Proprietary Notice The contents of this instruction manual are proprietary to Heath Consultants Incorporated. Reproduction of this manual, in whole or in part, is prohibited without the express written consent of Heath Consultants Incorporated. Heath Consultants Incorporated operates under a continual product improvement program and reserves the right to make improvements and / or changes without prior notification.
(methane) and propane are two most common gases typically odorized and distributed. Sample gas to be tested enters the ODORATOR 2 through the quick connect fitting at the gas supply inlet port. Internally, the sample gas passes through an internal five-micron filter to the input of a low-pressure regulator.
ANY SUSPICION OF A GAS LEAK BY UNEXPLAINED SOUND OR OPERATION OF THE ODORATOR 2 IS REASON TO IMMEDIATELY SHUT OFF THE GAS SUPPLY VALVES AND THEN SEEK THE CAUSE. IF INDOORS, DO NOT TURN THE ODORATOR 2 ON UNTIL A REFERENCE CGI INDICATES SA F E...
Page 7
ENVIRONMENT. LIKEWISE, ALLOW THE ODORATOR 2 TO PURGE FOR AT LEAST ONE MINUTE AFTER A REFERENCE CGI ALSO INDICATES SAFE ENVIRONMENT. DO NOT PERFORM TESTS IN CLOSED OR CONFINED SPACES FOR PERSONNEL SAFETY. PERFORM ALL MAINTENANCE IN A LOCATION KNOWN TO BE NON-HAZARDOUS.
ODORATOR 2 to sub-freezing temperature. If this condition occurs, as indicated by “LO FLOW” on the display, allow the ODORATOR 2 to warm up above freezing and then gradually expose it to sub-freezing temperature.
errors are detected, then “ERR” and an error number will be displayed and held. Optionally a “LO BAT” message and beep may appear. Instrument model type and model. Example: METHANE 1.00% User ID number if configured. Operator name if configured. If not configured, “NO USER” will display. Actual UTC Date.
1. Prior to attaching the sample hose, sniff the sample hose and verify no odor is detected. If odor is detected replace the sample hose. 2. Connect the sample hose to the gas supply outlet valve and to the ODORATOR 2 inlet port. Squeeze the inlet port lock to disengage while connecting or disconnecting.
Page 12
4. Open the gas supply outlet valve and the ODORATOR 2 Flow Metering Valve fully for about 10 seconds to purge the air out of the sample hose and condition the instrument. Do not sniff the exhaust while conditioning. 5. It is recommended to perform a Flow Self-Test at the start of each day of testing or when a test reading is suspect.
Page 13
“RDL” “HIGH” for 5 seconds along with an audible beep from the instrument to alert the operator of the RDL test result. The ODORATOR 2 will now go into PURGE mode and the display will show the PURGE icon, a flashing “ * “ and the live reading. The Flow Metering Valve must be closed while in PURGE mode and the live reading should be “0.00”.
Bluetooth hardware is installed and configured but connectivity is not supported. **NOTE** A future firmware and OdorSeer / OdorSeer Remote release will provide support. ODORATOR 2 Operator Controls and Connectors Figure 1. ODORATOR 2 Top Panel Layout 1. Accessory door/pull knob 7. Keypad 2. Gas inlet 8.
LCD with “LO BAT”, an alarm will beep and the instrument will go back to sleep. The ODORATOR 2 must be recharged in this case prior to use. In order to obtain full battery capacity, charge the battery pack when the ambient **NOTE** temperature is above 50 ºF (10 ºC) and below 85...
To charge the ODORATOR 2, plug the 18 Vdc power supply into an approved AC outlet and then connect the bayonet connector to the external power connector on the rear of the case. It will twist and lock into position.
3. If no odor is detected, turn the ODORATOR 2 “OFF”. The ODORATOR 2 has not absorbed the odorant and is usable. 4. If an odor is detected, turn the ODORATOR 2 “OFF” and send it to the nearest Heath Repair Center (Regional Office) listed on page 21. The odor is an indication that the ODORATOR 2 has absorbed the odorant and must have its internal tubing replaced and gas flow components cleaned.
The RMA process may be found on page 18 of this manual. Parts List The following list specifies ODORATOR 2 parts and accessories that are available from the factory or regional offices. HPN 104175 ODORATOR 2, complete Operator’s manual, ODORATOR 2...
Appendix A - Periodic Test Records A Periodic Test should be conducted every 30 days. The procedure for this test may be found on page 13 of this manual. The forms on pages 16 and 17 are provided for your convenience. They should remain in the manual and be kept in a safe location as a permanent record of periodic testing.
Page 20
Periodic Test Records Test Date Results Operator Name Action...
Page 21
Periodic Test Records Test Date Results Action Operator Name...
CHAPTER FOUR Warranty and Warranty Repair All instruments and products manufactured by Heath Consultants Incorporated are warranted to be free from defects in material and workmanship for one (1) year from the date of shipment. Furthermore, the warranty on authorized repairs in the Houston Factory Service Center (FSC) and other regions is ninety (90) days materials and thirty (30) days labor.
Return Procedure (RMA) The following steps will expedite the repair of your instrument: 1. Contact Heath Factory Service at 800-432-8487 to request a repair form. The form is also available on-line at http://heathus.com/info-center/repair-forms/ 2. Package carefully using the original shipping carton and cushions if available and return all components including the repair form.
Has been demonstrated to be in compliance with the European EMC directive (2004/108/EC), FCC 47 CFR, Part 15 and Industry Canada ICES-003 and is in accordance with the following standards: EN 61326-1: 2006 Date of Issue: March 28, 2014 Signed by: (On File) Jim Rutherford Sr. Vice President, Heath Consultants Inc.
REGIONAL SALES, CUSTOMER SERVICE AND INSTRUMENT REPAIR Eastern Region Heath Consultants Incorporated 575 Park Way Monongahela, PA 15063 Phone: 724-242-3145 Fax: 724-872-3206 Southwest Region Heath Consultants Incorporated 9030 Monroe Road Houston, TX 77061 Phone: 713-844-1300 Fax: 713-844-1309 Western Region Heath Consultants Incorporated...
Need help?
Do you have a question about the Odorator 2 and is the answer not in the manual?
Questions and answers
My older rater 2 was working fine then it shut off on me I charged it trying to turn it back on but it won't turn on
we have the oderator 2 and it is saying error 40 on the screen and its not allowing us to use the machine its