Test Products International 318C Instructions

Pocket digital thermometer

Advertisement

Quick Links

318C and 319C
Pocket Digital Thermometer
To measure temperature follow these simple instructions:
Selecting Fahrenheit or Celsius mode:
• With the unit "OFF" press and continue to hold the ON/OFF button. The LCD will alternate between °F
and °C
• Release the ON/OFF button when desired measurement mode appears on the LCD.
Remove the protective sheath from the stem.
Press the ON/OFF button
Insert stem into area or object to be tested. At least ½ inch of the stem must be inserted.
Warning: The stem is a sharp and pointed object which can be dangerous
The Data Hold Function is used by pressing the D-H button when taking a measurement. The measurement will
flash on the LCD when Data Hold is activated. To return to normal operation, press and release the D-H button
a second time.
Press the ON/OFF button to end operation or the unit will automatically power off after approx. 35 minutes.
Specifications:
Temperature Range
Resolution
Accuracy
Sampling Time
Battery
TPI 318C and 319C FIELD CALIBRATION
Fill a plastic or metal container with crushed ice and add clean water to a depth of at least 4 inches. Stir the ice
and water for 2 to 3 minutes prior to performing calibration to ensure the water is completely chilled. Make
certain there is plenty of ice in the mixture and always use clean water. Distilled water works well. The
temperature of an ice bath is approximately 32°F (0°C).
Insert the stainless steel shaft of the 318C and 319C into the ice bath making sure at least one inch of the tip is
immersed. Allow the reading on the thermometer to stabilize.
Note: The temperature reading must be within 23°F to 41°F (-5°C to 5°C) for calibration to have effect.
Press and hold down the HOLD button for approximately 8 seconds until "CAL" is displayed. "CAL" will
display for approximately 2 seconds and then the 318C and 319C will return to normal operation. Calibration is
complete.
Note: If the temperature reading was not within 23°F to 41°F (-5°C to 5°C) when step 3 was performed no
change in calibration occurred. The 318C and 319C was designed with this feature to prevent improper
calibration.
-58 to 300 °F/-50 to 150 °C
0.1 Degrees
+/- 2 °F, +/- 1 °C
1.5 Seconds