Installation & Calibration
Monitor Panel Dimensions
AFA500 or Remote Airfl ow Sensor Installation Diagram
Normal Sash
working height
Visit www.HEMCOcorp.com
HEMCO
HEMCO Corporation
711 S. Powell Road
Independence MO, 64056
Lab Planning Solutions
MM-TEL-AIRFLOW-500.07.14
Monitor Panel Cutout Dimensions
Fume hood
Fume hood
inner skin
outer skin
The monitor or sensor must be positioned
where it can 'see' the room pressure of the
laboratory. The back connection spigot of the
monitor or sensor will accept the 25mm (1")
hose which should be connected to the inner
chamber of the fume hood. The ideal position
for the hose connection is 100mm (4") back
from the sash and 100mm (4") higher than the
normal sash opening height.
Call Toll Free: (800) 779-4362
Phone (816) 796-2900
Fax. (816) 796-3333
R
Email info@HEMCOcorp.com
Installation, Operation, Maintenance Manual
TEL Airfl ow Monitor Model AFA 500 MK 3
Visit www.HEMCOcorp.com
HEMCO
HEMCO Corporation
711 S. Powell Road
Independence MO, 64056
Lab Planning Solutions
Call Toll Free: (800) 779-4362
Phone (816) 796-2900
Fax. (816) 796-3333
Email info@HEMCOcorp.com
R

Advertisement

loading
Need help?

Need help?

Do you have a question about the AFA 500 MK 3 and is the answer not in the manual?

Questions and answers

Summary of Contents for HEMCO AFA 500 MK 3

  • Page 1 Installation, Operation, Maintenance Manual Installation & Calibration TEL Airfl ow Monitor Model AFA 500 MK 3 Monitor Panel Cutout Dimensions Monitor Panel Dimensions AFA500 or Remote Airfl ow Sensor Installation Diagram Fume hood Fume hood Normal Sash inner skin outer skin...
  • Page 2 Model # Airfl ow Monitor Installation & Calibration AFA 500 MK3 STARTUP The AFA500 MK3 must be fi eld-calibrated once the room air supply and exhaust is balanced. When the unit is powered up, the following sequence of events occurs: 1.