Page 1
C.A 6541 Mégohmmètres Megohmmeters Megohmmeter C.A 6543 Megaohmmetri Megaóhmetros F R A N Ç A I S Notice de fonctionnement E N G L I S H User's manual D E U T S C H Bedienungsanleitung I T A L I A N O Libretto d’Istruzioni...
The voltage of parts indicated with this symbol can be ≥ 120 V DC. For safety reasons, this symbol lights up on the LCD screen as soon as a voltage is generated. Thank you for purchasing a C.A 6541 or C.A 6543 megohmmeter. To get the best service from this instrument : read this user’s manual carefully...
Megohmmeters Megohmmeters C.A 6541 and C.A 6543 are portable instruments, each housed in robust casing with lid and operating on batteries (C.A 6541), or on a rechargeable battery and the mains (C.A 6543). They can be used to measure :...
Page 5
“Megohm View” PC software (option for the C.A 6543) The PC software makes it possible to: - retrieve data from memory, to plot graphs of the changes in insulation as a function of the time over which the test voltage is applied, R (t), - to print out protocols of personalised tests depending on the user’s needs,...
1 yellow START / STOP button: start/stop of the measurement 6 buttons (C.A 6541) or 8 buttons (C.A 6543) in elastomer, each with a first and a second function: Selection of the second function (in yellow italics below each button)
Mains plug (direct operation on AC network/ battery recharge) RS 232 serial INTERFACE male plug (9 pin) for connection to a PC or printer Note : The battery compartment (C.A 6541) or the rechargeable battery compartment (C.A 6543) is situated inside the unit casing.
Page 8
During insulation measurement, the elapsed time or the output voltage is displayed. After recording a set of data (C.A 6543), the small display also indicates the time and date in MR (memory recall) mode. It also indicates the memory address with the OBJ. TEST number (see § 2.2.3.
3. MEASUREMENT FUNCTIONS AC / DC Voltage Select any insulation, resistance or continuity measurement position with the rotary selector switch and the device is automatically in AC / DC voltage measurement mode. The voltage is permanently measured and is shown on the small display. Proceeding with a measurement is prohibited if an external voltage is present at the terminals before pressing START / STOP.
V-TIME button Main Small Small display Display display if the MR button is pressed (C.A 6543) Resistance duration (min. sec) date, time, test voltage, OBJ : TEST duration (min. sec) date, time, test voltage, OBJ : TEST duration (min. sec)
4. SPECIAL FUNCTIONS Button Use this button to select the second function activated by the buttons, it is always associated with the symbol. This symbol disappears upon pressing the function button chosen, except if the button is activated. In this case, it only disappears when the button is pressed again or if other function buttons are pressed.
Page 12
On these kinds of parts, measurement is disturbed at the beginning due to interference currents (from capacitive charges, dielectric absorption) which eventually cancel each other out. Therefore, in order to get an accurate measurement of the leakage current that is representative of the insulation, it is necessary to carry out the measurement over a long period of time so as to reduce the influence of any interference currents when measurement is commenced.
SET-UP (the default value is 30 seconds). It is possible to save more than 20 samples, available processing memory space permitting. On the C.A 6543, the number of samples that can be saved is only limited by the read-write memory space available.
/ SMOOTH Button First function Used to select the desired parameter to be modified – the selected parameter flashes. It is modified using the button (see § 4.6). Second function SMOOTH Used to activate a digital filter for insulation measurement. It only has an effect on the displayed values (which are smoothed) and not the actual measurements.
Page 15
Parameters Command button Display to modify main small symbols values Duration of 15 : 00 min. sec, 1 - 59 min “ time-controlled test” and 2 times for R-DAR-PI second time first time min : sec 00 : 59 PI calculation (10 min) (1 min) Time between interval...
Page 16
4.7.1 Wiping the memory In SET-UP, press the MEM button: - The MEM symbol flashes - The small display indicates ALL - The main display indicates cLR To wipe the entire memory, press the MEM button again for 2 seconds: - The MEM symbol is displayed without flashing - The main display indicates FrEE To wipe the contents of a specific OBJ : TEST number:...
4.7.5 Barring of insulation measurement This function prohibits insulation measurement from being carried out at certain test voltages. This makes it possible the entrust the instrument to less experienced operators, for specific applications for example (telephone ...). In SET-UP, press 5 times or more (depending on voltage to be barred): - The test voltages appear in succession on the small display with the button and On/OFF on the...
The open circuit voltage is equal to the voltage of the battery (C.A 6541) or rechargeable battery voltage (C.A 6543). NOTE : the R-DAR-PI and V-TIME buttons are not active in this function.
6. MEMORY / RS 232 (C.A 6543) RS 232 Specifications The signalling speed in Bauds can be set to 300, 600, 1200, 2400, 4800, 9600, or “Parallel” for the printing of results on parallel printers via the optional serial-to-parallel adapter.
Page 20
An object represents a ‘box’ into which we can place 99 tests. Thus, an object can represent a machine or an installation upon which a certain number of various measurements (insulation, resistance, continuity) are carried out. 1. When the button is pressed, the MEM symbol flashes and the small display indicates the first free OBJ : TEST number, e.g.
Page 21
If a new is OBJ is selected, TEST is automatically set to the maximum memorised number. At this stage it is possible to consult the entire memory with the buttons, since the measurement values corresponding to the selected OBJ : TEST number are displayed on the main display.
Printing out measured values (PRINT/PRINT MEM button) (C.A 6543) If you use a serial printer, choose the appropriate communication speed in the SET-UP menu, between 300...9600 bauds, then program the printer to the format run by the instrument (see § 6.1).
Page 23
Continuity or resistance measurement CHAUVIN ARNOUX C.A 6543 Instrument number : 000 001 CONTINUITY TEST or RESISTANCE TEST OBJECT : 01 TEST : 01 (only printed in MR) Description: ........................Date : ......... 31.03.1998 Start time : ........14 : 55 Test current : ........
5. To start printing out values that haven’t been saved (immediate printing), press PRINT after a measurement. 6. To start printing out saved measurements, press the “PRINT MEM” button WARNING: This adapter is designed exclusively for use with the C.A 6543 and is not suitable for any other application. 7. SPECIFICATIONS...
Page 25
Accuracy : ±1% R ±3 cts, AC sinusoidal or DC voltage Input resistance : approx. 750 kΩ WARNING: This instrument is Cat III 600 V, according to EN 61010-1 + A2 (ed. 95). It shall not be connected to voltages > 600 V in relation to the earth. 7.2.2 Insulation resistance Method: Voltage-current method according to EN 61557-2 (ed.
Page 26
100 V test voltage Resistance (k 250 V test voltage Resistance (k 500 V test voltage Resistance (k...
Page 27
1000 V test voltage 1200 1000 1000 Resistance (k Typical settling time of the measurement as a function of the elements tested These values include the influences on the auto-range and supply voltage regulation system due to the charge of the capacitive component. Non-capacitive load Load with 1 µF capacitor 1 MΩ...
Power supply The device supply is via: - 8 x 1.5 V batteries, LR 14 (C.A 6541) - Rechargeables battery NiMh (C.A 6543) Recharge: 85 to 256 V / 50-60 Hz (electrical safety: 256 V Cat. III) Instrument C.A 6541 C.A 6543...
Temperature influence on measurement accuracy (for all functions) ±0.15 % per °C Altitude: < 2000 m Mechanical specifications Outside dimensions of casing (L x W x H): 240 x 185 x 110 mm Weight : approx. 3.4 kg Conformity to international standards Electrical safety according to: EN 61010-1 + A2 (ed.
8.1.2 Recharging the battery (C.A 6543) If the symbol appears flashing, then it is necessary to recharge the battery. Connect the instrument to the mains AC network via the connector, charging starts automatically: bAt on the small display and CHrG on the main display, signifies fast charging in progress...
Set of 2 test probes (red + black) .................. P01.1018.55 Set of 3 x 3 m safety leads (red + blue + black guarded) ..........P01.2951.70 PC software (C.A 6543) ....................P01.1019.38 Serial printer (C.A 6543) ....................P01.1029.03 Serial-to-parallel adapter (C.A 6543) ................P01.1019.41 Spares : 3 x 1.5 m safety leads straight-straight (red + blue + black guarded ) ......