Millar Mikro-Tip Catheter Pressure Transducer Instructions For Use Manual

Advertisement

Quick Links

World Headquarters
Millar Instruments, Inc.
6001-A Gulf Freeway
Houston, Texas 77023-5417 USA
Phone: 832-667-7000 or 800-669-2343 (in the USA)
Fax:
832-667-7001
Email: info@millarmail.com
Web site: www.millarinstruments.com
Millar Worldwide Distribution
Millar Instruments, Inc. has a network of Authorized Distributors in most countries around the world.
For information on the Millar distributor in your country, please contact the Millar Customer Service
Department at our headquarters in Houston.
European Authorized Representative
®
Sensors.Systems.Solutions.
Harald Hellmann
DeMeTec GmbH
Co: FMI
Lützelwiesen 5
35428 Langgöns
+49-6403-7874-0
Fax -30
GF: Dipl. Phys. Harald Hellmann
AG Gießen: HRB-Nr. 2604
Email:
info@fmigmbh.de
®
Mikro-Tip
Catheter Pressure Transducer
Model SPC-721
Instructions for Use
OBSERVE PRECAUTIONS
FOR HANDLING
0086
ELECTROSTATIC
NONSTERILE PRODUCT
SENSITIVE DEVICES
Cardiology catheters MUST be cleaned and sterilized with ethylene oxide gas prior to every use.
Follow directions in this IFU.
CAUTION: Federal (USA) law restricts this device to sale by or on the order of a physician.
© 2008 Millar Instruments, Inc. All rights reserved
Millar, Mikro-Tip and Sensors.Systems.Solutions. are registered trademarks of Millar Instruments,
Inc.
Products and company names used are the trademarks or trade names of their respective companies.
Models referred to herein are protected by USA and International patents.
M.I. P/N: 004-2184 Rev. --

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the Mikro-Tip Catheter Pressure Transducer and is the answer not in the manual?

Questions and answers

Summary of Contents for Millar Mikro-Tip Catheter Pressure Transducer

  • Page 1 Web site: www.millarinstruments.com Millar Worldwide Distribution Millar Instruments, Inc. has a network of Authorized Distributors in most countries around the world. For information on the Millar distributor in your country, please contact the Millar Customer Service Department at our headquarters in Houston.
  • Page 2 Millar’s control, directly affect the product and the results obtained from its use, Millar shall not be liable for any incidental or consequential loss, damage, or expense arising directly or indirectly from the use of this product.
  • Page 3: Table Of Contents

    ) ........................... 3 RESSURE ENSOR ) ......................4 ONNECTOR ABLE .................... 4 RANSDUCER ERIFICATION AND ETUP OPERATING INSTRUCTIONS WHEN USING A MILLAR PRESSURE CONTROL UNIT .. 4 OPERATIONAL NOTES ........................5 ....................... 5 HONOCARDIOGRAM ECORDING ..............6 ANDLING RECAUTIONS FOR IKRO ATHETERS ..............
  • Page 4: Recommended Accessories

    Recommended Accessories Schematics M.I. P/N: 851-5918, Model TC-510 Control Unit, No patient isolation M.I. P/N: 880-0129, Model PCU-2000 Control Unit with Patient Isolation M.I. P/N: 850-1118, Model TEC-5C Extension Cable M.I. P/N: 850-1108, Model TEC-10C Extension Cable M.I. P/N: 850-1308, Model TEC-10D Extension Cable M.I.
  • Page 5: Contraindications

    • Disconnect the Mikro-Tip catheter from the Millar Control Unit prior to defibrillation or electrosurgery. • DO NOT use the Mikro-Tip catheter in close proximity to high electrical noise-generating...
  • Page 6: Precautions

    2.0 ± 0.5 inHgA/min. (6.6 ± 1.7 kPa) Rate: Aeration (Hot Cell) CAUTION: If damage is found during inspection, DO NOT use the catheter. Contact Millar Duration: At least 8 hours Instruments or authorized distributor. 110 ± 10 ºF (43 ºC)
  • Page 7: Connector ( S ) And Transducerv

    Apply a known pressure to the Millar catheter and verify the signal at the monitor. drying the catheter. Follow the instructions for the Millar pressure control unit being used. Set up the mercury manometer CAUTION: Use only the listed cleaners for the times/temperatures indicated.
  • Page 8: Cleaning

    Corrective Action Excessive Drift Deposit of foreign material on the Follow Cleaning Instructions. If diaphragm of the pressure sensor. problem persists, contact Millar. Transducer will Moisture in the connector, damage to Follow Operating Instructions or not balance wires in the catheter, or fractured...