SHORT OPERATION INSTRUCTIONS
FOR STEAM STERILIZER HV SERIES
1. SAFETY DIRECTION
This short operating instructions do not replace the manual. Read in every case before start-up and/or before using
steamsterilizer the manual and obey the indicated safety directions!
2. DISPLAY AND OPERATION SWITCHES
(1) Indication of temperature, date, errors
(2) Sterilizing time, time, air evacuation %
(3) Unit display
(4) Cycle Display
(5) Mode Display
(6) Func. Display (timer, clock, print)
(7) Mode Switch for selection Mode and
checking set temperature or exhaust%
(8) Func. Switch for activating or
modification of functions
(9) Switch for modification of single parameters
(10) Switch for modification sterilizing parameters
3. OPERATION
Unit must be connected to supply (energy, demineralised-water, pressured air, sewerage)
WARNING: THE OPEN/CLOSE LEVER ONLY WORKS, WHEN POWER SWITCH IS TURNED TO „ON"!!.
Pour water into exhaust bottle till marking
Pour water through chamber until hole at the centre of heater cover.
Place the substance to be sterilised into the chamber.
Flexible temperature sensor
Be sure that temperature sensor is activated (press Switch Func. till „FL ON" comes
4. SELECTING MODE
Prog.
MODE
-No
Sterilization of agar medium (warmed for the
1
LIQ
prevention of coagulation after sterilization)
2
LIQ
liquids
3
SOLID Solids, instruments, glass
4
AGAR Dissolution of AGAR with warming cycle 50° C
Liquids with warming cycle 50° C
5
LIQ
6
LIQ
Liquids with cooling
Lid-opening-temperature
For precaution al Liquid programs have standardized take-from-temperature of 80° C
Modification lid-opening-temperature
Modification of lid-opening-temperature is only possible when flexible temperature sensor is installed and activated.
Select program,
Press switch for coming to programme mode, the temperature display blinks
Pressing switch to 80 oPn
Press switch up or down till needed lid-opening-temperature is reached for example 95 oPN
Press switch, the modified temperature is programmed.
Press switch, the lid opening temperature will be announced short time and the program starts
Direction:
80° C lid-opening-temperature for Liquids in close d glass bottles
90° C lid-opening-temperature for Liquids in closed plastic bottles
95° lid-opening-temperature for Liquids in open bo ttles
V:\HMC-Europe\Beschreibungstexte\HV\Bedienungsanleitungen\Kurzbedienungsanleitungen\Kurzbedienungsanleitung engl..doc
Description
first you reach the program number, then the standard opening temperature [80 oPN]
(11) Switch for selection parameters
(12) Switch starts and stops the program
Lid
opening.
80/90/95° C
80/90/95° C
98° C
80/90/95° C
80/90/95° C
only for fast
recooling
80/90/95° C
Sterilising
Sterilising
temp.
time
121° C
20
121° C
20
121° C
20
100° C
20
121° C
20
121° C
20
Stand Juli 2003
Exhaust
%
10
10
10
10
cooling
cooling