Panacol-Elosol UV-H 255 Operating Manual

Panacol-Elosol UV-H 255 Operating Manual

For hand-held uv lamp (polymerisation tool for industrial and laboratory applications)

Advertisement

Panacol-Elosol GmbH
Daimlerstr. 8
D- 61449 Steinbach/Ts
Germany
℡ +49 6171 6202 0
+49 6171 6202 590
info@panacol.de
Operating Manual
for hand-held UV lamp model UV-H 255
(polymerisation tool for industrial and laboratory applications)
To be observed prior to commissioning:
The instructions in this manual are binding. For a
disruption-free
that these instructions are adhered to.
In the event of
those
intended, all warranty is voided and Panacol-Elosol
rejects any
The operating instruction must be available near the UV lamp for
reference.
If you have any queries regarding the safety, operation or
maintenance of the device, please do not hesitate to contact Panacol-
Elosol.
BA_uv-h255 hand.held lamp 09_v1 US
GB.doc
operation of the UV lamp, it is imperative
improper use or use for purposes other than
liability for
damages.
safe
and
1

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the UV-H 255 and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for Panacol-Elosol UV-H 255

  • Page 1 Operating Manual for hand-held UV lamp model UV-H 255 (polymerisation tool for industrial and laboratory applications) To be observed prior to commissioning: The instructions in this manual are binding. For a safe disruption-free operation of the UV lamp, it is imperative that these instructions are adhered to.
  • Page 2: Table Of Contents

    => Risk to life and limb Caution! Risk of explosion! => Risk to life and limb Caution! UV radiation! => Damage to eyes, burns Panacol-Elosol GmbH Daimlerstr. 8 Caution! UV radiation! => Damage to eyes, burns D- 61449 Steinbach/Ts Caution! Hot surfaces! =>...
  • Page 3: General Safety Instructions

    Non flammable surface Risk of burns and dazzling UV radiation can damage skin and eyes. Always protect skin and eyes Panacol-Elosol GmbH against direct radiation. Daimlerstr. 8 D A N G E R D- 61449 Steinbach/Ts...
  • Page 4 The risk of exploding is thereby Prior to operation, inspect filter disk increased. Panacol-Elosol GmbH Safety measures for repair and cleaning work Daimlerstr. 8 Prior to starting repair or cleaning, disconnect device from the power D- 61449 Steinbach/Ts supply.
  • Page 5: Product Features

    UV radiation is used in a wide range of industrial laboratory applications, such as the detection of hairline cracks or the examination of bank notes to detect forgeries. The UV-H 255 lamp is exclusively designed for the hardening of –...
  • Page 6: Delivery And Accessories

    Never use the UV lamp as a tanning lamp or as a hair-dryer, as this could lead to serious burns and injuries. In the event of improper use or use for purposes other than those intended, all warranty is voided and Panacol-Elosol rejects any liability for damages. Delivery and accessories The following components are included in the delivery: –...
  • Page 7: Commissioning And Decommissioning

    To prevent damage to the device, do not connect it to a socket to which other high-powered devices are connected. Cable of the Hand Lamp should not be disconnected during operation of Panacol-Elosol GmbH control device. Daimlerstr. 8 The control device should be first shut D- 61449 Steinbach/Ts down and disconnected from plug.
  • Page 8: Operation

    Inspect filter disk goggles Switch on the control device by Front side pressing the power switch at the front of the housing (position I ). Panacol-Elosol GmbH Daimlerstr. 8 D- 61449 Steinbach/Ts Germany ℡ +49 6171 6202 0 +49 6171 6202 590 info@panacol.de...
  • Page 9 Never place the lamp on the fan slots or on all three plastic knobs of Non flammable surface the filter frame. Panacol-Elosol GmbH Daimlerstr. 8 D- 61449 Steinbach/Ts Germany ℡ +49 6171 6202 0 +49 6171 6202 590 info@panacol.de...
  • Page 10: Maintenance And Cleaning

    Unsuitable cables, metal halide tubes or fuses result in faults. Use only Panacol-Elosol GmbH original spare parts and accessories that fulfil the technical Daimlerstr. 8 requirements (see technical data). If in doubt contact Panacol-Elosol. D- 61449 Steinbach/Ts Germany ℡ +49 6171 6202 0 +49 6171 6202 590 info@panacol.de...
  • Page 11: Replacement Of Tube

    Caution! Hot surfaces! Before any maintenance or cleaning, allow the lamp to cool down for 30 minutes. >30 minutes Panacol-Elosol GmbH Daimlerstr. 8 D- 61449 Steinbach/Ts Germany ℡ +49 6171 6202 0 +49 6171 6202 590 info@panacol.de...
  • Page 12 Close the housing be completing the above steps in reverse order. Ensure that the filter is exactly centred within the housing. Panacol-Elosol GmbH Daimlerstr. 8 D- 61449 Steinbach/Ts Carry out a function test. The hand-held lamp is again ready for operation.
  • Page 13: Glass Filter Replacement

    Pull off the plastic cover (snap-on lid). Remove the dirty dust filter from the Dust filter housing. Panacol-Elosol GmbH Insert a new or cleaned filter. Daimlerstr. 8 To clean the filter, remove the visible dirt from its surface and clean it D- 61449 Steinbach/Ts subsequently with compressed air.
  • Page 14: Cleaning Of The Blue Filter

    It is recommended to regularly clean the housing. Clean the housing only when it is cold and after you have disconnected the device from the power supply! Panacol-Elosol GmbH Clean the housing with a moist soft cloth. Use only water. Daimlerstr. 8...
  • Page 15: Disposal

    Panacol-Elosol rejects any liability or warranty for damages caused by improperly completed repairs. Prior to returning the device to Panacol-Elosol, please check the device by following the steps in the table below. If none of the remedies below is successful, please contact your supplier.
  • Page 16 Fuse burnt out Contact appliance retailer to replace fuse Power failure Check other electrical devices and fuses Panacol-Elosol GmbH Daimlerstr. 8 D- 61449 Steinbach/Ts Germany ℡ +49 6171 6202 0 +49 6171 6202 590 info@panacol.de BA_uv-h255 hand.held lamp 09_v1 US...
  • Page 17: Technical Data

    4 metres (longer cable available on request) Weight Hand-held lamp: approx. 1.3 kg Control device: approx. 3.5 kg Noise: < 70 dB Panacol-Elosol GmbH Protection class according to IP 20 Daimlerstr. 8 VED 0710/part 1: D- 61449 Steinbach/Ts Germany ℡ +49 6171 6202 0 Subject to changes due to technical improvements.
  • Page 18: Lamp Replacement Maintenance Report

    Lamp replacement maintenance report Lamp type Hour counter Date Signature Panacol-Elosol GmbH Daimlerstr. 8 D- 61449 Steinbach/Ts Germany ℡ +49 6171 6202 0 +49 6171 6202 590 info@panacol.de BA_uv-h255 hand.held lamp 09_v1 US GB.doc...

Table of Contents