Maintenance and Troubleshooting

Maintenance
If you believe your Saunders Cervical Traction device requires
maintenance, contact your authorized Empi Traction Dealer or
contact Empi at 1-800-862-2343.
Cleaning
All surfaces of your Saunders Cervical Traction device may be cleaned
with a mild detergent or disinfectant. Do not immerse unit in water. If
unit is transferred between different users, all surfaces should be
cleaned using standard cleaning techniques paying particular attention to
the components that come in contact with the body; we recommend
using a new head strap.
Storage
Your Saunders Cervical Traction device should be stored in its
protective case.
Troubleshooting
The Saunders Cervical Traction device features a very accurate,
sensitive gauge. The pump is specifically designed not to leak in the
"HOLD" position. If you notice a drop in the gauge reading when in the
"HOLD" position it is probably because you have relaxed or moved
slightly. If this happens, return to the "PUMP" position and pump back
to the desired force, then reset to "HOLD."
Proper operation of the Saunders Cervical Traction device requires that
the air system be virtually leak-free. If the traction force is not maintained,
there may be a leak somewhere in the system.
Questions to Ask:
• Is the pump in the "PUMP" position when pumping?
• Is the pump in the "HOLD" position when receiving treatment?
• Are you lying still? (Any movement will cause fluctuation.)
• Are you pumping slowly?
• Are all connections tight (i.e. hose cap, lens cap, good hose
connection at the cylinder?)
7