○This instrument has been designed and tested according to IEC Publication
○ The
○Following symbols are used on the instrument and in the instruction manual.
DIGITAL AC CLAMP METER
KEW SNAP
SERIES
KEW SNAP 2007A
KYORITSU ELECTRICAL INSTRUMENTS
WORKS,LTD.
6−2 AC Voltage Measurements
5.Preparation for Measurement
5−1 Checking Battery Voltage
  Set the function selector switch to any position other than "OFF".
When the
BATT
display is clear without "
" showing, proceed to measurement. When the
display blanks or "
BATT
" is indicated, replace the batteries according to the
instructions described in section 8. Battery Replacement.
NOTE
  (1) Set the function selector switch to the "400V" or "750V" position. "
The sleep feature automatically turns the instrument off in a certain period of
  (2) Plug the red test lead into the V terminal and the black test lead into the
time after the last switch operation. Therefore, the display may be blank with the
function selector switch set to a position other than "OFF". To operate the
instrument in this case, set the switch back to the "OFF" position, then to the
  (3) Connect the test lead prods to the circuit under test and take the reading
desired position, or press any button. If the display still blanks, the batteries are
exhausted. Replace the batteries.
5−2 Checking Switch Setting and Operation
  Make sure that the function selector switch is set to the correct position and
the data hold function is deactivated. Otherwise, desired measurement cannot
be made.
6.Measurement
6−1 AC Current Measurement
6−3 Resistance Measurement
#
WARNING
● This instrument is designed to work in distribution systems where the line to
earth has a maximum voltage of 300V. Be sure to use it within this rated
voltage.
● Transformer jaw tips are designed to minimize the possibility of shorting
conductors in the circuit under test. If equipment under test has exposed
conductive parts, however, extra precaution should be taken to avoid possible
  (1) Set the function selector switch to the "Ω/
shorting.
  (2) Plug the red test lead into the Ω terminal and the black test lead into the
●Do not make measurement with the battery compartment cover removed.
● Do not make current measurement with the test leads connected to the
instrument.
  (3) Check that the display reads "OL". With the test lead prods shorted
●Keep your fingers and hands behind the barrier during measurement.
  (4) Connect the test lead prods to the circuit under test and take the reading
  (1) Set the function selector switch to the "400A" or "600A" position.
  (2) Press the trigger to open the transformer jaws and clamp onto one
NOTE
conductor only.
  ◇ When shorting the test lead prods together, the display may show a very
  (3) Take the reading on the display.
Note:
  ◇If one of the test leads has an open, the display reads "OL".
  ◇ During current measurement, keep the transformer jaws fully closed.
Otherwise, accurate measurement cannot be made. The maximum
conductor size is 33mm in diameter.
  ◇ When measuring a larger current, the transformer jaws may buzz. This does
not affect the instrument's accuracy.
Correct
Wrong
1. Safety Warnings
61010: Safety Requirements for Electronic Measuring Apparatus. This instruction
manual contains warnings and safety rules which must be observed by the user to
ensure safe operation of the instrument and retain it in safe condition. Therefore,
read through these operating instructions before using the instrument.
# WARNING
● Read through and understand instructions contained in this manual before
starting using the instrument.
● Save and keep the manual handy to enable quick reference whenever
necessary.
● Be sure to use the instrument only in its intended applications and to follow
measurement procedures described in the manual.
● Be sure to understand and follow all safety instructions contained in the
manual. Failure to follow the above instructions may cause injury, instrument
damage and/or damage to equipment under test.
#
symbol indicated on the instrument means that the user must refer to
related parts in the manual for safe operation of the instrument. Be sure to
#
carefully read the instructions following each
symbol in this manual.
# DANGER
is reserved for conditions and actions that are likely to cause
serious or fatal injury.
# WARNING
is reserved for conditions and actions that can cause serious or
fatal Injury.
# CAUTION
is reserved for conditions and actions that can cause minor injury
or instrument damage.
Attention should be paid to each symbol to ensure your safety.
#
Refer to the instructions in the manual.
○Measurement categories (Over-voltage categories)
Indicates an instrument with double or reinforced insulation.
  To ensure safe operation of measuring instruments, IEC61010 establishes safety
Indicates that this instrument can clamp on bare conductors when
measuring a voltage corresponding to the applicable Measurement
category, which is marked next to this symbol.
Indicates AC (Alternating Current).
Indicates DC (Direct Current).
Indicates AC and DC.
Indicates Earth.
Crossed-out wheel bin symbol (according to WEEE Directive: 2002/96/
EC) indicating that this electrical product may not be treated as
household waste, but that it must be collected and treated separately.
# DANGER
● This instrument is designed to work in distribution systems where the line to earth
has a maximum voltage of 300V. Be sure to use it within this rated voltage.
● Do not attempt to make measurement in the presence of flammable gasses,
fumes, vapor or dust. Otherwise, the use of the instrument may cause
sparking, which can lead to an explosion.
● Transformer jaw tips are designed not to short the circuit under test.
  If equipment under test has exposed conductive parts, however, extra
precaution should be taken to minimize the possibility of shorting.
●Never attempt to use the instrument if its surface or your hand is wet.
●Do not exceed the maximum allowable input of any measurement range.
●Never open the battery compartment cover when making measurement.
● Never try to make measurement if any abnormal conditions, such as broken
Transformer jaws or case is noted.
7. Other Functions
# DANGER
7−1 Sleep Function
● This instrument is designed to work in distribution systems where the line to
earth has a maximum voltage of 300V. Be sure to use it within this rated
voltage.
●Do not make measurement with the battery compartment cover removed.
●Keep your fingers and hands behind the barrier during measurement.
This is a function to prevent the instrument from being left powered on in order to
conserve battery life. This function causes the instrument to automatically enter the
sleep (powered-down) mode about 10 minutes after the last switch or button
COM terminal.
operation.
To exit the sleep mode, turn the function selector switch back to "OFF", then to any
on the display.
other position, or press any button.
How to disable the sleep function:
Red lead
Black lead
7−2 Data Hold Function
  This is a function used to freeze the measured value on the display. Press the
# DANGER
●Always make sure that the circuit under test is powered off.
  To exit the Data Hold mode, press the Data Hold button again.
●Do not make measurement with the battery compartment cover removed.
●Keep your fingers and hands behind the barrier during measurement.
NOTE:
  ◇ If the instrument in the Data Hold mode enters the sleep mode, the Data
" position.
COM terminal.
8.Battery Replacement
together, also check that the buzzer beeps and the display reads "0" .
on the display. The buzzer beeps the reading is below about 50Ω.
small resistance instead of "0". This is the resistance of the test leads.
  When "
  (1) Set the function selector swith to the "OFF" position.
  (2) Unscrew and remove the battery compartment on the bottom of the
Black lead
Red lead
  (3) Replace the batteries observing correct polarity. Use two new R03 or
  (4) Mount and screw the battery compartment cover.
● The instrument is to be used only in its intended applications or conditions.
2.Features
Otherwise, safety functions equipped with the instrument doesn't work, and
instrument damage or serious personal injury may be caused.
●Tear-drop-shaped jaws for ease of use in crowded cable areas and other tight places
●Safety design conforming to the following provisions of IEC61010
# WARNING
   Measurement category III 300V, pollution degree 2,
   Measurement category II  600V, pollution degree 2,
● Never attempt to make any measurement if any abnormal conditions are
   Measurement category I
noted, such as broken case, cracked test leads and exposed metal part.
● Do not turn the function selector switch with plugged in test leads connected to
●Data Hold function for easy reading in dimly light or hard-to-read locations
●"Sleep" feature to extend battery life
the circuit under test.
● Do not install substitute parts or make any modification to the instrument.
●Beeper permits easy continuity check
●Provides a dynamic range of 4,000 counts full scale
Return the instrument to Kyoritsu or your distributor for repair or re-calibration.
●Do not try to replace the batteries if the surface of the instrument is wet.
●Uses shrouded transformer jaws to further improve safety
● Always switch off the instrument before opening the battery compartment
3.Specifications
cover for battery replacement.
● Verify proper operation on a known source before starting to use the
instrument or taking action as a result of the indication of the instrument.
●Measuring Ranges and Accuracy
 AC Current (
# CAUTION
● Make sure that the function selector switch is set to the appropriate position
before making measurement.
● Always make sure to insert each plug of the test leads fully into the appropriate
 AC Voltage (
terminal on the instrument.
● Make sure to remove the test leads from the instrument before making current
measurement.
● Do not expose the instrument to the direct sun, extreme temperatures or dew
 Resistance (Ω/
fall.
● Be sure to set the function selector switch to the "OFF" position after use.
When the instrument will not be in use for a long period of time, place it in
storage after removing the battery.
● Use a damp cloth and detergent for cleaning the instrument. Do not use
●Operating System
●Display
abrasives or solvents.
●Low Battery Warning
●Overrange Indication
●Response Time
●Sample Rate
standards for various electrical environments, categorized as CAT I to CAT IV,
●Location for use
and called measurement categories. Higher-numbered categories correspond to
●Temperature and Humidity
  for Guranteed Accuracy :23±5℃, relative humidity up to 85% without condensation
electrical environments with greater momentary energy, so a measuring
●Operating Temperature
instrument designed for CAT III environments can endure greater momentary
  and Humidity
energy than one designed for CAT II.
●Storage Temperature
CATⅠ: Secondary electrical circuits connected to an AC electrical outlet
  and Humidity
through a transformer or similar device.
●Power Source
CATⅡ: Primary electrical circuits of equipment connected to an AC electrical
●Current Consumption
outlet by a power cord.
●Sleep function
CATⅢ: Primary electrical circuits of the equipment connected directly to the
distribution panel, and feeders from the distribution panel to outlets.
●Standards
CATⅣ: The circuit from the service drop to the service entrance, and to the
power meter and primary over-current protection device (distribution
panel).
Incoming wire
●Location for use
Interior wiring
●Overload Protection
CAT.III
CAT.IV
CAT.II
●Withstand Voltage
Transformer
CAT.I
●Insulation Resistance
Socket
●Conductor Size
●Dimensions
NOTE
The instrument still consumes small amount of battery power in the sleep mode.
Make sure to set the function selector switch to the OFF position after use.
To disable the sleep function, power on the instrument with the Data Hold
9.Optional Accessories
button pressed. "P.OFF" is shown on the display for about 3 seconds after the
instrument is powered on.
●Model 8008 Multi-Tran
To enable the sleep function, power the instrument off, then power it on without
  Multi-Tran Model 8008 is designed to increase the measuring capability of a
pressing the Data Hold button.
  (1) Set the function selector switch to the "400A" position.
  (2) As shown in the figure, clamp KEW SNAP 2007A onto the pickup coil of
Data Hold button to freeze the reading. The reading will be held regardless of
subsequent changes in input. "
H
" is shown on the upper left corner of the
display while the instrument is in the Data Hold mode.
  (3) Clamp Model 8008 onto the bus-bar or conductor under test.
  (4) Take the reading on KEW SNAP 2007A and multiply it by 10.
Hold mode will be cancelled.
# WARNING
To avoid electric shock hazard, make sure to set the function selector switch to
"OFF" and remove the test leads from the instrument before trying to replace the
batteries.
# CAUTION
●Do not mix new and old batteries.
● Make sure to install battery in correct polatiry as indicated inside the battery
compartment.
BATT
" is shown on the display, replace the batteries. Note that when
the battery is completely exhausted, the display blanks without "
BATT
" shown.
instrument.
equivalent batteries.
●Weight
●Accessories
  )
●Optional Accessories
●Material of this product Material group I 600 ≦ CTI
750V, pollution degree 2,
4.Instrument Layout
A)
Range
Measuring Range
Accuracy
400A
 0∼399.9A
 ±1.5%rdg±4dgt ( 50/60Hz)
600A
 0∼599A
 ±2.0%rdg±5dgt ( 40∼400Hz)
V) (Input impedance: approx. 2MΩ)
Range
Measuring Range
Accuracy
400V
 0∼399.9V
 ±1.2%rdg±3dgt ( 50/60Hz)
750V
 0∼749V
 ±1.5%rdg±4dgt ( 40∼400Hz)
)(Auto-ranging)
Barrier:
It is a part providing protection against electrical shock and ensuring the minimum
Range
Measuring Range
Accuracy
required air and creepage distances.
400Ω/
 0∼399.9Ω
 ±1.5%rdg±2dgt
4kΩ
 0.150∼3.999kΩ
 (Buzzer beeps below 50±35Ω)
: Dual Integration
: Liquid crystal display (maximum count: 3999)
:"BATT" is shown on the display
:"OL" is shown on the display
:Approx. 2 seconds
:About 2.5 times per second
: Indoor use, Altitude up to 2000m
:0-40℃, relative humidity up to 85% without condensation
:-20-60℃, relative humidity up to 85% without condensation
:Two R03 or equivalent (DC1.5V) batteries
Use of our Protective Cap offers different lengths of the exposed metal part suitable
:Approx. 2.5mA.
for the test environments.
: Automatically powered down in about 10 minutes after the
Please attach the Cap onto the metal part under CAT.III or higher test environments.
last switch operation (power consumption in the sleep mode
is about 35 μA.)
:IEC61010-1 CAT Ⅲ 300V, pollution degree 2
            CAT Ⅱ 600V, pollution degree 2
The Cap should be firmly attached to the Probes.
           CAT Ⅰ 750V, pollution degree 2
Keep your fingers and hands behind the barrier during measurement.
IEC61010-031
IEC61010-2-032
 ●LCD INDICATOR
IEC61326(EMC)
In-door use, altitude 2000m or less
:AC current ranges: 720A AC for 10sec
AC voltage ranges: 900V AC for 10sec
Resistance ranges: 600V AC for 10sec
: 3470VAC(RMS,50/60Hz) for 5 sec. between electrical circuit
and housing case
: 10MΩ or greater at 1000V between electrical circuit and
housing case
:Approx. 33mm diameter max.
:195(L)x78(W)x36(D)mm
DISTRIBUTOR
Battery Compartment Cover
Screw
Battery
clamp meter. With the use of the Multi-tran, you can measure AC current up to
3000A and a large bus-bar or conductor.
Model 8008.
MAX 100
MAX
150
3-14
:Approx. 260g (including batteries)
:Test leads M-7066A
Two R03 batteries
Carrying case M-9097
Instruction manual
:Multi-Tran M-8008
Transformer Jaw
Barrier
Jaw Trigger
Function Selector Switch
Data Hold Switch
Display
Ω Terminal
V Terminal
COM Terminal
Drop Prevension Strap
Exposed metal part
Protective Cap
Barrier
Uncapped condition for CAT.II environment
Barrier
Capped condition for CAT.III/ IV environments
# CAUTION
Data Hold Mode
Voltage
Resistance
Current
Battery Voltage Alert
Kyoritsu reserves the rights to change specifications
or designs described in this manual without notice
and without obligations.
KYORITSU ELECTRICAL
INSTRUMENTS
WORKS, LTD.
No.5-20, Nakane 2̶ chome, Meguro-ku,
Tokyo, 152̶0031 Japan
Phone:81̶ 3̶3723̶0131
Fax:81̶ 3̶3723̶0152
URL:http://www.kew-ltd.co.jp
E-mail:info @ kew-ltd.co.jp
92-2204

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the KEW SNAP 2007A and is the answer not in the manual?

Questions and answers

ERIC DELROSARIO
February 25, 2025

how to calibrate this clamp meter?

Subscribe to Our Youtube Channel

Table of Contents