IDEAL Networks FiberTEK III Manual

Fiber optic adaptor module

Advertisement

Quick Links

FiberTEK III

Advertisement

Table of Contents
loading

Summary of Contents for IDEAL Networks FiberTEK III

  • Page 1 FiberTEK III...
  • Page 2 All rights reserved Publication ref: 164802 Issue 4 - 02/2016 (Applies to LanTEK III software revision 3.067 onwards) IDEAL INDUSTRIES LTD. Stokenchurch House Oxford Road Stokenchurch High Wycombe Buckinghamshire HP14 3SX www.idealnetworks.net FiberTEK III Document 164802 Iss 4 User Manual Page 1...
  • Page 3: Safety Precautions

    If you have any questions concerning these precautions, the operating instructions, or any ® other concerns about the safe use and disposal of the battery packs used in LanTEK III, please contact a representative of IDEAL INDUSTRIES, LTD. FiberTEK III Document 164802 Iss 4 User Manual Page 2...
  • Page 4 Boldface Refers to a button on LanTEK III Cable Certifier. • Italics Refers to a menu option in this Operating Manual. • Quotation marks " " identify a "Screen Message". FiberTEK III Document 164802 Iss 4 User Manual Page 3...
  • Page 5: Table Of Contents

    Working with Jobs ........................19 4.4.5 Display of Result Details ......................... 19 Chapter 5. FiberTEK™ III Analyse Mode .................. 20 5.1.1 Continuous Run ..........................20 5.1.2 Power Meter Mode ..........................21 FiberTEK III Document 164802 Iss 4 User Manual Page 4...
  • Page 6: Chapter 1. Specifications

    100 to 999 nW 1.00 to 9.99 uW 10.0 to 99.9 uW 100 to 999 uW Length: 0.1m / 0.5ft Display update rate Approximately 2 updates per second on power meter function FiberTEK III Document 164802 Iss 4 User Manual Page 5...
  • Page 7 9/1310 or 0.75 ea/ OF-300 1550 1.80 1.5 total >35 50 or 0.75 ea/ OF-500 62.5/850 3.25 1.5 total >20 50 or 0.75 ea/ OF-500 62.5/1300 2.25 1.5 total >20 FiberTEK III Document 164802 Iss 4 User Manual Page 6...
  • Page 8 4.50 >26 Multimode- 0.75 ea/ 10GBase-SR 62.5/850 1.5 total 2.60 >20 Multimode- 0.75 ea/ 10GBase-SR 62.5/805 1.5 total 2.50 >20 Multimode/ 0.75 ea/ 10GBase-SR OM2-OM3 50/850 1.5 total 2.20 >20 FiberTEK III Document 164802 Iss 4 User Manual Page 7...
  • Page 9: Chapter 2. Product Description

    The adaptor modules are hot-swappable, ie. they can be changed without switching the LanTEK III power off. The two adaptor modules have an identical layout, as described below. 2.1.1 Controls, Ports, Indicators and Connectors Fig. 1 The Single-mode FiberTEK™ III module FiberTEK III Document 164802 Iss 4 User Manual Page 8...
  • Page 10 Do not look directly into the fiber end - it may be carrying a dangerous level of invisible light! FiberTEK III Document 164802 Iss 4 User Manual Page 9...
  • Page 11: Fiber Adapters

    The use of this conditioning unit introduces some Insertion Loss and therefore, before it is used, a Field Calibration must be performed, with the conditioning unit connected to the adaptor Tx port. FiberTEK III Document 164802 Iss 4 User Manual...
  • Page 12: Chapter 3. Calibration

    The Field Calibration should be performed using the same arrangement of jumpers as will be used in the tests. Use Arrow Keys to navigate on ready screen to " Field Calibration" display and confirm with Enter. 2. Select the required method: FiberTEK III Document 164802 Iss 4 User Manual Page 11...
  • Page 13 Tx and Rx Ports of the FiberTEK™ III . Fig. 5 Loopback Field Calibration (1-Jumper) FiberTEK III Document 164802 Iss 4 User Manual Page 12...
  • Page 14 FiberTEK™ III measuring adapters with an optical cleaning sponge. • Check launch cords with a fiber continuity tester (IDEAL order number: VFF5) for continuity. • Replace suspected damaged launch cords. FiberTEK III Document 164802 Iss 4 User Manual Page 13...
  • Page 15: Chapter 4. Testing Using Fibertek™ Iii

    NOTE The launch cord used for the field calibration must not be disconnected from the source otherwise the calibration is invalid. FiberTEK III Document 164802 Iss 4 User Manual Page 14...
  • Page 16: Set Autotest Pref

    Use Arrow Keys to navigate on ready screen to "LASER" display and confirm with Enter. 2. Use Arrow Keys to select the desired fiber typeset standard and confirm with Enter. FiberTEK III Document 164802 Iss 4 User Manual Page 15...
  • Page 17: Selection Of Refractive Index

    After selecting Fiber type, the cable properties screen is displayed. 2. Use soft key to select the fiber size. 3. Use soft keys to select test wavelength Fig. 12 Select Fiber size and wavelength FiberTEK III Document 164802 Iss 4 User Manual Page 16...
  • Page 18: Determine Loss Budget

    If you want to keep the offered Budget value, confirm with Enter, If not, use keypad to enter a number or use the soft key to open input box for loss budget calculator. Fig. 14 Loss Budget Calculator FiberTEK III Document 164802 Iss 4 User Manual Page 17...
  • Page 19: Evaluation Of Autotest Results For Fibers (Lwl)

    A name is automatically assigned to finished tests if Auto Increment function has been enabled. Use soft key to save the result using the default name. If another name is preferred, use option to manually assign another name. FiberTEK III Document 164802 Iss 4 User Manual Page 18...
  • Page 20: Manual Saving Of Autotest Results (Autosave Off)

    2. Mark the desired job. Press Enter to open the job. 3. Use Arrow Keys to mark the desired test record. 4. Press Enter to open the record. 5. Press Escape to return to previous screen at any time. FiberTEK III Document 164802 Iss 4 User Manual Page 19...
  • Page 21: Chapter 5. Fibertek™ Iii Analyse Mode

    There is no save function. In the Analyze screen select Continuous Mode. Fig. 16 Continuous Run (Ready) 2. Use soft key to start the test FiberTEK III Document 164802 Iss 4 User Manual Page 20...
  • Page 22: Power Meter Mode

    The reference level is shown by a black dot. Future power measurements are then shown in dB relative to the reference level. FiberTEK III Document 164802 Iss 4 User Manual...
  • Page 23 6. Use soft key to stop the test. 7. Use soft key to save the results. 8. Use soft key to clear the reference level and return to absolute power measurements in dBm. FiberTEK III Document 164802 Iss 4 User Manual Page 22...