Mojave V3 Pre-Installation Manual

Mojave V3 Pre-Installation Manual

Dry vacuum system

Advertisement

Quick Links

Dry Vacuum System
Part Numbers V3, V5, V7, 2V3, 2V3CT, 2V5, 2V5CT, 2V7, 3V5 and 4V5
P R E - I N S TA L LAT I O N G U I D E
All pumps comply with NFPA 99C level 3 requirements.
All installations must conform to local codes.
System being installed:
q V3
q V5
q 2V5
q 2V5CT
Doctor:
__________________________________________________
Address:
__________________________________________________
Phone#:
__________________________________________________
Dealer:
__________________________________________________
Dealer Address:
__________________________________________________
(AS CHECKED)
q V7
q 2V3
q 2V7
q 3V5
q 2V3CT
q 4V5

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the V3 and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for Mojave V3

  • Page 1 Dry Vacuum System Part Numbers V3, V5, V7, 2V3, 2V3CT, 2V5, 2V5CT, 2V7, 3V5 and 4V5 P R E - I N S TA L LAT I O N G U I D E All pumps comply with NFPA 99C level 3 requirements.
  • Page 2 MT10 / MT12 Tank Stacked One V3, V5 or V7 Two V3, V5 or V7 Three Assembly 10 Gallon Continuum onto One V3, V5 or V7 Pump Pump Pumps Stacked V5 Pumps Stacked Width 13.5 in. (34 cm) 29 in. (74 cm) 25 in.
  • Page 3 (84 cm) (44 cm) (107 cm) 17 in. (43 cm) MT10 / MT12 10 Gallon Tank Dimensions V3, V5 and V7 Vacuum Pump Dimensions CT20 / CT22 Continuum Tank Dimensions 23 in. (58 cm) 50 in. (127 cm) 50 in.
  • Page 4 TREATMENT ROOM PLUMBING INSTALLATIONS SUB FLOOR INSTALLATION - OVERHEAD INSTALLATION - Notes Recommended system installation layout should be used Alternate system installation layout should be used only when 1. 10-foot Maximum Height from Main whenever possible. unable to use the sub-floor plumbing layout. Line to Tank.
  • Page 5 E. MASTER CONTROLLER ELECTRIC OUTLET - Master Controller requires a dedicated standalone 120V, 5 AMP or 220V, 5 AMP grounded receptacle. F. PUMP ELECTRIC SERVICE - Each Mojave pump is wired directly with a dedicated 220V, 20 AMP , single phase 60 Hz circuit. If Main Circuit panel is not located in equipment room, a disconnect box with approved ground is needed for each pump.
  • Page 6 SITE REQUIREMENTS Electrical V3 & V5 2V3 & 2V3CT 2V5 & 2V5CT Master Controller Voltage Rating Volts AC All pumps 120 / 220 220 Volts Single Phase AC, 50/60 Hz olts AC or 108/132 V Voltage Minimum/Maximum 198/242 Volts AC All pumps...
  • Page 7 Facility Piping from Operatory. 2” Flexible Coupling Connector V3, V5 & V7 Hose from Manifold Connected to Pump PUMP HOSE via Air Input Filter Gasket Connector CONNECTION DETAIL WITH RESPECT TO ELECTRICAL SHOCK, FIRE, MECHANICAL MEDICAL ELECTRICAL EQUIPMENT AND OTHER SPECIFIED HAZARDS ONLY IN ACCORDANCE WITH UL-60601-1, CAN/CSA C22.2 NO.601.1 66CA...
  • Page 8 EXHAUST VENTILATION REQUIREMENTS Heat HEAT EXHAUST CONNECTION NOTES Exchanger 1. VENT LINE - The exhaust vent line required for MOJAVE systems using Vent Pipe the Heat Exchanger and systems with the Heat Exchanger removed Exhaust Vent See Note 1. have different requirements.