Global Water WL16S Instruction Manual

Covers and data loggers wl16 series

Advertisement

Quick Links

Global Water
800-876-1172 • globalw.com
WL16 Data Logger
Covers WL16S and WL16U Data Loggers
Global Water
Instrumentation, Inc.
151 Graham Road
P.O. Box 9010
College Station, TX 78742-9010
T: 800-876-1172
Int'l: (979) 690-5560, F: (979) 690-0440
globalw@globalw.com
11/30/09
01-404
Publication Number 38300112
- 1 -

Advertisement

Table of Contents
loading

Summary of Contents for Global Water WL16S

  • Page 1 Global Water 800-876-1172 • globalw.com WL16 Data Logger Covers WL16S and WL16U Data Loggers Global Water Instrumentation, Inc. 151 Graham Road P.O. Box 9010 College Station, TX 78742-9010 T: 800-876-1172 Int’l: (979) 690-5560, F: (979) 690-0440 globalw@globalw.com 11/30/09 01-404 Publication Number 38300112...
  • Page 2: Table Of Contents

    Global Water 800-876-1172 • globalw.com Congratulations on your purchase of a Global Water WL16 Data Logger. This instrument has been quality tested and approved for providing accurate and reliable measurements. We are confident that you will find the logger to be a valuable asset to your applications. Should you require assistance, our technical staff will be happy to help.
  • Page 3: Checklist

    Inspection: a. Your Data Logger was carefully inspected and certified by our Quality Assurance Team before shipping. If any damage has occurred during shipping; please notify Global Water Instrumentation, Inc. and file a claim with the carrier involved. Warranty: III.
  • Page 4: Wl16 Specifications

    Storage Capacity: 81,759 Recordings for two analog inputs, battery voltage monitor and date/time stamp Communication Ports: WL16S: RS-232 4-pin circular connector WL16U: USB Type B Selectable Baud Rates: Auto Detect or fixed, rates at 9600, 19200, 28800, 38400, 57600, 115200...
  • Page 5 Global Water 800-876-1172 • globalw.com Sensor: Sensor Element: Silicone Diaphragm, Wet/Wet Transducer Range: Level: 0-3', 0-15', 0-30', 0-60', 0-120', 0-250' Optional Temperature: 32°-122°F (0°-50°C) Linearity and Hysteresis: ±0.1% FS Accuracy: Level: ±0.1% FS at constant temperature ±0.2% over 32° to 70°F range Optional Temp: Smaller of 0.5°F or ±1% of reading...
  • Page 6: System Requirements

    This will install the Global Logger application in the Program Files folder and automatically put the Global Logger icon on the desktop. Serial communication (WL16S Only): If you are using the RS-232 serial port to access the data logger, no further action is needed. You can start using the logger by opening the Global Logger application program.
  • Page 7 Global Water 800-876-1172 • globalw.com WL16U Only The following menus apply to the Windows XP operating system. 1) Do not let Windows search for software. Click “No, not at this time” when asked if Windows Update can search for software.
  • Page 8 Global Water 800-876-1172 • globalw.com WL16U Only 4) The Hardware Wizard searches for the driver. If Windows advises that the software is not certified, click: “Continue Anyway”. This only means that Microsoft has not licensed the software. Second party software is allowed and no harm will come to your computer.
  • Page 9: Direct Connection

    There are 3 ways to connect to the WL16 data Loggers, direct connection through the RS-232 serial port (WL16S only) or USB port (WL16U only), as well as dial out through an installed modem (requires WL16S only with external serial modem).
  • Page 10: Modem Connection

    COM port settings. VIII. Modem Connection (WL16S Only): Access to the WL16S can be made through dial-out to a remote modem by clicking the “Modem” option. The connection from modem to logger requires an RS-232 serial modem, USB modems are not supported as an interface to the logger;...
  • Page 11 Global Water 800-876-1172 • globalw.com WL16S Only Two baud rates are supported 9600 and 19200. Should you experience difficulty establishing a reliable connection at 19200, try the slower connection speed. Note that due to the slower baud rates and the additional handshaking performed during modem communication, The Global Logger software will operate significantly slower than during a direct connection.
  • Page 12 All options and features work the same but at a reduced communication speed. When connecting the WL16S Data Logger to an external modem, a special cable is required. Wiring diagrams are provided at the end of this manual and pre-wired cables are available from Global Water.
  • Page 13: Main Screen

    Global Water 800-876-1172 • globalw.com Main Screen: The main screen is where the status and option menus are accessed. The WL16 data logger will continue to log at normal intervals while connected to the Global Logger software. While connected, the logger will draw an addition current from the power supply of 5-10mA so it is advised that, if you plan to not access the logger for extended periods of time, you disconnect to increase battery life.
  • Page 14 Global Water 800-876-1172 • globalw.com Tool Bar: The tool bar at the top of the screen has 3 options; File, Action and Help. In this screen, the File option is used to exit the program only. Action allows a menu driven interface to the various functions and displays the hot-keys, which can be used for fast access.
  • Page 15 Global Water 800-876-1172 • globalw.com Records: This field indicates the number of data recordings in memory. Logging State: This shows whether the WL16 is currently logging or waiting to be started or stopped by the alarm timers. These features are described in detail later.
  • Page 16 Global Water 800-876-1172 • globalw.com Bar Graphs: The bar graphs show a visual representation of the most recent data reading taken by using the Sample Continuously or Get Sample functions. The upper and lower limits of the display are defined by the user as discussed in detail later.
  • Page 17: On-Line Help Files

    Global Water 800-876-1172 • globalw.com On-Line Help Files: Clicking on Help displays the on-line help files which contain all of the information contained in this manual. There are 3 ways to access the information. The first tab is Contents which accesses the table of contents for the manual. Clicking on any of...
  • Page 18: Get History

    Global Water 800-876-1172 • globalw.com Get History: Clicking the Get History button begins the downloading of the historical data into an information screen. This screen allows the viewing of the recorded data and also provides a path to export the data to an Excel compatible file (.CSV file). Note that the logger continues to log as data is downloaded.
  • Page 19: Setup Menus

    Global Water 800-876-1172 • globalw.com Save to File: Clicking this button allows the data to be exported to an Excel compatible .CSV file. Specify the path and the file name and click the Save button Options: This opens a menu which...
  • Page 20: General Setup

    A warning message will appear before this action is taken. Help: The Help menu provides access to the on-line help files and Global Water contact information as described previously. XIII. General Setup: In the General Setup menu, the logger name, sensor warm-up time, logging mode, logging interval and alarm times are entered.
  • Page 21 Only change this to a fixed baud rate if planning to use this logger in a telemetry application, such as using Global Water’s Global Access radio modem software. Once programmed in a fixed rate mode, you must set the baud rate in the opening connection menu of the Global Logger II software to this speed, or the logger will not communicate.
  • Page 22 Global Water 800-876-1172 • globalw.com Logarithmic Timing: Typically used in groundwater pump studies, this mode speeds up the sample rate over time. Click the “Logarithmic” option to select this mode. This a logarithmic approximation that steps up the sample rate: a.
  • Page 23: Analog Setup

    Global Water 800-876-1172 • globalw.com Alarm Times: Check the box associated with each alarm to enable that feature and enter the desired start or stop time. The logger will start at the next occurrence of the Alarm 1 time and stop at the next occurrence of the Alarm 2 time after the Alarm 1 timer has expired.
  • Page 24 Global Water 800-876-1172 • globalw.com The calibration of channel one to a water level sensor is discussed in the following example where it is not possible to calibrate the sensor over its full range. The water level sensor shown is designed to measure over the range of 0-15 feet but a water column of 15 feet may not be available.
  • Page 25 Global Water 800-876-1172 • globalw.com Disp High: This field sets the upper limit of the bar graph displayed in the main screen and provides a way of scaling the graph beyond the upper limit used to calibrate the logger. In this example, the upper calibration point was a water level of 8 feet but the maximum range of the sensor is 16 feet.
  • Page 26 Global Water 800-876-1172 • globalw.com Click on the Low EU field and enter the low engineering unit that will be used to calibrate the logger, not necessarily the lowest reading the sensor can output. In this example, the value is zero feet, which is the lowest calibration point and water level.
  • Page 27: Programming

    Global Water 800-876-1172 • globalw.com XV. Programming Program Settings: The Program Settings button at the bottom of the setup menus will cause all the setup and calibration information to be programmed into the data logger. A warning will appear requiring the use to confirm the action.
  • Page 28: Modem Cable Diagram

    800-876-1172 • globalw.com XVI. Modem Cable Diagram (WL16S Only): The following wiring diagram is used to make a cable to connect the WL16S Data Logger to an external modem. A pre-wired cable is available from Global Water. - 28 -...
  • Page 29: Adjusting The Zero Reference Level

    Global Water 800-876-1172 • globalw.com XVII. Adjusting the Zero Reference Level: The sensor on the WL16 level logger measures the depth of water from the end of the sensor to the surface. However, it is sometimes desirable to know the water depth from a reference level other than the surface.
  • Page 30 Global Water 800-876-1172 • globalw.com • Install the WL16 as it will be in the final installation. • Place a mark on the cable or logger housing at the point where it meets the top of the well pipe (or other desired reference level).
  • Page 31 Global Water 800-876-1172 • globalw.com Assume that you wish to set the zero reference point to the lowest water level in the data record. This will show this low level as zero and all others as a positive level above it.
  • Page 32: Maintenance

    Verify that good batteries are installed. b. Check the communication cable. c. If the connection is to the WL16S using the serial cable provided by Global Water, verify that the correct COM port is selected in the connection menu of the Global Logger software.
  • Page 33: Customer Support

    Global Water 800-876-1172 • globalw.com XX. Customer Support: Call Global Water for tech support: 800-876-1172 or 979-690-5560 (many problems can be solved over the phone). Fax: 979-690-0440 or Email: globalw@globalw.com. When calling for tech support, have the following information ready: 1.