Sea-Bird Scientific SBE 33 Product Manual

Sea-Bird Scientific SBE 33 Product Manual

Carousel deck unit for use with pdim, sbe 32 carousel water sampler, or sbe 55 eco water sampler
Hide thumbs Also See for SBE 33:
Table of Contents

Advertisement

Quick Links

Release Date
01/30/18
Version
017
Firmware
3.0c & later
Software
Seasave V7 7.26.7 & later
service@seabird.com | support@seabird.com | Tel: +1 425 643 9866
Copyright 2017 Sea-Bird Scientific. All rights reserved.
SBE 33 Carousel Deck Unit
For use with PDIM, SBE 32 Carousel Water Sampler,
Product Manual
or SBE 55 ECO Water Sampler
Front Panel – switches & LEDs
Back Panel – connectors & fuses

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the SBE 33 and is the answer not in the manual?

Questions and answers

Summary of Contents for Sea-Bird Scientific SBE 33

  • Page 1 Front Panel – switches & LEDs Back Panel – connectors & fuses Release Date 01/30/18 Version Firmware 3.0c & later Software Seasave V7 7.26.7 & later service@seabird.com | support@seabird.com | Tel: +1 425 643 9866 Copyright 2017 Sea-Bird Scientific. All rights reserved.
  • Page 2: Limited Liability Statement

    Limited Liability Statement Extreme care should be exercised when using or servicing this equipment. It should be used or serviced only by personnel with knowledge of and training in the use and maintenance of oceanographic electronic equipment. SEA-BIRD ELECTRONICS, INC. disclaims all product liability risks arising from the use or servicing of this system.
  • Page 3: Declaration Of Conformity

    Manual revision 017 Declaration of Conformity SBE 33 Declaration of Conformity...
  • Page 4: Table Of Contents

    Mounting SBE 33 .................... 29 Wiring System ....................30 Water Sampler or PDIM to CTD .............. 30 Sea Cable from SBE 33 to Water Sampler or PDIM ........ 30 SBE 33 to Computer ................. 30 Power to SBE 33 ..................31 NMEA Navigation Device to SBE 33 ............
  • Page 5 Manual revision 017 Table of Contents SBE 33 Problem 4: Unsupported Modem Message (FFFFFFFF Unsupported modem message: XX XX XX) from Water Sampler........ 42 Problem 5: Yellow NMEA Transmit LED Not Flashing ........43 Problem 6: Yellow NMEA Transmit LED Flashing, but Lat/Lon Data Not Displaying ......................
  • Page 6: Section 1: Introduction

    This section includes photos of a typical SBE 33 shipment. About this Manual This manual is to be used with the SBE 33 Carousel Deck Unit. It is organized to guide the user from installation through operation. We’ve included detailed specifications, setup and operation descriptions, and helpful notes throughout the manual.
  • Page 7: Unpacking Sbe 33

    Manual revision 017 Section 1: Introduction SBE 33 Unpacking SBE 33 Shown below is a typical SBE 33 shipment. Inclusion of test cables, etc. is dependent on the order. SBE 33 Carousel Deck Unit SBE 33 power cable Rack mount kit...
  • Page 8: Section 2: Description Of Sbe 33

     The SBE 32 Carousel is a 12-, 24-, or The SBE 33 Carousel Deck Unit provides power and real-time data acquisition and control for an SBE 32 Carousel Water Sampler that has the SBE 33 36-bottle water sampler. The SBE 55 ECO is a 3- or 6-bottle water sampler.
  • Page 9 DPT – Depth compatible with NMEA depth data in Decoded NMEA data is appended to the CTD data stream in the SBE 33, and the data stream. are passed to the computer for storage and/or display with the CTD data.
  • Page 10 Manual revision 017 Section 2: Description of SBE 33 SBE 33 NOTES: 1. The SBE 32 Carousel Water Sampler can be used with several Sea-Bird CTD types, and can be controlled in several ways: Data Acquisition Control and Control SBE 9plus...
  • Page 11: Specifications

    SBE 33 Specifications The SBE 33 supplies a constant 250 VDC power to the sea cable. The Water Sampler or PDIM receives this voltage (minus the sea cable IR drop) and regulates it to a constant 64 VDC. This 64 VDC is input to a high-efficiency DC/DC converter, which outputs + 15 VDC.
  • Page 12: Pdim Dimensions And Connectors

    PDIM Dimensions and Connectors If deploying the CTD without a water sampler, a PDIM can be used to connect the CTD to the SBE 33. Order PDIM with desired type of bulkhead connector:  PN 90227 - PDIM with XSG connectors ...
  • Page 13: Sbe 33 Front Panel

    Bottle to Fire pushbutton switch - selects which bottle to fire when Fire  Bottles can also be fired from the button is pushed; if set to 99, SBE 33 will fire bottles in sequential order computer using Seasave real-time ...
  • Page 14: Cables And Wiring

    Manual revision 017 Section 2: Description of SBE 33 SBE 33 Cables and Wiring PN 171887 DB9 to DB9 cable (2) – SBE 33 Serial Data connector and Carousel Data connector to computer 2-pin to 2-pin Test Cable - SBE 33 Sea...
  • Page 15 Manual revision 017 Section 2: Description of SBE 33 SBE 33 Slip Ring Cable -- DN 31371 2-pin SBE 33 Sea Cable to slip ring 2-pin to DB9 Test cable - SBE 33 NMEA Input to computer...
  • Page 16 See the appropriate drawing for pinout details.  SBE 33 can also be used with a Satlantic (Sea-Bird Scientific) Surface PAR sensor. Wiring diagram for that cable is not currently available. 4-pin SBE 33...
  • Page 17: System Communications

     SBE 25: If the SBE 25 is integrated with more than 4 auxiliary voltage sensors, and the SBE 33 is appending both NMEA and Surface PAR, you must set the SBE 25 real-time output rate to 4 scans/sec or less (see the SBE 25 manual for command details).
  • Page 18: Nmea And Surface Par

    NMEA 0183 protocol. Decoded NMEA sending, see Section 7: data can be appended to the end of the CTD data stream in the SBE 33 and Troubleshooting for information on passed to the computer for storage and/or display with the CTD data (see viewing the raw NMEA message using the SBE 33’s diagnostic mode.
  • Page 19 Manual revision 017 Section 2: Description of SBE 33 SBE 33 The SBE 33 has three operating modes and two diagnostic modes, as well as Notes: two NMEA depth options and two Surface PAR options, and selections for  In Seaterm, use the @ setting and displaying communication parameters and start-up behavior.
  • Page 20: Section 3: Setting Up System

    Setting the baud rate in the CTD  Setting up and testing the SBE 33  Changing communication settings between the SBE 33 and computer  Setting up the CTD configuration (.xmlcon or .con) file Installing Software Seasoft V2 was designed to work with a PC running Windows XP service pack 2 or later, Windows Vista, or Windows 7.
  • Page 21: Setting Ctd Baud Rate And Operating Parameters

    SBE 19plusV2 or 25plus below. Connect the CTD directly to the computer (not through the SBE 33 and Water Sampler), using the data I/O cable supplied with the CTD. Double click on Seaterm.exe. If this is the first time the program is used,...
  • Page 22 Record the baud rate - you must also set the CTD baud rate  Once all baud rates are set, you can in the SBE 33 (see Setting Up and Testing System). also communicate with and change the setup of the CTD through the 10.
  • Page 23: Sbe 19Plus V2 Or 25Plus

    4800 baud. See Setting Up shows the new baud rate. and Testing System. C. Record the baud; you must also set the CTD baud rate in the SBE 33  Once all baud rates are set, you can (see Setting Up and Testing System).
  • Page 24: Setting Up And Testing System

    Click OK to save the settings and exit the dialog box. connector will interfere with communication with the SBE 33. Turn on power to the SBE 33. The SBE 33 status message displays in Seaterm: SBE 33/36 Deck Unit V3.0 setup:...
  • Page 25  Changing PC baud rate or parity disrupts communications, Type 1 and press the Enter key to put the SBE 33 in mode 1 (echo because they must match the characters to and from the instrument), which is the mode required for settings in Seaterm’s Configure...
  • Page 26 12. Send QS to put the CTD in quiescent (sleep) state (not applicable to SBE 49). 13. If using the SBE 33 with a NMEA device: Type @ to return to the setup menu. Type 3and press the Enter key to switch to mode 3. The display...
  • Page 27: Changing Communication Settings Between Sbe 33 And Computer

    SBE 19 or 25, which require 7 data bits and even parity. The procedure below applies to using an SBE 19 or 25 with an SBE 33 set at factory defaults. Once you change communication settings in the SBE 33, it will retain the new settings, even if you remove power.
  • Page 28: Checking Ctd Configuration (.Xmlcon Or .Con) File In Seasave

    Seasave of NMEA position data, NMEA depth data, and/or Surface PAR data extension. For example, for a CTD is dependent on the configuration file settings, not on the SBE 33 settings. with serial number 2375, Sea-Bird names the file 2375.xmlcon. You To verify the contents of the .xmlcon or .con configuration file:...
  • Page 29: Section 4: Mounting And Wiring System

    (www.seabird.com/products/third-party-equipment#CableAndWinch). Mounting SBE 33 Detachable rack mount ears are provided for mounting the SBE 33 to standard 19-inch electronics bays. Ensure that the back of the SBE 33, specifically the cooling fan and its vent, is not obstructed.
  • Page 30: Wiring System

    Water Sampler. The Serial Data channel sends commands to and channel is not connected to the receives replies from the CTD (through the Water Sampler). Connect the computer. computer to Carousel Data and Serial Data on the SBE 33 back panel using the supplied cables.
  • Page 31: Power To Sbe 33

    Power to SBE 33 WARNING! Some oceanographic vessels Verify that the SBE 33 back panel power selector switch is in the correct isolate (un-ground) the AC power position for your mains power source. Connect power to the AC Input ground circuit. If the SBE 33 is connector on the SBE 33 back panel using the supplied cable.
  • Page 32: Section 5: Operating System

    Click Configure Inputs. The Configure Inputs dialog box appears. A. Click the Serial Ports tab. - CTD Serial Port - Select the computer port connected to the SBE 33 Serial Data connector (which sends commands to and receives replies from the CTD, through the Water Sampler), and the baud rate, data bits, and parity.
  • Page 33 SBE Carousel (SBE 32, 32C, or 32SC), SBE ECO (SBE 55), or None (when used with PDIM) are only valid selections for SBE 33. Define serial port for water sampler operation on Serial Ports tab. Total number of bottles to be closed - up to ...
  • Page 34 C. If you selected NMEA position data added in the .xmlcon or .con file, Seasave initializes NMEA communications. Seasave automatically sends a command to put the SBE 33 in Mode 1 (do not transmit NMEA data) or Mode 2 (append NMEA and/or Surface PAR data to...
  • Page 35 Manual revision 017 Section 5: Operating System SBE 33 D. If you selected Check Scan Length in the Options menu, Seasave checks the .xmlcon or .con file to verify that the scan length defined by the configuration file matches the CTD (i.e., number of sensors and inclusion of NMEA and/or Surface PAR is as defined in the file).
  • Page 36: Firing Bottles From Sbe 33 Front Panel

    Firing Bottles from SBE 33 Front Panel Instead of firing bottles in Seasave, you can fire bottles directly from the SBE 33 if desired. Bottles fired from the SBE 33 can be in random or sequential order, regardless of the firing sequence you input in Seasave.
  • Page 37: Recovering Ctd And Water Sampler

    Manual revision 017 Section 5: Operating System SBE 33 Recovering CTD and Water Sampler WARNING! Rinse the CTD, auxiliary sensors, and Water Sampler with fresh water. If the CTD or Water Sampler stop working while underwater, are unresponsive to commands, or...
  • Page 38: Uploading Ctd Data From Memory

    When the cast is complete, turn the CTD switch off. Turn power to the SBE 33 off. Disconnect the CTD from the Water Sampler. Connect the CTD directly to the computer, using the data I/O cable provided with the CTD.
  • Page 39: Editing Raw Data File

    Manual revision 017 Section 5: Operating System SBE 33 Editing Raw Data File Sometimes users want to edit the raw data file before beginning processing, to remove data at the beginning of the file corresponding to instrument soak time, to remove blocks of bad data, to edit the header, or to add explanatory notes about the cast.
  • Page 40: Section 6: Routine Maintenance

    Cleaning SBE 33 Deck Unit To clean the SBE 33: Disconnect the power and any other cables from the back of the SBE 33. Using a soft cotton cloth dampened with warm water, clean the exterior of the SBE 33 with gentle pressure. Use special care cleaning around any connectors, to avoid getting water into them.
  • Page 41: Section 7: Troubleshooting

    Check the AC power source. Turn off the SBE 33, remove the power cable, wait 1 minute, and check the main fuse. Replace if necessary. If the main fuse blows again, there is probably a short in the main chassis AC wiring.
  • Page 42: Problem 3: Water Sampler Or Ctd Does Not Respond

    Problem 3: Water Sampler or CTD Does Not Respond Cause/Solution 1: WARNING! There may be no voltage on the sea cable. Turn off the SBE 33, remove the Use extreme caution when making power cable, and wait 1 minute. Connect a voltmeter (range set to at least these tests.
  • Page 43: Problem 5: Yellow Nmea Transmit Led Not Flashing

    Wiring may be incorrect. Check cables and connections between the SBE 33, NMEA navigation device, and computer. Cause/Solution 2: NMEA navigation device may be set to the wrong baud rate (SBE 33 requires 4800 or 9600 baud). Reset device’s baud rate. Cause/Solution 3: NMEA navigation device may not be transmitting data.
  • Page 44: Glossary

    Seasoft V2 includes Seaterm, SeatermV2, Seasave V7, SBE Data Processing. Seaterm – Sea-Bird’s Windows terminal program used to communicate with the SBE 33 and/or the CTD. SeatermV2 – Windows terminal program launcher. Depending on the instrument selected, it launches Seaterm232 (RS-232 instruments), Seaterm485 (RS-485 instruments), or SeatermIM (inductive modem instruments).
  • Page 45: Safety And Electrical Symbols

    Manual revision 017 Glossary SBE 33 Safety and Electrical Symbols Some or all of the following symbols may be used on the SBE 33: Symbol Description Potentially hazardous voltage. Hazardous! Voltage > 30 VDC may be present. Attention! There is a potential hazard; consult the manual before continuing.
  • Page 46: Appendix I: Nmea Navigation Device Message And Data Formats46

    Manual revision 017 Appendix I: NMEA Navigation Device Message and Data Formats SBE 33 Appendix I: NMEA Navigation Device Message and Data Formats Message Formats GGA - Global Positioning System Fix Data Time, position, and fix related data for a GPS receiver.
  • Page 47: Data Formats

    Note: If a Surface PAR sensor is also Seven bytes of NMEA position data are appended to each scan of hex data connected to the SBE 33, the from the instrument. Surface PAR data (three bytes) is appended before the position data.
  • Page 48: Appendix Ii: Nmea Message Simulation Program

    COM port. part of the SBE Data Processing installation). Use the NMEA Interface test cable (PN 801422 - supplied with the SBE 33) to connect NMEA Input on the SBE 33 to the simulation computer. The simulation...
  • Page 49 If the system works with the simulation program but does not work when connected to the actual NMEA navigation device, the problem is with the cable from the NMEA navigation device to the SBE 33, or in the NMEA navigation device itself.
  • Page 50: Appendix Iii: Pdim Electronics Disassembly/Reassembly

    PDIM Disassembly/Reassembly Manual revision 017 Appendix III: SBE 33 Appendix III: PDIM Electronics Disassembly/Reassembly Sea-Bird provides a jackscrew kit with the PDIM, to assist in removal of the connector end cap. The kit contains:  Jackscrew 2 Allen wrenches ...
  • Page 51: Appendix Iv: Replacement Parts

    Quantity Number Description Connect Serial Data and 9-pin DB-9F to Carousel Data on 171887 9-pin DB-9M cable, SBE 33 to computer 3 m (10 ft) COM ports 2-pin Connect NMEA Input on MS3106A-12S-3S to SBE 33 to computer 801422 9-pin DB-9S...
  • Page 52: Appendix V: Manual Revision History

     02/02 Create separate manual for SBE 33 (had been combined with SBE 32 manual). SBE 33 manual provides detailed information on how to use SBE 33.SBE 32 manual now explains how to assemble and use Carousel, and provides brief description of each interface method.
  • Page 53 03/14 Add information that Seasave 7.23.2 allows Surface PAR acquisition with SBE 49.  Add information on use of SBE 33 with PDIM in place of Water Sampler, when a Water Sampler is not available or needed.  Remove references to requiring Seasave version 7.19 or later (7.19 released > 4 years ago).
  • Page 54: Index

    Manual revision 017 Index SBE 33 Index .con file · 28 NMEA · 18, 24, 31 .xmlcon file · 28 data formats · 46 message formats · 46 simulation program · 48 troubleshooting · 43 NMEATest · 9, 20, 48 Back panel ·...
  • Page 55 Manual revision 017 Index SBE 33 Unpacking SBE 33 · 7 Winch · 29 Uploading data · 38 Wiring · 14, 30 Versions · 52...