APAR AR915 User Instruction

Temperature calibrator

Advertisement

APAR - SALES OFFICE
05-090 Raszyn, ul. Gałczyńskiego 6
Tel. +48 22 853-48-56, +48 22 853-49-30, +48 22 101-27-31
E-mail:
automatyka@apar.pl
Website:
www.apar.pl
USER INSTRUCTION
TEMPERATURE CALIBRATOR
AR915
Version 1.1.0
2015-06-12

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the AR915 and is the answer not in the manual?

Questions and answers

Summary of Contents for APAR AR915

  • Page 1 APAR - SALES OFFICE 05-090 Raszyn, ul. Gałczyńskiego 6 Tel. +48 22 853-48-56, +48 22 853-49-30, +48 22 101-27-31 E-mail: automatyka@apar.pl Website: www.apar.pl USER INSTRUCTION TEMPERATURE CALIBRATOR AR915 Version 1.1.0 2015-06-12...
  • Page 2: Table Of Contents

    -1 - Thank you for choosing our product. This instruction is intended to facilitate correct operation, safe use, and taking full advantage of the calibrator's functionalities. Before you start the device, please read and understand this instruction. In the event of any additional questions, please contact our technical adviser.
  • Page 3: Principles Of Safe Use

    -2 - 1. PRINCIPLES OF SAFE USE before you start to use the device, become familiar with the present instructions; ensure proper operating conditions, as specified in the user’s technical data (humidity, temperature) The device is designed so as to ensure an appropriate level of immunity to most interferences that may occur in industrial environments.
  • Page 4: Technical Data

    -3 - TECHNICAL DATA Operation mode (set from the keyboard) .... input (measurement) or output (simulation) Universal input/output (programmable with the parameter 0:StYP), measurement and settings range: RTD : - Pt100 (3- or 2-wire)..-100 ÷ 850 °C (factory setting of the sensor) - Ni100 (3- or 2-wire)..
  • Page 5: Dimensions Of The Housing And Description Of Connections And

    -4 - 4. DIMENSIONS OF THE HOUSING AND DESCRIPTION OF CONNECTIONS AND EXTERNAL ELEMENTS input/output connections LED diodes - operating mode indication (input - IN, output - OUT) LCD display keyboard Dimensions in mm DESCRIPTION OF CONNECTIONS termin description INPUT TC, mV 1-2-3 input Pt100, Ni100, Pt500, Pt1000, resistance, 3-wire...
  • Page 6: Button Functions

    -5 - 5. BUTTON FUNCTIONS Button Function - switches the calibrator on/off (press for more than 1 second) - sets the operating mode: input - IN (measurement) or output - OUT (simulation) - accesses the parameter configuration mode (press for more than 2 seconds in the measured/set value display mode), - shows the type of sensor set (in the measured/set value display mode) - edits the current parameter in the configuration mode (display of the parameter value),...
  • Page 7 -6 - Table 1.Configuration parameters Parameter name change - Parameter name readout/saving - Settings SET, Parameter value change - Name Description of parameter Parameter value and range of default user’s variability type of StYP Pt= Pt100, ni= Ni100 sensor rES= 10÷540 Ω-IN or 0÷3200 Ω Resistance (input and (max)-OUT...
  • Page 8: Setting Output Signal Values

    -7 - 7. SETTING OUTPUT SIGNAL VALUES By pushing in the simulation mode (OUT), you can change the set value by the preset step (parameter 9:StEP). By pressing SET and at the same time, you set the output to the lower of the following values: the upper limit of the simulated sensor or the value of parameter 8:LiHi;...
  • Page 9 -8 -...