DeVilbiss ® 4 & 5-Liter Series Oxygen Concentrators Service Manual For Product Numbers: 515DS 515DZ 515KS 515KZ 515NS 515UK CAUTION-Federal (U.S.A.) law restricts this device to sale by or DANGER – NO SMOKING on the order of a physician.
T A B L E O F C O N T E N T S GENERAL INFORMATION Introduction .............................................4 Important Safeguards..........................................5 Safety Precautions and General Warnings..................................5 UNPACKING AND SETUP Initial Inspection............................................6 Patient Setup............................................6 Operating Instructions ..........................................6 MAINTENANCE Patient Alert System ..........................................8 Routine Patient Maintenance.......................................8 Periodic Homecare Provider Preventative Maintenance ..............................8 Between Patient Maintenance ......................................9...
G E N E R A L I N F O R M A T I O N INTRODUCTION This service manual was designed to provide Sunrise Medical Respiratory Products Division qualified service technicians and homecare providers with the proper main- tenance, service, safety, and repair procedures for the DeVilbiss Oxygen Concentrator.
G E N E R A L I N F O R M A T I O N IMPORTANT SAFEGUARDS Read all instructions before operating the oxygen concentra- tor. Important information is highlighted by these terms: G. WARNING: Electric shock hazard. Do not remove cab- inet.The cabinet should only be removed by a qualified Sunrise Medical homecare provider.
U N P A C K I N G A N D S E T U P INITIAL INSPECTION Oxygen Tubing Only Connection (Figure 1) It is suggested that an initial inspection be performed upon 1. Thread the cannula fitting (part #CN100) onto the oxygen receiving the oxygen concentrator.
Page 7
WARNING:The plug on the DeVilbiss 515DZ and off and the red “Service Required” light will turn on. 515DS concentrators has one blade wider than the NOTE: Refer to “Specifications” for specific alarm settings. other.To reduce the risk of electric shock, this plug is intended to fit in a wall outlet only one way.
M A I N T E N A N C E PATIENT ALERT SYSTEM Cannula/Mask and Tubing The DeVilbiss Oxygen Concentrator patient alert The patient should clean and replace the cannula or mask and system will detect unit component failure.This system is com- tubing as instructed by the manufacturer.
M A I N T E N A N C E b. The concentrator must operate for a minimum of 20 min- 7. Leak test the HEPA filter fittings. Apply a leak test solution utes before checking the oxygen concentration. such as Epi-SEAL ®...
T R O U B L E S H O O T I N G SYSTEM OPERATION NORMAL OPERATING SEQUENCE The DeVilbiss Oxygen Concentrator uses a pressure swing When the concentrator is turned “On,” the following normal adsorption system.The air is drawn into the unit through air cycling sequence should be observed by attaching pressure filters and into a double-head compressor.
T R O U B L E S H O O T I N G SIMPLIFIED TROUBLESHOOTING The key to simple troubleshooting is to recognize which type of problem exists and select the most effective approach to WARNING: Mechanical Hazard. Keep fingers, loose solving the problem.The different types of problems and the clothing, etc.
T R O U B L E S H O O T I N G TROUBLESHOOTING CHART A Visible Alarm Audible Alarm Compressor Power Light Other Symptoms Possible Cause Possible Remedy Pulsating air noise Intake filter not in place or defective Check filter and replace if necessary Compressor intake hose disconnected Reconnect hose...
T R O U B L E S H O O T I N G TROUBLESHOOTING CHART B Visible Alarm Audible Alarm Compressor Power Light Blinking Pulsing Other Symptoms Possible Cause Possible Remedy Fan off Line cord not properly installed or defective Insert plug in receptacle or replace line cord No power at receptacle Check building circuit breaker or fuse, or have...
T R O U B L E S H O O T I N G TROUBLESHOOTING CHART F (OSD UNITS ONLY) Compressor Power Light Other Symptoms Possible Cause Possible Remedy No OSD lights are illuminated. Defective OSD or malfunctioning concentrator. Check concentration with an oxygen analyzer. If the concentration is within specification, replace the OSD (early OSD models only).
C O M P O N E N T T E S T I N G , R E P A I R , A N D R E P L A C E M E N T PROPER REPAIR PROCEDURES The DeVilbiss Oxygen Concentrator is designed for ease of service.To aid service personnel, a Service Kit (part #444-501) WARNING: Electric shock hazard.
C O M P O N E N T T E S T I N G , R E P A I R , A N D R E P L A C E M E N T ACCUMULATOR PRESSURE TEST To check accumulator pressures: 1.
Page 17
C O M P O N E N T T E S T I N G , R E P A I R , A N D R E P L A C E M E N T To test the compressor operating voltage (Figure 4): place.When removing the heads, be sure to keep each head and its components with the correct compressor side.
C O M P O N E N T T E S T I N G , R E P A I R , A N D R E P L A C E M E N T If the four-way does not shift, the same bed will continue to To clean the four-way valve: (Figure 13) pressurize, and the pressure relief valve will relieve the pressure.
C O M P O N E N T T E S T I N G , R E P A I R , A N D R E P L A C E M E N T PRESSURE REGULATOR To remove and replace the PC board (without integral OSD) (Figure 7A): The pressure regulator stabilizes the flow of oxygen to the...
F I G U R E S , D I A G R A M S , A N D V I E W S Exterior Views The following figures show the exterior of the DeVilbiss 4 & 5-Liter Series Concentrators. Figure 1 Front Figure 2...
Page 23
F I G U R E S , D I A G R A M S , A N D V I E W S Figure 1. Exterior Front View OSD Indicator Lights Service Required Power Light (OSD units only) Light (Red) (green) Power Switch...
Page 24
F I G U R E S , D I A G R A M S , A N D V I E W S Figure 2. Exterior Side View Front cover Rear cover Cabinet screw Power Cord Strap Power Cord Caster Exhaust Grill...
Page 25
F I G U R E S , D I A G R A M S , A N D V I E W S Figure 3. Exterior Rear View Cabinet Screws Filter Door Filter Communication Port (OSD units only) Cabinet Screws...
Page 26
F I G U R E S , D I A G R A M S , A N D V I E W S Figure 4A. Interior Rear View (early OSD models) Bib Screws Intake Filter Oxygen Sensor Assembly (OSD units only) Accumulator Tank Communication Port...
Page 27
F I G U R E S , D I A G R A M S , A N D V I E W S Figure 4B. Interior Rear View (later OSD models) Bib Screws Accumulator Tank Intake Filter Communication Port (OSD units only) Exhaust Muffler (Hidden)
Page 28
F I G U R E S , D I A G R A M S , A N D V I E W S Figure 5A. Interior, Front View (early OSD models) Wire Harness (main) PC Board Right Sieve Bed Left Sieve Bed Wires to Power Switch and Circuit...
Page 29
F I G U R E S , D I A G R A M S , A N D V I E W S Figure 5B. Interior, Front View (later OSD models) PC Board Right Sieve Bed Left Sieve Bed Wire Harness (main) Wires to Power Hose to Flow Meter...
Page 30
F I G U R E S , D I A G R A M S , A N D V I E W S Figure 6A. Interior, Front Lower View (early valve version) 1/2" ID Exhaust Hose (Nitrogen) Sieve Bed Cable Tie Slot for Bib Tab Pressure Inlet...
Page 31
F I G U R E S , D I A G R A M S , A N D V I E W S Figure 6B. Interior, Front Lower View (later valve version) Sieve Bed Cable Tie 1/2" ID Hose Slot for Bib Tab (Braided) Pressure Inlet...
Page 32
F I G U R E S , D I A G R A M S , A N D V I E W S Figure 7A. Interior, Front Upper View (early OSD models) Four-Way Pressure Tubing to Sieve Bed Oxygen Outlet Oxygen Outlet Sieve Bed...
Page 33
F I G U R E S , D I A G R A M S , A N D V I E W S Figure 7B. Interior, Front Upper View (later OSD models) Four-Way Tubing to Oxygen Outlet Hour Meter Sieve Bed Oxygen Outlet Sieve Bed...
Page 34
F I G U R E S , D I A G R A M S , A N D V I E W S Figure 8. Interior, Behind Bib View Final Bacteria Filter Oxygen Outlet Handle Port Screw Power Switch Light Panel Final Check Valve Ribbon Connector...
Page 35
F I G U R E S , D I A G R A M S , A N D V I E W S Figure 9A. Interior,Top View (early OSD models) Oxygen Sensor Sieve Bed Accumulator Tank Sieve Bed Hose Clamp Pressure Regulator Hour meter...
Page 36
F I G U R E S , D I A G R A M S , A N D V I E W S Figure 9B. Interior,Top View (later OSD models) Fixed Orifice Sieve Bed Sieve Bed Pressure Regulator Accumulator Tank Hour meter 1/4"...
Page 37
F I G U R E S , D I A G R A M S , A N D V I E W S Figure 10. Interior, Rear Lower View Pressure Exhaust Hose Hose Clamp for (Silicone) 1/2" ID 1/2"...
Page 38
F I G U R E S , D I A G R A M S , A N D V I E W S Figure 11. Compressor Intake Fitting Compressor Compressor Filter Pressure Relief Exhaust Fitting Valve Motor Mount Electrical Connector Compressor Mounting Plate...
Page 39
F I G U R E S , D I A G R A M S , A N D V I E W S Figure 12.Compressor With Head Removed Retainer Screw Valve Flapper Head Gasket Valve Plate Compressor Cup Retainer Sleeve Compressor Head Pressure Relief Valve...
Page 40
F I G U R E S , D I A G R A M S , A N D V I E W S Figure 13. Four-Way Valve End Cap Screws End Cap Left Solenoid Valve Body Right Solenoid End Cap End Cap Screws Spring...
Page 41
F I G U R E S , D I A G R A M S , A N D V I E W S Figure 14 Compressor (exploded view) 1 Connecting rod, eccentric 10 Head and bearing assembly 11 Valve plate assembly 2 Piston cup 12 Valve keeper strip 3 Cylinder sleeve...
FITNESS FOR A PARTICULAR PURPOSE, ARE LIMITED TO Limited Warranty THE DURATION OF THE EXPRESS LIMITED WARRANTY The DeVilbiss 515DS/515DZ Oxygen Concentrators, manufac- AND TO THE EXTENT PERMITTED BY LAW ANY AND ALL tured and sold by Sunrise Medical, are warranted as stated IMPLIED WARRANTIES ARE EXCLUDED.THIS IS THE EXCLU-...
O R D E R I N G I N F O R M A T I O N A N D P A R T S L I S T ORDERING INFORMATION DAYS. COMPONENTS WILL THEN BE BILLED AT THE REBUILT COST;THERE WILL BE A CHARGE FOR SHIPPING.
**Note:This corresponds with a sound level of 50 dbA overall *Note:The OSD performance at 10°C to 35°C, 95% R.H. average as measured according to the guidelines of the through the voltage range has been verified on a 515DS now defunct ANSI Z79.13-1981 standard. at 670m.