Land Instruments International SPOT User Manual page 8

Modbus controls

Advertisement

Modbus
Name
Type
Bytes
address
(16 bit)
14 lo
LED
uc
1
14 hi
Display booleans
uc
1
14 hi
Display booleans
uc
1
15 - 18
Name
c(8)
8
19 lo
Calibration Current
uc
1
Value
19 hi
IOMode booleans
uc
1
1
20
Current output low
s
2
temperature (0 or 4mA
equivalent tempera-
ture)
21
Current output high
s
2
temperature limit
(20mA equivalent tem-
perature)
22
High target alarm
s
2
threshold
23
Low target alarm
s
2
threshold
24 lo
High instrument alarm
c
1
threshold
24 hi
Low instrument alarm
c
1
threshold
25 lo
AlarmMode booleans
uc
1
Bits
Default
Permitted Range
Flash rate
5
1 - 30 flashes per sec-
ond
7 set IP
0x81
0 = OHCP,
1 = Set IP
6 Use back-
0=off, 1=on
ground correction
5 Automatic
0=off, 1=on
calculation of
background tem-
perature
4 Process win-
0 = none, 1 = process
dow
window
3 lock (internal
use)
2 LED on
0 = off, 1 = on
1 Demo mode
0 = off, 1 = on
0 Camera or data
0 = data, 1 = live
screen on display
camera
Default
Alphanumerics
text
10
0 - 20mA
7 Display un-
0
0 = off, 1 = on
der-range values
6 CMD IN mode
0 = track & hold, 1 =
peak picker reset
5 Raw data
0 = off, 1 = on
4 Calibration
0 = off, 1 = on
current output
3 CMD OUT
0 = use for alarms (if
set), 1 = use as watch-
dog
2 CMD IN
0 = don't use, 1 = use
(according to CMD IN
mode)
1 NONC
0 = normally open, 1 =
normally closed
0 Analogue Mode
0 = 4 - 20mA,
1 = 0 - 20mA
0
0 - 5000 deg C or F
1600
0 - 5000 deg C or F
1600
0 - 5000 deg C or F
0
0 - 5000 deg C or F
70
0 - 70 deg C or F
0
0 - 70 deg C or F
4 Low Signal %
0
0 for off, 1 for on
Access Notes
rw
rw
rw
rw
rw
rw
rw
Use bit 2 of register 134 to switch the LED on and off as in example 5 below - this bit of register 14 is just an indicator of LED
current state.
rw
In demo mode, the LED is switched on when any button is pressed on the keypad, and remains on for 15 minutes rather than
30s.
rw
rw
A user-configurable name
rw
Set bit 12 of register 19 to switch on the fixed calibration current of this value. It will switch off automatically after 15 minutes
rw
rw
rw
The bit will be set to zero and the calibration current switched off automatically after 15 minutes.
rw
The watchdog setting toggles the CMD Out line every second.
rw
rw
rw
rw
rw
rw
rw
rw
rw
All bits can be set independently. If multiple bits are set, an alarm will be raised if any threshold is exceeded. Note that low sig-
nal% alarm is only available in ratio modes.