1. Summary of Non-Contact Infrared Thermometer
The FDIR-V16 is a high technology medical thermometer for measuring eardrum and forehead
temperature of human body via in frared rays emitted from the eardrum and skin surface.
Following the correct way to measure is very important to obtain a precise temperature.
To ensure proper use, please be sure to read this user manual carefully.
In order to properly use this product, please carefully read the full text of this manual before using,
in particular the "safety precautions" section.
Please keep these instructions handy in a safe place for reference.
2. Safety precautions
All the materials used in the FDIR-V16 have passed the non toxicity testing of the
international regulations.
Warning
Use of this ther m ometer is not intended as a substitute for consultation with your physician. It is
dangerous for users to perform a self-evaluation and self-treatment based on the measuring result. Be
sure to follow doctor's instruction.
Keep the thermometer out of reach of children. If the battery or other component is swallowed
nent, please consult a doctor immediately.
Don't throw the battery into fire.
Notice
The device is a precision instrument, don't drop, tramp or impose any vibration or impact on the ther
mometer.
Do not touch the lens of the probe with your fingers, and do not dissemble the device by yourself.
Before measuring the temperature, make sure there is no hair or sweat in the way.
After you do some exercise, eating and bathing, you should stay still indoors about 30 minutes before
measurement.
To make the measurement data reliable and stable, when ambient temperature varies a lot, the ther
mometer should be placed indoors for about 30 minutes before using.
When we measure somebody continuously, the temperature should be measured every minute, if
you need to measure yourself continuously for a short time, there are some slight errors when you
read the temperature ,which is a normal phenomenon. At this time, we should choose the average.
There is no absolute standard about the temperature of the human, so please try to collect the re
cording of individual temperature in normal conditions, as a reference for having a fever or not.
Do not measure the sites of scarred tissue or tissue compromised by skin disorders, because sensing
body temperature from sites of scarred tissue or tissue compromised by skin disorders.
Do not measure the forehead temperature if the patient has trauma on forehead.
Do not measure when the measuring site (forehead) is exposed to direct sunlight, fireplace heat,
cold compress therapies, direct air conditioner flow. If you are in these environments please leave
and wait for 30 minutes to measure.
Do not immerse the device into water or any other liquid, and don't expose directly to sunlight.
In order to ensure the accuracy of measurement data, please don't take measurement of body tem
perature in strong electromagnetic interference environment (such as microwave, high frequency
equipment operation environment).
3. Intended use
Only use this product for the temperature measurement of ear and forehead. It can be used for anybody,
e.g. for new-born, for children and adults.
For safety reasons, measurement of children or baby's temperature must be measured by adults.
Children and babies should not operate the thermometer.
4. Temperature Measurement Modes and Range Description:
The infrared thermometer has the following measurement modes:
1) Forehead temperature measurement mode -- measures the skin surface of human forehead's
temperature accurately.
2) Eardrum temperature measurement mode—measures the skin surface of a humans eardrum
temperature accurately.
Normal temperature range for di erent measuring position:
Measuring position
Normal temperature (°C)
Anus
36.6-38.0
Oral
35.5-37.5
Armpit
34.7-37.3
Forehead
35.8-38.0
ear
35.5-37.8
Normal Forehead Temperature Range Based on Di erent Ages:
Ages
Normal temperature (°C)
0-2 years old
36.4-38.0
3-10 years old
36.1-37.8
11-65 years old
35.9-37.6
> 65 years old
35.8-37.5
NOTE: The normal temperature and di erence between the di erent body parts can vary between
individuals. To de ne your average temperature, measure your temperature for a least 2 weeks
in the same ear canal and forehead position.
NOTE: When consulting your physician, communicate the temperature that the thermometer
recorded and in what position and room environment as well as your normal body
temperature range as additional reference.
NOTE: Because the forehead temperature is a ected by the external environment (eg:
environment, air conditoning and color, etc),we advise that you take the forehead
temperature only as reference. When you have a doubt about the measurement
result, please use the ear temperature to con rm it.
5. Features
Two user design
The product has a two user mode, each user's test data is stored separately
.
High reliability
This product has passed the life and reliability test of the manufacturer, the time to failure
is≥1000h of use.
High accuracy
This product has passed the European Union infrared thermometer performance standards
for measuring clinical requirements, measuring clinical repeatability is no more than ±0.5°F
(0.3°C).
LED display
When the temperature exceeds the range, LCD will display the Lo or Hi prompt.
When the operating environment exceeds the design speci cations, the LCD will display the Err
prompt. When the thermometer battery power is insu cient, it has low voltage icon.
Has the hardware self-test function, when hardware malfunction is detected, the Err will display
prompts.
Power saving function
-
When the thermometer is not in use after approximately 30 seconds it will automatically enter standby
mode and the screen will display the time and environmental temperature alternately.
Memory storage function
The thermometer is able to record the previous 32 temperature recordings for your reference.
-
Backlight indication function
With the backlight indication function, it is easy for people to identify the temperature in a dark
environment.
Two-color LED indication function
In the forehead temperature measure mode:
If the body temperature is within 89.6°F -99.5°F (32°C -37.5°C), the Green LED will light up indicating
-
a safe body temperature.
If the body temperature is within 99.6°F -109.2°F (37.6°C -42.9°C), the Red LED will light up indicating
potential fever symptoms.
Flashlight function
The product is equipped with a ashlight, so you can take babies temperature at night time whilst
baby is sleeping.
6. Care and Cleaning
The measuring probe tip is the mode delicate component of the thermometer and it must
be clean and undamaged to ensure accurate readings. To clean the probe tip please follow
these instructions:
1. Very gently wipe the surface of the probe tip with a cotton swap or soft cloth
moistended with alcohol or an alcohol wipe.
2. After cleaning, inspect the probe tip to ensure that is it clean and undamaged.
3. If you nd the probe tip is damaged do not use the thermometer and contact
customer service.
4. Use a soft dry cloth to clean the thermometer LCD screen and surface of the body.
Do not use abrasive cleaners and never immerse the product in water.
7. Overall description
Normal temperature (°F)
97.9-100.4
95.9-99.5
Power/Scan Switch
94.5-99.1
96.4-100.4
95.9-100
Memory/User Switch
Flashlight Switch
Normal temperature (°F)
97.5-100.4
97.0-100.0
96.6-99.7
96.4-99.5
8. Operation Instruction
1) Replace the batteries to ensure e cient power supply if there is a low battery icon showing.
2) Ensure the sensor is clean. Refer to the section above, Care and Cleaning for details.
If the sensor's lens is damaged, please stop using it and contact Customer Care.
3) When you press the [Power/Scan] button, the system self tests the software and hardware.
If there are problems, the LCD will display "Err" symbol. Check if the sensor laser is dirty or
damaged.
4) To ensure an accurate measuring result, put the thermometer in the measurement
environment for 30 minutes so the thermometer is acclimatized.
5) Accuracy of unexpected uctuations in ambient temperature may decrease the measurement
results. When the thermometer displays di erent ambient temperatures in the same position it
may not be able to provide an accurate result. Be sure to take the temperature in an
environment where the room termpature is stable (ie not in front of an air conditioner).
6) Clean forehead and move hair aside. The forehead should be unobstructed and clean, in order
to ensure the accuracy of the measurement.
7) The probe tip is required to be cleaned after each use so it can be re-used. See chapter 6
above "Care and Cleaning". After ensuring the thermometer is clean you should always
store the thermomter in accordance with Chapter 9 "Storage".
【Instruction for use】
1) Before measurement begins.
Power on
For the rst use, put the battery inside the battery compartment. The product into the o state, press
the [Power /Scan] button to turn on. It will then boot as follows:
- Red backlight on, then Green backlight on, then boot LCD 2 seconds.
- Green LCD light turns on at the same time and the LCD shows last temperature
recording
- The thermometer then enters the measurement state.
(2) User selection
Press memory key for 3 seconds to select user 1 or user 2.
(3) Measurement unit selection
Press unit key to set measurement unit (°F or °C).
(4) Measure selected site
Put on the probe cover to switch to forehead mode or leave it o and and the thermometer enters
into the ear measuring mode.
When measuring the ear temperature, insert the probe into the ear first, then press the power
key. One second later you will hear "Beep", and you can remove the thermometer.
The measurement is completed.
Measuring probe
TEMPERATURE TAKING HINTS
(Sensor)
The right ear temperature may di er from the left ear. Therefore, always take the
temperature in the same ear.
Flashlight
The ear must be free from obstruction or excess earwax build-up to take an accurate reading.
External factors may in uence ear temperatures, including when an individual has:
- been lying on one ear or the other
- had their ears covered
- been exposed to very hot or very cold temperatures
Rating label
- been recently swimming or bathing.
In these cases, remove the individual from the situation and wait 30 minutes prior to taking a tem-
perature.
Use the untreated ear if prescription ear drops or other ear medications have been placed in
Battery Compartment
the ear canal.
Don't use the thermometer outdoors.
(2) Forhead temperature measure
Install the forehead cap, and press the scan button directly to measure the forehead temperature.
When measuring the forehead temperature, put the tip of the probe on your forehead and press the
scan button. The display will then show the tempreature reading in one second.
TEMPERATURE TAKING HINTS
Before take the temperature, please brush back any hair to prevent measurement deviation.
The sweat on head or cosmetic can a ect the accuracy of measurement. Please maintain the
cleanness of the forehead when measuring.
It is normal that there may be temperature di erence depending on various skin types and
color, since di erent skin type will re ect di erent voltage of infrared ray.
Don't use the thermometer outdoors.
(3) Temperature test information
If the temperature range is 89.6°F -99.5°F (32°C -37.5°C), the green backlight will display.
If the temperature is above 99.6°F -109.2°F (37.6°C~42.9°C), the red backlight will display.
If the measured temperature is higher than 109.2°F (42.9°C) , it will display " Hl".
If the measured temperature is below 89.6°F (32°C) , it will display "Lo".
If the ambient temperature outside the range of 50°F -104°F (10-40°C), the thermometer shows "Err";
4) Memory query mode
At the measure mode, Press memory key to enter measure memory query mode of the user
In the memory query mode, Press Memory/User key to move through previous measurements,
number will start from 1, 2,... 32 and the screen will show the corresponding number records.
In the memory query mode, without any action in 3s, it will turn off the current mode and return
to measure mode

Advertisement

loading

Summary of Contents for Cherub Baby FDIR-V16

  • Page 1 Be sure to take the temperature in an All the materials used in the FDIR-V16 have passed the non toxicity testing of the If the ambient temperature outside the range of 50°F -104°F (10-40°C), the thermometer shows “Err”;...
  • Page 2 Cherub Baby declares that the thermometer complies with following applicable standards: (2) Follow the polarity marking in the battery compartment, put the new batteries to the battery further details on the warranty and to make a warranty claim please In the forehead measurement mode: compartment and keep it fixed well.