5s) Print Print real-time data ----- ------ Attentions:OC-1000 portable gas detector only avaible to detect the gas concentration when the pump on. ② Treatment for the over measuring range. Please avoid use the device in concentration which exceeds the measurement range.
Page 4
, then after the ▲ “beep”, will keep the measuring data. Menu entrance:Under the testing interface, long press “ok” for 5s then will enter the menu. ⑥ Parameter revise instruction All parameter for OC-1000 could be revised with the “Back”, “OK”, ▼. ▲,...
1.Description OC-1000 series multi gas and dust particle detector, can be used for multi gas and dust, at most for 5 sensors. With the imported sensor and micro control technology, respond quikly, high accuracy and stability, good repeatability. With the rechargeable 4000mA lithium battery, continuous working for longer time.
◆ Support Chinese or English operation menu. ◆ With the function of temperature compensation, water-proof and dust-proof, equiped with the dust filter. ◆ With the temperature and humidity detection. 3.Technical parameter Product: Multi gas and dust particle detector Measuring Reffering to the gas list range:...
Page 7
Explosion-Pr Case Material : ExdII CT6 ABS+PC oof grade: -30 ~ Protection Working IP66 grade: temperature: 60℃ ≤95%RH(no Working 4000mA rechargeable lithium battery Working humidity: n-condensin power: -30Kpa ~ Size and 220*85*55 mm(L×W×H) Working pressure: weight: 0.5 Kg(Net weight) 100Kpa Accessory:...
4.Structure 5.Operation instruction 5.1 Keys 、▼、Back、OK、run/stop、 Total seven keys: and print. ▲ Three interfaces:Testing interface、Menu interface、Setting interface...
Page 9
5s) Print Print real-time data ----- ----- Attentions:OC-1000 portable gas detector only avaible to detect the gas concentration when the pump on. 5.2 Power on When power off, long press key for 5s to activate the backlight and red indicator, then “beep”, then the detector power on. Then the display will be sensor testing(3s), sensor information(1s), preheating for 60s with the pump on.
6.Operation interface 6.1 Testing interface display After enter into the testing interface, for different gas quantity, will display different interface: as the following picture, displaying one gas, two gas, three gas, four gas orderly. As the above displaying, time, temperature, pump running, battery and humidity.
Page 11
6.2 Function menu instruction Long press “OK” key for 5s, will enter to the menu, then will display nine sub menu: Basic set, History data, Data analyse, Zero cali, Range calibration, Alarm param, Detection /Storage/Print , Date time, Reset factory. On this menu by “ ▼”...
Page 12
Attention: zero calibration must be in the clean air. 6.2.5 Target Calibrate(Only professionl engineer operation allowed) OC-1000 multi gas and dust particle detector can make the six level target concentration calibration for the detecting gas, as the picture shows, this must with the standard concentration gas, pressure relief valve, flowmeter, calibration cap, and good connection of gas, or will be prohibited.
Operation details: user move the cursor to “Alarming mode”, press “ok”, by “ ▼” can change the alarming mode, after switch then ▲” or “ press “ok” will save the setting. 6.2.7 Detection storage With this function user can set the detecting mode, storage mode, print mode, the detecting mode including the realtime detecting and timing detecting, storage mode including the manual save , auto save and close(not save), print mode including hand, auto, close.
Prohibit drop from high space. ◆ Under the high temperature or pressure is not allowed to use, if high ◆ humidity, then need to add the device to decrese the humidity. Human damage is not free repairment. ◆ Befor enter into dangerous area, people need to be electrostatic ◆...
5V, then need to send back to manufacturer for testing. 9.Data logger OC-1000 with the data logger function, the charging port is the data output port, the operation steps as the following: 1、Install the export software on the computer.
Page 16
5 、 On the software,set the detector address and channel number(same with the “address setting”and “channel setting”in the “Setting”), start time and deadline(The save time of history data should be in this period range). 6、Click the “search”can find the above period gas concentration for the according channel gas,the data also could be saved as the EXCEL form by click the “Read Data”...