How to use the sampler
CAUTION!
son. Take a new lancet for each sampling and
GlucoJect Dual PLUS
®
do not leave a used lancet in the sampler.
1.
Wash hands or
4.
Pull out the sliding
the alternative site
sleeve to lock the sam-
(AST). Unscrew the
pler. Press and hold the
sampler's tip by tur-
sampler against your
ning it counterclock-
fi nger and press the
wise.
trigger.
Press the lancet ejec-
5.
Insert a new
2.
tor with your thumb
sterile lancet into
while pulling out the
the lancet holder and
sliding sleeve for dis-
push it fully.
posing of the used
lancet.
Sampling AST
Adjust the puncture
3.
Place the sampler on the
depth by rotating the
skin at selected plug area.
tip. 1-2 for soft or thin
Press and hold for a few
skin, 3 for normal skin,
seconds, then press the
4-5 for thick or harde-
shutter release button
ned skin.
to take a blood test.
NEVER use sampler on more than one per-
Error codes and messages
ER 1
System error hardware. Remove and insert the batteries again and repeat the
test with a new test strip. If error code persists, please contact customer service.
ER 2
Test strip is damaged or used, or have the blood sample applied before
"blood drop symbol" began to fl ash on the screen. Repeat the test
with a new strip.
Er 3
Test strip was given the small amount of blood. Stick again to repeat
the test with a new test strip and the right amount of blood.
Er 4
Test strip was removed or dislodged during the countdown. Repeat
the test with a new test strip.
Er 5
NFC was disconnected during transmission. Repeat the NFC-transfer.
If the problem persists, contact customer service.
HI
Your test result is higher than 33,3 mmol/L!
LO
Your test result is lower than 1,1 mmol/L!
CAUTION! If HI or LO appears, contact your health care provider before
changing your treatment. Shown HI or LO repeatedly even though your
blood sugar levels are high or low, contact customer service.
Customer Service: 040-32 12 70 info@menarinidiagnostics.se
NOTE! This is a concise quick reference guide. Detailed information can be found in the manual supplied with the meter.
Medeon Science Park Per Albin Hanssons väg 41, hus D SE-205 12 Malmö
Tel 040–32 12 70 Fax 040–32 12 71 www.menarinidiagnostics.se
Quick Guide
Blood glucose measurement with wireless transmission
English