Summary of Contents for RGB Omicrom FT
Page 1
Portable Vital Signs Monitor Omicrom FT Service Manual RGB Medical Devices, S.A. C/ Alfonso Gómez, 42 - 28037 MADRID (SPAIN) Cat. I101-7305 - Rev. 21 September 1998...
Page 2
RGB MEDICAL DEVICES. In order to maintain and improve standards of manufacturing, methods of functioning and to increase reliability, RGB MEDICAL DEVICES equipments are periodically reviewed. For this reason, the contents of this publication are subject to...
OMICROM FT SERVICE MANUAL LIST OF CONTENTS LIST OF CONTENTS ................i 1. INTRODUCTION ................1.1 1.1. Quality, Reliability and Safety..............1.1 1.2. General Precautions.................1.2 2. SERVICE INFORMATION ............. 2.1 2.1. Monitor and accessories cleaning............2.1 2.2. Technical specifications................2.2 2.2.1. Classification according IEC601.................2.2 2.2.2.
Page 5
OMICROM FT SERVICE MANUAL 4. DISASSEMBLY AND REASSEMBLY .......... 4.1 4.1. Back Cover Removal................4.1 4.2. Battery .......................4.2 4.3. Front cover....................4.2 4.4. Front Panel....................4.4 4.5. Measurement boards fastening..............4.6 4.6. NIBP Board (P000PLAC01 Internal Code) ..........4.8 4.7. Pump (C000BOMB01 Internal Code)............4.9 4.8.
The technical qualified personnel must read carefully and thoroughly this manual before performing any disassembly or maintenance procedure The family of Omicrom FT monitors have several models that monitors different physiological signs. All models have a common base that contains a metallic chassis, an external plastic enclosure and two common boards that provide power supply and communications for the rest of boards.
All service operations must be carried out by qualified personnel. Besides, these operations can only be performed by authorised personnel by RGB MEDICAL DEVICES. Before cleaning the equipment, always switch it off and disconnect it from the AC power supply.
Omicrom FT monitors is included. 2.1. Monitor and accessories cleaning. The Omicrom FT monitor and accessories must be kept clean on a routine basis, in accordance to the instructions provided here. To clean the monitor or their accessories it is necessary to disconnect the monitor from the AC Power and the different accessories from the monitor.
In this case, it is mentioned the code of the board that monitors the vital sign, because the specifications can vary according to the board used. The cross reference of the boards used into each model of Omicrom FT is included in Chapter 5 (“Schematics and spare parts list”).
OMICROM FT SERVICE MANUAL • IEC601-2-34(1994. Medical electrical equipment. Part 2: Particular requirements for the safety of direct blood pressure monitoring equipment. 2.2.3. Environmental Specifications. Storage temperature: 20°C to 60°C Operating temperature: 0°C to 40°C Relative humidity: 5 % to 90%, non condensing...
OMICROM FT SERVICE MANUAL 2.2.6. Respiration specifications: Board P000PLAC20 - “ECG-R 280598 Rev. 03” Patient Safety: Class I, type CF, protected against defibrillator (according IEC601). Respiratory Rate: Range: 0-150 breaths per minute. Resolution: 1 breath per minute. Accuracy: 2 breaths per minute.
OMICROM FT SERVICE MANUAL 2.2.8. Invasive Arterial Pressure (PRES) Specifications. Board P000PLAC01 - “TENSIOMETRO 2030594 Rev. 01” Patient Safety: Class I, type CF, protected against defibrillator (according IEC601). Number of channels: One. Measurement range: -15 to 250 mmHg Bandwidth: 0 to 15 Hz Transducer sensitivity: 5 µV / V / mmHg...
OMICROM FT SERVICE MANUAL 2.2.11. Temperature (TEMP) Specifications. Board P000PLAC04 - “CONTROTEMP 200695 Rev. 30” Patient Safety: Class I, type CF, protected against defibrillator (according IEC601). 15 to 45 °C / 59 to 113 °F Measurement Range: 0.1 °C / 0.2 °F Resolution: ±0.1 °C (it is not included the probe tolerance).
OMICROM FT SERVICE MANUAL 2.2.14. Operating Specifications. Sweep Speed: The sweep speeds available are 12.5 mm/s and 25 mm/s. Information by display: The display can monitor the waveforms of ECG, NIBP, , invasive pressure and respiration. The display shows the numeric value of each measured parameter.
Page 15
OMICROM FT SERVICE MANUAL Configuration Options: Configurations options available depend on each model and can be stored to be used the next time that the monitor is turned on: • Sweep speed: 12.5 or 25 mm/s. • QRS beep: Enabled or disabled.
In this section it is described the verification associated with each signal. Only it is necessary to carry out those verifications, according the Omicrom FT model (so each model monitors several signals). To verify each signal only it is necessary to activate the signal at the monitor.
Page 17
OMICROM FT SERVICE MANUAL 21 September 1998 2.10...
Page 18
ECG modes). b) Select at simulator QRS detection signal of 0,5 mV, 40 ms and 80 bpm. Select at Omicrom FT as Patient type neonates and check that the heart rate showed at monitor display is correct (all ECG modes).
Page 19
Disconnect the three leads and check that the same message is shown. d) With all cables and leads connected, select at simulator a normal QRS with an amplitude of 1 mV and at the Omicrom FT the 1mV ECG scale. Check that Omicrom FT shows an alarm message indicating “Signal saturated”.
Page 20
Pressure measurement: a) Select at simulator a static pressure of -10 mmHg. b) Select at the Omicrom FT a negative pressure scale and check that the pressure values showed at Omicrom FT display are correct. c) Select the maximum scale at the Omicrom FT.
Page 21
The actual heart rate can be measured manually over the wrist. Alarm system verification. a) Do not clip the sensor on to the finger and check that Omicrom FT shows an alarm message indicating “Sensor off patient”. b) Place an opaque means between the emitter and the receiver of the sensor and check that Omicrom FT shows a message indicating “Insufficient light”.
Page 22
SERVICE MANUAL Non invasive pressure signal verification. Activate the non invasive pressure signal at the Omicrom FT (for some models, it is necessary to disactivate previously the invasive pressure signal) and activate the displaying of the oscillating pressure at the cuff.
Page 23
OMICROM FT SERVICE MANUAL c) Change the non invasive pressure alarm levels at Omicrom FT to check that the patient alarms are being activated correctly. Verification of auxiliary operating aspects. This section includes a general verification of some characteristics of the Omicrom FT.
Omicrom FT. 2.4.1. Initial self-test. When the Omicrom FT is powered on, it is made a self-test for some seconds. If some failure is detected at anyone of the measurement boards it is indicated at the display and the relating measurements are not available.
OMICROM FT SERVICE MANUAL 2.4.3. SpO messages. During the normal operation of Omicrom FT monitor, it can show several alarm messages regarding the SpO signal (P000PLAC19 board). These messages are: 1) INTERNAL FAILURE The self calibration has failed or it is not possible to communicate with the SpO board.
OMICROM FT SERVICE MANUAL 2.4.4. NIBP messages. During the normal operation of Omicrom FT monitor, it can show several alarm messages regarding the NIBP signal (P000PLAC01 board). These messages are: 1) INTERNAL FAILURE The self calibration has failed or it is not possible to communicate with the NIBP board.
The list of accessories can vary depending on the Omicrom FT models. This list does not include the optional accessories that may have been selected and which may also be included in the box. If an item is missing or damaged, call RGB MEDICAL DEVICES by mail, fax or telephone.
SERVICE MANUAL 3.2. Monitor installation. For normal use, Omicrom FT can be placed on a horizontal surface; four feet are attached to the base of the monitor to provide a firm placement. It can also be attached to a vertical or horizontal bar using the adapters that are supplied as optional accessories.
3.3. Adjustments and option jumpers. This section describes the different adjustments that can be made and the use of the jumpers placed at the several boards included into the Omicrom FT monitor. To make these operations it is necessary to disassembly the Omicrom FT according the instructions given in chapter 4.
OMICROM FT SERVICE MANUAL 3.3.2. Base board (P000PLAC07 Internal Code) J1,J2 J3,J4 The function of the jumpers in this board are: 1) Jumpers J1 and J2 are used to select the operation mode of 68HC11 microcontroller. Function Open Open Expanded multiplexed mode - Standard...
OMICROM FT SERVICE MANUAL 3.3.3. Display board (P000PLAC14 Internal Code) J1,J2 The function of the jumpers in this board are: 1) Jumpers J1 and J2 are used to select the operation mode of 68HC11 microcontroller. Function Open Open Expanded multiplexed mode (Standard)
OMICROM FT SERVICE MANUAL 3.3.4. Main board (P000PLAC04 Internal Code) J1,J2 The function of the jumpers in this board are: 1) Jumpers J1 and J2 are used to select the operation mode of 68HC11 microcontroller. Function Open Open Expanded multiplexed mode (Standard)
OMICROM FT SERVICE MANUAL 3.3.5. Pulsioximeter board (P000PLAC19 Internal Code) J3,J4 JP3,JP4 J1,J2 The function of the jumpers in this board are: 1) Jumpers J1 and J2 are used to enable the SPI external communication from the board and their standard position is open.
OMICROM FT SERVICE MANUAL 3.3.6. ECG and respiration board (P000PLAC20 Internal Code) J1,J2 JP2,JP3 The potentiometer P1 is used to adjust the notch filter (50 or 60 Hz). The signal can be watched in PT3 with an oscilloscope. The function of the jumpers in this board are: 1) Jumpers J1 and J2 are used to select the operation mode of 68HC11 microcontroller.
OMICROM FT SERVICE MANUAL 3.3.7. NIBP and IBP board (P000PLAC01 Internal Code) J6,J7 J1..J5 The function of the jumpers in this board are: 1) Jumpers J1 to J5 are used to enable the SPI external communication from the board and their standard position is open.
Page 37
OMICROM FT SERVICE MANUAL 21 September 1998 3.10...
OMICROM FT SERVICE MANUAL 4. DISASSEMBLY AND REASSEMBLY This chapter includes the instructions for the disassembly of the equipment. Unless otherwise stated at the text, reassembly is the reverse of the disassembly procedure. During the disassembly procedure, be sure to keep all screws, nuts, plastic spacers and other accessories for the posterior assembly of the monitor.
OMICROM FT SERVICE MANUAL 4.2. Battery a) Disconnect the connectors: the lower positive (1) and the upper negative (2). Assembly: Place the battery in the rear chassis with the negative connector (black) (2) in the top and the positive connector (red) (1) in the bottom.
Page 40
OMICROM FT SERVICE MANUAL b) Remove the two nuts (2 and 3) of the upper rims of the metallic chassis. Assembly: Place the ground wire in the left screw (2) and fasten it with one spring washer and one nut. Place one spring washer and a nut in the right screw (3).
OMICROM FT SERVICE MANUAL 4.4. Front Panel a) Disconnect the Interface-display cable (1), the keyboard cable (2), and the LED’s board cable (3). P000CABL50 C000CARC01 P000CABL20 b) Remove the four nuts and the spring washers that fasten the display board (P000PLAC14 Internal Code).
Page 42
OMICROM FT SERVICE MANUAL c) Remove the board by pulling out it from the four hexagonal metallic spacers. Assembly: Before placing the display board it is necessary to mount the four hexagonal metallic spacers (15 mm length). d) To separate the...
OMICROM FT SERVICE MANUAL e) To separate the LED’s board, remove the two nuts (3) and spring washers. Under this board there are two nylon nuts placed as spacers. Take out the keyboard cover removing the nuts (1) and disconnecting the ground wire (2).
Page 44
OMICROM FT SERVICE MANUAL b) Remove the nuts with the screws with its washers that are fastening the boards to the metallic chassis. For the Controtemp board (P000PLAC04 Internal Code) it is mandatory to use a nylon screw and a nylon nut.
OMICROM FT SERVICE MANUAL 4.6. NIBP Board (P000PLAC01 Internal Code) a) After performing the procedures of the previous section, disconnect the two cables that come from the pressure switch (1) and the pump (2). b) Disconnect the electrovalve tube that come from the pump and the upper tube of the lower Y.
OMICROM FT SERVICE MANUAL 4.7. Pump (C000BOMB01 Internal Code). a) To fasten the pump it is used a metallic clamp where the pump is placed by pressure. This metallic clamp is fastened to the chassis using a screw and a nut.
OMICROM FT SERVICE MANUAL b) Disconnect the AC/DC-Line Filter cable (2) and remove the two nuts (3) with its washers. Remove the board, bringing it out from the plastic hook spacer (4). Assembly: Replace the two spacers in the screws of the chassis (in the right side of the chassis - 3 -).
OMICROM FT SERVICE MANUAL 4.11. Power Line Filter a) Remove the two screws and the nuts with the spring washers. P000FILT01 b) Dismantle the ground wire unscrewing the ground nut. Assembly: Replace the ground wire in the chassis ground socket fastening it with a nut 21 September 1998 4.11...
This chapter includes the schematics of the printed circuit boards used into the Omicrom FT monitors and the spare parts list. The PCB codes are the internal codes used in RGB MEDICAL DEVICES and they must be used to order these boards as spare parts.
Page 50
• U9 PAL Code “SA-16X8-01”. P000PLAC04 BOARD. • U2 EPROM Code “M1150-10202 / 020798 Rev. 02”. A50 Omicrom FT model in spanish language. • U2 EPROM Code “M1150-20202 / 020798 Rev. 02”. A50 Omicrom FT model in english language. • U2 EPROM Code “M1150-30202 / 020798 Rev. 02”.
Page 51
OMICROM FT SERVICE MANUAL • U2 EPROM Code “M1121-10202 / 020798 Rev. 02”. A21 Omicrom FT model in spanish language. • U2 EPROM Code “M1121-20202 / 020798 Rev. 02”. A21 Omicrom FT model in english language. • U2 EPROM Code “M1121-30202 / 020798 Rev. 02”.
SERVICE MANUAL 5.2. Spare parts list. This section includes the spare parts list for all Omicrom FT monitors models. First of all, it is included the list of the more important spare parts that are not mounted in any board and, after that, it is included the list of components of each board.
OMICROM FT SERVICE MANUAL 5.3. Board schematics. This section includes the schematics of all the boards that can be used into Omicrom FT monitors. 21 September 1998 5.27...
OMICROM FT 5.28 SERVICE MANUAL 21 September 1998 5.3.1. P000PLAC01 Schematics.
Page 77
OMICROM FT 5.29 SERVICE MANUAL 21 September 1998 MICROPROCESADOR RED1 PRESION OSCILOMETRICA VAL1 VAL2 SSCB PROS MISO SWIT MOSI PRESION INTRAARTERIAL OFFSET DIGITAL SWIT PRIN MOSI S[0..7] S[0..7] PE[0..3] -12V +12V Vbom 220uF 220uF 220uF 220uF MISO MOSI SSCB RED1...
Page 78
OMICROM FT 5.30 SERVICE MANUAL 21 September 1998 MISO MOSI 100nF 100nF 100nF MC34064P 15pF 15pF 100nF 100nF 47nF RESET 12MHz 100nF SRAM 32Kx8 RELOJ CSRAM CSROM XTAL 74HC373 74HC00 RESET XIRQ 100nF ADRESS BUS 68HC11 EPROM 64Kx8 100nF D[0..7] DATA BUS D[0..7]...
Page 79
OMICROM FT 5.31 SERVICE MANUAL 21 September 1998 -12V 100nF BIT 8(LSB) VOUT U12A BIT 7 VOUT SENSE A BIT 6 VOUT SENSE B 5K11 BIT 5 AD648 BIT 4 BIT 3 100nF BIT 2 BIT 1(MSB) 100nF AD557 +12V...
Page 80
OMICROM FT 5.32 SERVICE MANUAL 21 September 1998 EXCITACION SENAL TRANSDUCTOR CALIBRACION AMPLIFICADOR DIFERENCIAL +Vex (Ganancia = -114 o -14) AISLAMIENTO DE LA PRESION 100nF P.INVASIVA U18A 100K 0.1% +12V 1 2 3 4 5 6 7 OP200 100nF 100nF...
Page 92
OMICROM FT 5.44 SERVICE MANUAL 21 September 1998 MISO MOSI 100nF 100nF 100nF MC34064P 15pF 15pF 100nF 100nF 47nF RESET 12MHz 100nF SRAM 32Kx8 RELOJ CSRAM CSROM XTAL 74HC373 74HC00 RESET XIRQ 100nF ADRESS BUS 68HC11 EPROM 64Kx8 100nF D[0..7] DATA BUS D[0..7]...
Page 93
OMICROM FT 5.45 SERVICE MANUAL 21 September 1998 -12V 100nF BIT 8(LSB) VOUT U12A BIT 7 VOUT SENSE A BIT 6 VOUT SENSE B 5K11 BIT 5 AD648 BIT 4 BIT 3 100nF BIT 2 BIT 1(MSB) 100nF AD557 +12V...
Page 98
OMICROM FT 5.50 SERVICE MANUAL 21 September 1998 100K PUNTO TEST 4,7uF L[0..6] U12A LM393 BC546 100K 100nF 332K 100nF 100nF PUNTO TEST 332K 332K GENERADOR INTERMITENTE 74HC4066 LEDS * NO COLOCAR T10 ALARMA MONTAR PUENTE ENTRE LOS BC546 PINES 1 Y 4...
Page 112
OMICROM FT 5.64 SERVICE MANUAL 21 September 1998 Z O N A A I S L A D A +2V5 +12Vg +12Vg JPAD50 100nF U10A 300K 1/2W AMPLIFICADOR DE INSTRUMENTACION JPAD50 AD648 CG90L 120p G = 12,3 DG509 -12Vg -12Vg...