Page 2
Declaration of conformity INTEGRA Biosciences AG – 7205 Zizers, Switzerland declares on its own responsibility that the device Description Model FIREBOY eco 144010 complies with: EU Directives (DoW: Date of Withdrawal) Before DoW After DoW Low Voltage Equipment 2006/95/EC 20.04.2016...
This symbol cautions about potential material damage or data loss in a microprocessor controller. Follow the instructions. This symbol identifies important notes regarding the correct operation of the device and labor-saving features. 1.1.2 Hazard warnings on the device TTENTION Risk of burns from the flame. www.integra-biosciences.com...
INTEGRA Biosciences, the protection provided by the FIREBOY eco may be impaired. FIREBOY eco may not be altered in design or in any of its safety aspects. If alterations are made, INTEGRA Biosciences will be relieved from all liability for any resulting damage. Safety features of the FIREBOY eco The FIREBOY eco is equipped with the following safety devices: •...
GLP, GMP, FDA, must be observed. Please visit our website www.integra-biosciences.com on a regular basis for up to date information regarding REACH classified chemicals contained in our products.
Page 8
• rotating it and using it upside down • using it as a cooker without the optionally available long burner head, see “7.1 Accessories” on page 22 • carrying it by the burner head because the device may fall down INTEGRA Biosciences AG...
3 Red LED 4 GAS KNOB: to adjust the gas supply (flame size). 5 AIR KNOB: to adjust the air supply (flame temperature). 6 PUSH BUTTON: to switch the device on and off and for the C operating mode. ONTINUOUS www.integra-biosciences.com...
Page 10
9 Electrode of the flame monitor Rear view 10 Switch connection (jack) 11 Mains plug socket (mains plug) 12 Gas adapter opening (for engaging) Bottom 13 Folding stand, to position device at an angle 14 Brass screw, access to the nozzle INTEGRA Biosciences AG...
FIREBOY eco may only be operated with the nozzle that corresponds to the gas type in use. Only use the nozzles supplied. Prior to every nozzle installation, inspect the O-ring in the brass screw for damage and if necessary replace it. www.integra-biosciences.com...
Page 12
Insert the requested new adapter until it audibly clicks into position. The adapter is now replaced. INTEGRA Biosciences AG...
Page 13
Then open the gas cartridge adapter: Turn the milled knob fully counterclockwise in the "ON" direction. The gas cartridge is now open and ready for use. The flame can now be ignited using any of the three operating modes. www.integra-biosciences.com...
Page 14
(-), rotating it counterclockwise increases it (+): The AIR KNOB (air) can be used to regulate the air supply and therefore the flame temperature. Rotating the knob clockwise reduces the air supply (-), rotating it counterclockwise increases it (+). INTEGRA Biosciences AG...
AUTION The minimum distance between the tip of the burner head and the object to be heated, e. g. cover plate of quadruped, must be at least 60 mm. The duration of continuous operation should not exceed 60 min. www.integra-biosciences.com...
PUSH BUTTON ONTINUOUS Alternatively, FIREBOY eco can be ignited and also extinguished by pressing the PUSH BUTTON. The maximum burning time is 60 min. ARNING Never let the burner burn unattended! INTEGRA Biosciences AG...
The warnings and notes are displayed via the red LED. In the event of an error, the flame is automatically extinguished. Every message must be confirmed by pressing the PUSH BUTTON. The following table lists the possible errors and the recommended action for elimination. www.integra-biosciences.com...
Page 18
Electrode of the flame monitor is flashing slowly flame monitor. is kinked or dirty. (every 3 sec), Visually inspect, if necessary green LED is on. clean Red LED is on, Technical error. Call service technician green LED is off. INTEGRA Biosciences AG...
Do not spray directly on the instrument but use a lint-free cloth, lightly soaked with a disinfectant and wipe dry directly after decontamination. Never use acetone or other solvents! Follow the instructions provided by the disinfectant manufacturer. www.integra-biosciences.com...
FIREBOY eco - Operating instructions V07 Maintenance Disposal The FIREBOY eco device must not be disposed of with unsorted municipal waste. Dispose of the FIREBOY eco in accordance with the laws and regulations in your area governing disposal of devices. INTEGRA Biosciences AG...
Nominal thermal rating Propane/butane gas (liquid gas) 2.0 kW Natural gas E (~94% methane) 1.0 kW Natural gas LL (~81% methane) 0.8 kW Nominal gas input pressure Propane/butane gas (liquid gas) 50 mbar Natural gas E, LL 20 mbar www.integra-biosciences.com...
Gas cartridge adapter for 230 g butane 144 052 cartridge, e.g. Campingaz CV 300 Plus, with safety stand and connecting hose. Foot switch and connecting cable. 143 200 Benchtop switch and connecting cable. 171 081 Gas cartridge adapter for EXPRESS 444. 144 053 INTEGRA Biosciences AG...
Page 23
Mains adapter JP 100 VAC 156 634 Gas safety tubing Ø 10 mm. Length 2 m, 140 055 max. pressure 100 mbar, DVGW approved Adapter for gas hose, inner diameter 10 mm 144 240 Adapter for 1/4" left-hand thread 144 227 www.integra-biosciences.com...
Need help?
Do you have a question about the FIREBOY eco 144010 and is the answer not in the manual?
Questions and answers