Shortwave diathermy device

Advertisement

Instructions for Use
The use of shortwave diathermy for heat therapy has the advantage of penetrating deeper than other conventional methods, such as
infrared lights, heating pads, and even microwave diathermy. The heat generated by shortwave diathermy induces a whole range of
physiological effects. It relaxes muscles, tendons and other connective tissues, and increases blood circulation in the treatment area.
When shortwave diathermy is applied in short, high-energy pulses, the depth of penetration is increased, having a particularly positive
effect on the blood circulation, while the temperature sensitive skin hardly feels the heat.
Replexa+ has been shown to be most beneficial when twice a day for 30 minutes per treatment.
An initial recommended setting has been pre-set by the manufacturer. For instructions to change the recommended setting, please
contact the ProMedTek service department at 866.388.2410 or your local dealer.
Turn on the Replexa+ Device
1.
Connect the Power cord to a grounded wall receptacle.
2.
Press the black POWER switch, located on the left side of the Device, to turn the Device on.
3.
The screen on the Device will light up and provide step-by-step instructions.
Connect Treatment Applicator to Device Port
1.
Place the Device close enough for the Treatment Applicator to easily reach the patient's treatment site – must be within 8 feet.
2.
Connect the Treatment Applicator cable to a Port on the front of the Device.
3.
The Port indicator will light blue if the Treatment Applicator is properly connected to the Device port.
The Treatment Applicator must be connected to the Device Port to begin treatment.
Position the Treatment Applicator
1.
Position the patient so they are resting comfortably and the treatment site is easily accessible.
2.
Place the Treatment Applicator directly on treatment site.
a.
The white side of the Treatment Applicator should face away from the treatment site. The black side of the Treatment
Applicator should touch patient.
b.
The center of the Treatment Applicator should be focused directly over the principal site of treatment.
c.
The weight of the body should not rest on the Treatment Applicator.
3.
Secure the Treatment Applicator to stay in place for the duration of the treatment.
Start Replexa+ Treatment
1.
Push the blue button to begin treatment.
2.
The Screen will indicate "Treatment in Progress" and display remaining treatment time.
Connecting or disconnecting a Treatment Applicator during therapy will stop the treatment.
Stop Replexa+ Treatment
1.
At the end of 30 minutes, treatment will automatically stop.
a.
The Screen will indicate "Treatment Completed".
b.
You may either press the POWER switch to power off the Device or press the blue button to begin another treatment.
2.
Although the Replexa+ treatment timer is set for 30 minutes, the clinician may manually discontinue treatment at any time by
pressing the blue button during therapy, which allows for treatment duration variability from 0 to 30 minutes.
a.
The Screen will indicate "Treatment Stopped".
b.
You may either press the POWER switch to power off the Device, or press the blue button to begin another treatment.
c.
The Device will always begin a new 30-minute treatment when started again.
3.
When treatment is completed or discontinued, turn off the Device by pressing the POWER switch.
4.
Remove the Treatment Applicator from the patient.
Cleaning and Storage
1.
Treatment Applicator
a.
Unplug the Treatment Applicator from the Device.
b.
Wipe clean the Treatment Applicator and cables with a non-staining, water-based disinfectant.
c.
Set aside that patient's Treatment Applicator for next use (single patient use only).
2.
Device
a.
Unplug the Power cord.
b.
Wipe clean the Device and Power cord with a non-staining, water-based disinfectant.
c.
Set aside for next use.
Troubleshooting
1.
If the screen indicates "Treatment Stopped," but does not indicate maintenance required, check to make sure the Treatment
Applicator is connected to the Device Port. Then press the blue button to begin treatment. If treatment does not begin, call the
ProMedTek service department at 866.388.2410 for further instructions.
2.
If the screen indicates "Maintenance Required," call the ProMedTek service department at 866.388.2410 for further instructions.
3.
There are no user-serviceable parts. All service needs must be addressed by ProMedTek.