Unauthorised conversion and manufacturing replacement parts ..9 Warranty claims and liability ............... 9 ABOUT YOUR WATER PURIFIER ............10 The Select edi 60 range ..............10 Key components of the Select edi 60 ..........10 5.2.1 Touch screen display ..............10 5.2.2 Storage tank ..................
Page 3
OPERATING MANUAL FOR SELECT EDI 60 10.5 Water storage ................33 10.6 Electrical specifications / connections ........... 33 10.6.1 Mains supply ................33 10.6.2 Fuse rating / type................ 33 10.6.3 Alarm port connection details ............. 33 10.6.4 USB Mass storage device ............34 10.6.5...
OPERATING MANUAL FOR SELECT EDI 60 2 PREFACE This Operating Manual provides all of the information and instructions needed to install, operate and maintain your SELECT edi 60 water purification unit. Please read the instructions carefully and make sure that you fully understand the information given before operating the unit.
OPERATING MANUAL FOR SELECT EDI 60 3 CONTACT DETAILS Should you require any additional information relating to the servicing, maintenance, spares and consumables, simply contact SUEZ using the details below:- Water Purification Systems Ltd Bandet Way Thame Oxfordshire OX9 3SJ United Kingdom (Or your local distributor;...
Country-specific requirements standards and regulations must be observed. 4.3 Intended use. The Select edi 60 unit has been specifically designed for indoor use only to purify potable drinking water for use in laboratory applications or similar applications requiring highly purified deionised water. Refer to Section 11.5 for details of feed water limitations and Section 11.2 for treated water...
If the unit’s performance becomes impaired and any remedial work appears to be outside the scope of this manual, do not operate the unit and seek advice from SUEZ. The unit must only be serviced and maintained by SUEZ Service personnel or SUEZ approved and trained technicians.
9001:2008 quality control process. If, however, you are unsatisfied with the unit, please contact SUEZ Ltd. Any warranties guaranteed by SUEZ with respect to the Select edi 60 will be void if the equipment is not used in accordance with the written instructions provided within this Operating Manual.
5.1 The Select edi 60 range The Select edi 60 is only as one model producing a makeup flowrate of 10 litres/hour. The above output is based on a typical 8 hour working day at 60psi (4 bar) inlet pressure, 25 C and maximum of 400 ppm total dissolved solids.
OPERATING MANUAL FOR SELECT EDI 60 5.2.3 Removable front cover. The front cover is easily removed and provides access to the consumable cartridge media pack. Touch screen display Use the two grab slots to gently pull away the front cover to gain...
OPERATING MANUAL FOR SELECT EDI 60 6 OPERATING YOUR WATER PURIFIER Your Select edi 60 is operated via the touch screen display. Buttons are provided on the display to access start menus, set up functions and dispense water. The display also provides a mimic of the system when in operation.
OPERATING MANUAL FOR SELECT EDI 60 6.2 Menu structure Power-On Screen Start User Menu Clean routine Setting Menu Time & Date Alarms Maintenance System Menu Product Information Engineering menu Set Unit ID Auto Restart Set Log Frequency Data Logging 6.3 Clean routine menu Refer to Section 8 for details of disinfection procedure.
OPERATING MANUAL FOR SELECT EDI 60 6.4.1 Time and date If you select “Set Time” the screen opposite will be displayed. Set Time The flashing cursor identifies which BACK digit you can adjust. By using the ▲/▼buttons you can either increase or decrease the 16:03 number value.
OPERATING MANUAL FOR SELECT EDI 60 6.4.2 Alarms The Select edi 60 unit can be programmed to sound an alarm buzzer if certain performance limits are exceeded or if the unit has detected a serious fault. Within the “Alarms” menu alarm set points can be set for the final purified water quality and for the temperature of the purified water.
Page 16
OPERATING MANUAL FOR SELECT EDI 60 Set quality alarm A minimum water quality is required to protect the EDI Cell from fouling. An alarm can be set that will warn if the water quality exceeds the set value (5-10us nom). Enter the desired water quality alarm limit in the set quality alarm menu.
OPERATING MANUAL FOR SELECT EDI 60 Set temperature alarm If dispensed water must be supplied below a certain temperature an alarm can be set. Enter the maximum water temperature in the “”Set Perm Temp alarm” menu. During “Processing” should the water temperature rise above this, the navigation bar will display “Temperature High”...
OPERATING MANUAL FOR SELECT EDI 60 The maintenance menu provides the capability to set defined dates for replacement of consumable items in the device. These dates, when expired, will trigger an alarm which will be displayed on the mimic screen 6.4.4 Systems menu...
Displays a six digit number; the number can be changed within this menu. “Unit Type” Can be set to “Select edi 60” or “Select edi 60 + PP”. “Maintenance” Is used to quantify the number of EDI operations and water volume produced since commissioning.
Page 20
OPERATING MANUAL FOR SELECT EDI 60 “Line Cell Constants” There are three line cells within the unit, each cell constant can be set independently and are displayed. The accepted input range will be 0.05 – 1.99. If a value outside of this range is entered, the value will be ignored, and the previous value retained.
Page 21
OPERATING MANUAL FOR SELECT EDI 60 detect if a USB memory device is connected. If no device is detected, a message “Memory Stick Not Detected” shall appear for a few seconds on the screen. The stages of data transfer and programming sequence will be displayed.
Page 22
OPERATING MANUAL FOR SELECT EDI 60 Set log frequency Set Log Frequency BACK MINUTES ENTER The device has an inbuilt data logger that record status of the unit and any alarm events and performance data. Data is written to the memory.
OPERATING MANUAL FOR SELECT EDI 60 Data is stored to the data stick as a .CSV file which may be viewed using Microsoft Excel. “Clear data logger” After downloading data it is recommended to clear the data logger memory as it has a limited capacity which when exceeded historical data will be over written Selecting and accepting clear data logger clears the data file log.
OPERATING MANUAL FOR SELECT EDI 60 6.5.3 Engineer menu defaults ENGINEER MENU Default settings Serial Number 000000 Unit Type Select edi 60 Select edi 60 + PP Maintenance EDI Run Hours EDI On / Offs Total Volume Setup Line Cell Constants Outlet Line Cell 0.12...
OPERATING MANUAL FOR SELECT EDI 60 7 MAINTENANCE 7.1 Replacing the media pack Scan QR code for video clip on how to replace the media pack • Switch the unit off by pressing the STOP button. • Leave the unit to stand for 30 seconds to relieve any internal pressure.
OPERATING MANUAL FOR SELECT EDI 60 7.2 Cleaning the inlet screen filter It is recommended that this filter is cleaned at least once per year. The inlet screen filter is located in the rear of the unit and can be cleaned as follows: •...
OPERATING MANUAL FOR SELECT EDI 60 • To obtain access to the filter screen unscrew the black knurled top anti- clockwise. The screen will then be exposed. Pull the screen away from the housing and clean away any debris. • Once cleaned, replace the screen and black top.
Page 28
OPERATING MANUAL FOR SELECT EDI 60 Location of lithium battery on main PCB Main PCB Lithium PCB battery Part No. (R083349) Ref Section 11.4.5 L300180 Iss:A...
• Pull away the front door cover to reveal the media cartridge packs. • Remove the left hand Pre-Treatment pack and discard. o To sanitise the unit insert a cleaning pack adaptor (Suez part number L998549) into the left hand media pack location.
OPERATING MANUAL FOR SELECT EDI 60 10 TROUBLE SHOOTING The touch screen communicates faults with the system via both the mimic screen and text lines. A fault finding guide with corrective actions is shown below. 10.1 General fault conditions Symptom/s...
Page 32
NOTE: If remedial actions do not resolve the problem, turn off the water supply, isolate the unit from the electrical supply and contact either your authorised supplier or call SUEZ Service Department. DO NOT ATTEMPT TO DISMANTLE THE UNIT AND CARRY OUT ANY REPAIRS, UNLESS TRAINED AND APPROVED TO DO SO, WITHOUT FIRST CONTACTING SUEZ LIMITED OR YOUR AUTHORISED SUPPLIER.
OPERATING MANUAL FOR SELECT EDI 60 10.6.4 USB Mass storage device Specification: FAT 16 formatted USB memory stick 10.6.5 Main PCB Battery Specification Voltage: Type: Lithium, CR2032 10.7 Feed water specification 10.7.1 Feed water quality The unit has been designed to only operate on a potable water supply conforming to current EC Directive “Relating to the quality of water...
OPERATING MANUAL FOR SELECT EDI 60 10.8 Water services connections All water feed connections are found on the rear of the unit. Using the installation kit provided attach the water feed, Low pressure drain, high pressure drain and overflow tubing as shown below.
OPERATING MANUAL FOR SELECT EDI 60 10.9 Weights and dimensions Height Unit Weight kg Width (mm) Depth (mm) (mm) 35 (shipping) Select edi 60 55 (working)* *Weight with full tank of 20 litres 10.10 Environmental Room storage and operating 5 to 40...
OPERATING MANUAL FOR SELECT EDI 60 11 INSTALLATION For details on installing the unit scan the QR code below to gain access to video clip or alternatively follow the procedure below. The unit can either be wall or bench mounted. Refer to Section 12.4.2 for details on how to install the wall mounting kit.
Page 38
OPERATING MANUAL FOR SELECT EDI 60 L300180 Iss:A...
OPERATING MANUAL FOR SELECT EDI 60 11.3 Cartridge media pack Your Select edi 60 is supplied with a Pretreat PT8 cartridge pack and a blank ‘Flushing pack’ Note: The PT8 pack must be fitted in the left hand position in the front of the unit and the Flushing pack should be fitted in the right hand position.
OPERATING MANUAL FOR SELECT EDI 60 11.4.2 Wall mounted Installation steps Find a suitable mounting position. The wall bracket MUST be fitted to a load bearing wall of masonry or concrete material capable of supporting the weight of the bracket, unit, tank and full load of stored water.
Page 41
OPERATING MANUAL FOR SELECT EDI 60 If the unit is supplied with an internal tank, fix the tank support bracket to the bracket base plate using 3 off washers and nylock nuts supplied. Remove the feet from the underside of the unit and discard the feet and fixing screws.
OPERATING MANUAL FOR SELECT EDI 60 Tighten the front and rear retaining screws to fix the unit firmly to the plate and bracket. Bracket operation Empty the storage tank of water to make the unit lighter to move. Loosen the four slide plate fixing screws (DO NOT REMOVE). Pull the unit and slide plate forwards.
OPERATING MANUAL FOR SELECT EDI 60 12 APPENDIX 12.1 MSDS Material Safety Data Sheets for the following accessories are available upon request or can be downloaded from the SUEZ web site, see Section 3 for details. • PT 8 Media Pack L300180 Iss:A...
OPERATING MANUAL FOR SELECT EDI 60 12.2 Pushfit water connections To make a connection, the tube is simply pushed in by hand; an integral collet locking system then holds the tube firmly in place without deforming it or restricting flow.
OPERATING MANUAL FOR SELECT EDI 60 12.3 Process Refer to flow diagram on following page. All external hydraulic and electrical connections are positioned at the rear of the unit to protect them from accidental damage. On entering the Select Analyst an internal feed water strainer removes any large debris remaining in the feed water supply.
Page 46
OPERATING MANUAL FOR SELECT EDI 60 L300180 Iss:A...
OPERATING MANUAL FOR SELECT EDI 60 12.4 CE declaration A copy of Declaration of Conformity can be obtained from SUEZ upon request or from our website at www.suezwater.co.uk L300180 Iss:A...
OPERATING MANUAL FOR SELECT EDI 60 12.5 WEEE declaration Users in the United Kingdom who wish to discard electrical and electronic equipment that was supplied by SUEZ should contact:- Tel: No: 0844 873 1034 E-mail: info@complydirect.com Fax No: 0844 873 1035 SUEZ Registration No.
Need help?
Do you have a question about the SELECT EDI 60 and is the answer not in the manual?
Questions and answers