However, Hach Ultra assumes no responsibility for any inaccuracies that may be contained in this guide. In no event will Hach Ultra be liable for direct, indirect, special, incidental, or consequential damages resulting from any defect or omission in this guide, even if advised of the possibility of such damages.
Operating Information (cont) Identification and Authorization Level The statistical view offers statistical data that matches with Total Quality management tools. Statistics is a tool to better analyze how a process behaves. The statistics are calculated from the data in the measurement file.
Page 4
Measurement Continuous Mode vs. Sample Mode • Stability: The stop criteria is met when the variation of the gas concentration is The continuous mode is typically used for process smaller than that entered in “Variation”. measurement, whereas sample mode is aimed at •...
Page 5
Measurement (cont) Measurement Configuration (TC Sensor) Measurement Filter Configuration The measurement configuration for a TC sensor is same as for an EC sensor with the addition of one extra selection criteria: Purge gas: From the drop-down list, select the purge gas being used for the TC sensor. Measurement Alarms Configuration T h e f i l t e r s a r e a i m e d a t “...
Page 6
Measurement (cont) Advanced Configuration (EC Sensor) • Liquid to gas factor: Enable correction as appropriate. • If checked, the percentage correction factor must be entered. This value cannot be negative. Note : To manually set the TC sensor into a continuous purge mode, press the Continuous Purge button that is available from the Services - Diagnostic - Channel x - Amplifiers menu.
Page 7
Measurement (cont) Interferences Configuration (TC Sensor) cycle. The current measurement files are updated in volatile memory, and are regularly copied in A number of different interfering gas options are non-volatile memory (file back-up). At start up, the available (temperature, H O, He, O and H measurement files in volatile memory are updated...
Page 8
Calibration General Modify Calibration Parameters We define 2 types of calibration: • the calibration of the gas to measure • the calibration of interferences To calibrate the gas to measure (main gas), the user usually puts the sensor in the main gas without any interfering gas.
Calibration (cont) Calibration Results Verification Similar to the calibration procedure, but for A calibration screen will be displayed showing verification of the actual calibration values. The current measurement data which is continually result of the measurements made during the refreshed. verification is not stored and the actual calibration The value “% ideal current”...
Page 10
Calibration (cont) Sensor Calibration t h e M o d i f y b u t t o n . T h e l a s t c a l i b r a t i o n parameters are memorized, so this step can be The sensor is either in contact with: ignored if the correct parameters are already set.
Page 11
Calibration (cont) Calibration Results Verification Similar to the calibration procedure, but for A calibration screen will be displayed showing verification of the actual calibration values. The current measurement data which is continually result of the measurements made during the refreshed. verification is not stored and the actual calibration The value “% ideal slope”...
Calibration (cont) Interference Calibration (EC Sensor) Interference Calibration (TC Sensor) The hydrogen interference calibration is available This calibration is available for TC sensors and is for EC sensors measuring O concentrations, and required when one of the interferences has been enabled.
Calibration (cont) Calibration Reports External Pressure Calibration (optional) Two calibration methods for the external pressure sensor can be selected: • Two point calibration (recommended) • One point calibration By default the two point calibration is selected. Two Point Calibration (recommended) Once a measured gas calibration is completed, a report is issued showing the log file of the last actions.
Instrument Maintenance Any instrument maintenance should be carried out by a qualified Hach Ultra Service Technician. Please contact your local representative should The instrument can automatically remind the user y o u f e e l a n y m a i n t e n a n c e o r i n s t r u m e n t when the next sensor service is due.
Maintenance and Troubleshooting (cont) Type Name Description Measure Normal measurement mode. Filter enabled The gas measurements are filtered. Sample measurement The sample measurement is started. Meas. not ready The measurement is not ready (e.g. at startup) Autotest in progress The autotest is running. Information Autotest failed The autotest has failed.