  • Page 1 USER GUIDE Nano SPY Ref : 11816C...
  • Page 2: Table Of Contents

    ..........................6 ATTERY REPLACEMENT VI. M ............................6 AINTENANCE VII. C ............................. 6 HARACTERISTICS Compliance ................................9 VIII. S ..........................9 UITABILITY FOR USE IX. W ............................. 11 ARANTY ......................... 11 AINTENANCE CONTRACT XI. E ........................11 NVIRONMENTAL PROTECTION ©JRI...
  • Page 3: Introduction

    I. INTRODUCTION The Nano SPY is a digital probe pe capable of measuring 1 or 2 physical quantities (T or TH depending on the model) and transmitting by radio to monitoring software hosted on a Web platform through a Nano Link.
  • Page 4: B) Positioning

     For installations outside the enclosure, mount the units sufficiently high on the walls to avoid interference with obstacles and foot traffic.  Insofar as is possible, place the Nano SPY LINK in a central position relative to the measurement points.
  • Page 5: Use

    IV. USE The Nano SPY can only be used with the My Sirius software hosted on a Web platform and with a Nano SPY Link. See MySirius online help for Nano Spy configuration. Off state As delivered, the Nano SPY is turned off. It can neither emit nor receive signals.
  • Page 6: Battery Replacement

    Put the new battery  in place respecting the polarity  KEEP THE BATTERY AWAY FROM FIRE; DO NOT ATTEMPT TO RECHARGE OR SHORT-CIRCUIT IT ONLY USE BATTERIES* SUPPLIED BY JRI (REF : 11596) *Recommended batteries: Saft LS14500 type AA 3.6V 2250mAh VI. MAINTENANCE Clean the device with a soft cloth, either dry or slightly moistened with water.
  • Page 7 : adjustable from 5s to 24h Response time : ~ 2 min. to 90% of the variation Power source : 3,6V Lithium battery– 2-year autonomy depending on use Case : Polycarbonate – Food safe Dimensions : 63 mm x 42 mm x 25 mm ©JRI...
  • Page 8 Weight : ~ 60g (without probe) ©JRI...
  • Page 9: A) Compliance

    EN 12830 Yes: these devices must be verified regularly according to EN 13486 (recommendation is once per year) CE ERM EN 301 489 / EN 61000 / EN 61010 / EN 55022 / EN 300 220 VIII. SUITABILITY FOR USE ©JRI...
  • Page 10 ©JRI...
  • Page 11: Waranty

    The use of radiofrequency thus requires periodic monitoring by recognized specialists. It is for this reason that JRI has developed for you the maintenance contract. We simplify your procedures by offering you a fully-integrated solution. This global service offer includes both maintenance and a metrological service, ensuring the optimum functioning of your devices or of your installation.