4V ≤ 6mA 508.893.8999 508.429.5732 bioscience@harvardapparatus.com phone e-mail www.harvardapparatus.com • Pump 11 Elite OEM Module Series User’s Manual Publication 5420-007 Rev B...
  • Page 23: Utilizing Usb Interface

    Other Devices. Right click on either of the flash update applications and select Update Driver Software. 508.893.8999 508.429.5732 bioscience@harvardapparatus.com phone e-mail www.harvardapparatus.com • Pump 11 Elite OEM Module Series User’s Manual Publication 5420-007 Rev B...
  • Page 24 5. If a warning message regarding the driver publisher is displayed, choose Install this driver software anyway. 6. Click Close after the software has been successfully installed. 508.893.8999 508.429.5732 bioscience@harvardapparatus.com phone e-mail www.harvardapparatus.com • Pump 11 Elite OEM Module Series User’s Manual Publication 5420-007 Rev B...
  • Page 25: Rs-485

    1.0.6 or lower can be all RS-485 powered units deriving power returned to manufacturer for a hardware from it. update. 508.893.8999 508.429.5732 bioscience@harvardapparatus.com phone e-mail www.harvardapparatus.com • Pump 11 Elite OEM Module Series User’s Manual Publication 5420-007 Rev B...
  • Page 26: Legacy Rj-11 Option

    CableCat # 55-7760 70-2022 55-7760 or 72-2478 or 72-2478 Daisy-Chaining with Legacy RJ-11 Option Pumps 1...n also have RJ-11 telephone jacks. 508.893.8999 508.429.5732 bioscience@harvardapparatus.com phone e-mail www.harvardapparatus.com • Pump 11 Elite OEM Module Series User’s Manual Publication 5420-007 Rev B...
  • Page 27: Appendix D: Maintenance

    For more efficient cleaning, use a soft cloth dampened [not soaked] with water, an aqueous solution of 75% isopropyl alcohol, or a mild detergent. 508.893.8999 508.429.5732 bioscience@harvardapparatus.com phone e-mail www.harvardapparatus.com • Pump 11 Elite OEM Module Series User’s Manual Publication 5420-007 Rev B...
  • Page 28: Upgrading Pump 11 Elite Series Software

    5. Click Open Update File and browse to the file previously saved on your desktop and click Open. 6. Click Start Update. 7. After the update is complete, press Reboot to power cycle the pump. 508.893.8999 508.429.5732 bioscience@harvardapparatus.com phone e-mail www.harvardapparatus.com • Pump 11 Elite OEM Module Series User’s Manual Publication 5420-007 Rev B...
  • Page 29: Troubleshooting

    A value was entered or encountered in a pump Method that was beyond the pump’s limits. 4. Overload Protection The Pump 11 Elite OEM Module Series incorporates resettable fuses for overload protection. In the event of an overload, power to the pump may be disconnected.
  • Page 30: Appendix E: Ordering Information

    For additional items such as Syringes, Tubing, Syringe Heaters, Spill Sensors, Temperature Controllers, Nano-Fluidic Accessories, and much more please visit our website www.harvardapparatus.com or call (800) 272-2775. 508.893.8999 508.429.5732 bioscience@harvardapparatus.com phone e-mail www.harvardapparatus.com • Pump 11 Elite OEM Module Series User’s Manual Publication 5420-007 Rev B...
  • Page 31 Product Name: Pump 11 Elite OEM Module Syringe Pump Series Model No.: Pump 11 Elite OEM Module Series- Catalog # 70-4xxx To which this declaration relates, is in conformity with the applicable EC Directives, harmonized standards, and other normative requirements:...

This manual is also suitable for:

Pico plus elite oem

Table of Contents